Automated cardiac arrest detection using wrist-derived photoplethysmography during withdrawal of life-sustaining treatment: a prospective clinical validation study

Background Automated cardiac arrest detection and alerting using wearable technology has the potential to shorten recognition delays for unwitnessed out-of-hospital cardiac arrest. In DETECT-1A and -1B, a photoplethysmography-based detection model was developed and validated in patients with induced cardiac arrest. This study evaluates model performance in true cardiac arrests following withdrawal of life-sustaining treatment. Methods Prospective, single-center study in adult ICU patients with planned withdrawal of life-sustaining treatment. Patients wore a photoplethysmography-wristband (CardioWatch) until death. Continuous ECG and invasive arterial pressure served as reference standards. The previously developed rule-based algorithm was refined using two training cohorts and evaluated in a separate test cohort. Endpoints were sensitivity for cardiac arrest detection and false positive alerts. Findings Forty-four patients were included (training 1: n=10; training 2: n=11; test: n=23), median age 65 years; 75% male, all with non-shockable cardiac arrest. Sensitivity for cardiac arrest detection was 100% (10/10; 95% confidence interval [CI] 66–100%) and 90% (9/10; 95% CI 54–99%), in training 1 and 2, respectively. In the test set, sensitivity was 100% (23/23; 95% CI 82–100%), with one false positive alert. Cardiac arrest was detected at a mean arterial pressure of 30 mmHg (IQR 24–35) and pulse pressure of 13 mmHg (IQR 11–19). Interpretation Cardiac arrest can be detected with high sensitivity using wrist-derived photoplethysmography, providing first evidence on model performance in true cardiac arrest, specifically in non-shockable cases. Findings support further development of wearable-based cardiac arrest detection technologies to enable earlier recognition for unwitnessed cardiac arrest. Funding Dutch Heart Foundation, Radboudumc.

Authors

Institutions

Publication Details

Journal
The Lancet Regional Health - Europe
Published
2026-07-24
DOI
https://doi.org/10.1016/j.lanepe.2026.101791
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated cardiac arrest detection using wrist-derived photoplethysmography during withdrawal of life-sustaining treatment: a prospective clinical validation study

Aysun Cetinyurek‐Yavuz, Eelko Ronner, Roos Edgar, Niels van Royen et al.
The Lancet Regional Health - Europe
Cardiac Arrest and Resuscitation
article

Automated cardiac arrest detection using wrist-derived photoplethysmography during withdrawal of life-sustaining treatment: a prospective clinical validation study

Aysun Cetinyurek‐Yavuz, Eelko Ronner, Roos Edgar, Niels van Royen, Judith Bonnes, Kambiz Ebrahimkheil, Rypko Beukema, Peter Stas, Lente Pol, Catharina E. Jansen, Cornelia W. E. Hoedemaekers, E Boersma, Marc A. Brouwer, Ruud C. van Kaam
article en

Abstract

Background Automated cardiac arrest detection and alerting using wearable technology has the potential to shorten recognition delays for unwitnessed out-of-hospital cardiac arrest. In DETECT-1A and -1B, a photoplethysmography-based detection model was developed and validated in patients with induced cardiac arrest. This study evaluates model performance in true cardiac arrests following withdrawal of life-sustaining treatment. Methods Prospective, single-center study in adult ICU patients with planned withdrawal of life-sustaining treatment. Patients wore a photoplethysmography-wristband (CardioWatch) until death. Continuous ECG and invasive arterial pressure served as reference standards. The previously developed rule-based algorithm was refined using two training cohorts and evaluated in a separate test cohort. Endpoints were sensitivity for cardiac arrest detection and false positive alerts. Findings Forty-four patients were included (training 1: n=10; training 2: n=11; test: n=23), median age 65 years; 75% male, all with non-shockable cardiac arrest. Sensitivity for cardiac arrest detection was 100% (10/10; 95% confidence interval [CI] 66–100%) and 90% (9/10; 95% CI 54–99%), in training 1 and 2, respectively. In the test set, sensitivity was 100% (23/23; 95% CI 82–100%), with one false positive alert. Cardiac arrest was detected at a mean arterial pressure of 30 mmHg (IQR 24–35) and pulse pressure of 13 mmHg (IQR 11–19). Interpretation Cardiac arrest can be detected with high sensitivity using wrist-derived photoplethysmography, providing first evidence on model performance in true cardiac arrest, specifically in non-shockable cases. Findings support further development of wearable-based cardiac arrest detection technologies to enable earlier recognition for unwitnessed cardiac arrest. Funding Dutch Heart Foundation, Radboudumc.

The Lancet Regional Health - EuropeVol. 67
Radboud University Nijmegen (NL), Erasmus MC (NL), University Medical Center (US), Radboud University Medical Center (NL), Reinier de Graaf Hospital (NL), Radboud Institute for Molecular Life Sciences (NL), Wilhelmina Children's Hospital (NL)
Hartstichting, Radboud Universitair Medisch Centrum
Openalex Percentile: Top 6%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.