Validity, tolerability, and feasibility of automated arterial occlusion pressure across body positions

Objective To evaluate the validity, safety, and feasibility of automated algorithm-based arterial occlusion pressure (AOP) measurement compared to pulsed wave Doppler ultrasound across supine, seated, and standing positions.Approach This two-phase sequential study enrolled healthy adults. In Phase One (n=46), AOP was measured using both automated and Doppler ultrasound methods at three body positions in randomized order. Validity was assessed using intraclass correlation coefficients (ICC), Bland-Altman analysis, and proportional bias testing. Safety was evaluated through measurement duration and pain perception. Feasibility was assessed by the first-attempt success rate. In Phase Two (n=15), a post hoc study examined whether averaging two automated measurements improved accuracy in supine position.Main results Automated algorithm-based AOP measurement demonstrated good agreement withDoppler ultrasound in supine (ICC=0.76), seated (ICC=0.87), and standing (ICC=0.83) positions. However, 95% limits of agreement ranged from -38 to +48 mmHg across positions. A significant negative proportional bias was observed in supine (p=0.031), indicating underestimation at higher AOP values. Automated measurements required significantly longer duration than Doppler across all positions (p<0.001). Pain perception was comparable between methods. First-attempt success rate was significantly lower for automated measurement in standing compared to Doppler method (87% vs. 100%, p=0.041). In Phase Two, 73% of participants demonstrated improved accuracy when averaging two automated measurements.Significance Automated algorithm-based arterial occlusion pressure (AOP) assessment demonstrates acceptable validity and safety when performed in seated and standing positions, supporting its use for individualized pressure prescription in these postures. However, clinicians and practitioners should exercise caution when determining AOP in the supine position, as proportional bias and reduced agreement with Doppler ultrasound may result in underestimation at higher AOP values. When measuring AOP in standing positions, providing external support (e.g., upper limb support on a stable surface) to reduce lower limb muscle activation may improve first-attempt success rates.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2026-07-24
DOI
https://doi.org/10.1088/1361-6579/ae9034
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity, tolerability, and feasibility of automated arterial occlusion pressure across body positions

J. H. Yang, Nicholas Rolnick, Taylor Gubler, Caden Cloutier et al.
Physiological Measurement
Cardiovascular and exercise physiology
article

Validity, tolerability, and feasibility of automated arterial occlusion pressure across body positions

J. H. Yang, Nicholas Rolnick, Taylor Gubler, Caden Cloutier, Drew Martin, Zachary Hamilton, Erin Duffy, Elizabeth Kenneally, Grace Harty, Benjamin Senra
article en

Abstract

Objective To evaluate the validity, safety, and feasibility of automated algorithm-based arterial occlusion pressure (AOP) measurement compared to pulsed wave Doppler ultrasound across supine, seated, and standing positions.Approach This two-phase sequential study enrolled healthy adults. In Phase One (n=46), AOP was measured using both automated and Doppler ultrasound methods at three body positions in randomized order. Validity was assessed using intraclass correlation coefficients (ICC), Bland-Altman analysis, and proportional bias testing. Safety was evaluated through measurement duration and pain perception. Feasibility was assessed by the first-attempt success rate. In Phase Two (n=15), a post hoc study examined whether averaging two automated measurements improved accuracy in supine position.Main results Automated algorithm-based AOP measurement demonstrated good agreement withDoppler ultrasound in supine (ICC=0.76), seated (ICC=0.87), and standing (ICC=0.83) positions. However, 95% limits of agreement ranged from -38 to +48 mmHg across positions. A significant negative proportional bias was observed in supine (p=0.031), indicating underestimation at higher AOP values. Automated measurements required significantly longer duration than Doppler across all positions (p<0.001). Pain perception was comparable between methods. First-attempt success rate was significantly lower for automated measurement in standing compared to Doppler method (87% vs. 100%, p=0.041). In Phase Two, 73% of participants demonstrated improved accuracy when averaging two automated measurements.Significance Automated algorithm-based arterial occlusion pressure (AOP) assessment demonstrates acceptable validity and safety when performed in seated and standing positions, supporting its use for individualized pressure prescription in these postures. However, clinicians and practitioners should exercise caution when determining AOP in the supine position, as proportional bias and reduced agreement with Doppler ultrasound may result in underestimation at higher AOP values. When measuring AOP in standing positions, providing external support (e.g., upper limb support on a stable surface) to reduce lower limb muscle activation may improve first-attempt success rates.

Physiological Measurement
Lehman College (US), University of Hartford (US)
National Aeronautics and Space Administration
Openalex Percentile: Top 5%
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.