Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study

Background/Objectives: The erythrocyte sedimentation rate (ESR) is a widely used biomarker of systemic inflammation, with established roles in rheumatologic, infectious, and malignant disease. The Westergren method remains the reference procedure, but automated analyzers are increasingly used clinically, and methodological differences may introduce systematic bias. This study evaluated analytical agreement between the Vision C automated ESR analyzer (EDTA) and the Westergren reference (sodium citrate). Methods: This prospective, single-center study followed CLSI EP09-ED3. A total of 228 consecutive adult patients were enrolled. Agreement was assessed by Bland–Altman analysis, Passing–Bablok regression, and Pearson correlation, with subgroup analyses (<40, 40–80, >80 mm/h) and against RCPA allowable total error (TEa) specifications. Results: Bland–Altman analysis showed a significant mean bias of −3.30 mm/h (95% CI: −4.25 to −2.35), with increasing dispersion at higher ESR values indicating heteroscedasticity; regression-based limits of agreement therefore widened progressively with ESR magnitude. Passing–Bablok regression showed no proportional bias but a significant constant bias; subgroup regression was not performed due to truncated ranges. Correlation was strong overall (r = 0.961, p < 0.001) and declined across subgroups (r = 0.821, 0.740, 0.687), consistent with range restriction rather than reduced agreement. The analyzer met RCPA TEa criteria across both ranges (≤20 mm/h: bias −2.55 mm/h within ±6.0 mm/h; >20 mm/h: percentage bias −10.49% within ±30%), including the high-ESR subgroup (approximately 6–9%). Conclusions: The Vision C analyzer showed a significant, level-dependent negative bias versus Westergren, widening at higher ESRs; strong correlation and RCPA TEa compliance across the full range, including the high subgroup, support its suitability for routine use. Wide regression-based limits and the limited high subgroup sample size (n = 20) warrant caution and further study but do not support routine manual confirmation, since predefined acceptance criteria are met. Method-specific reporting and consistent instrument use in follow-up are recommended.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-07-24
DOI
https://doi.org/10.3390/diagnostics16152323
Primary Topic
Inflammatory Biomarkers in Disease Prognosis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study

Yusuf Kurtulmuş, Kenan Yörük, A Tuzcu, Mustafa Yılmaz
Diagnostics
Inflammatory Biomarkers in Disease Prognosis
article

Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study

Yusuf Kurtulmuş, Kenan Yörük, A Tuzcu, Mustafa Yılmaz
article en

Abstract

Background/Objectives: The erythrocyte sedimentation rate (ESR) is a widely used biomarker of systemic inflammation, with established roles in rheumatologic, infectious, and malignant disease. The Westergren method remains the reference procedure, but automated analyzers are increasingly used clinically, and methodological differences may introduce systematic bias. This study evaluated analytical agreement between the Vision C automated ESR analyzer (EDTA) and the Westergren reference (sodium citrate). Methods: This prospective, single-center study followed CLSI EP09-ED3. A total of 228 consecutive adult patients were enrolled. Agreement was assessed by Bland–Altman analysis, Passing–Bablok regression, and Pearson correlation, with subgroup analyses (<40, 40–80, >80 mm/h) and against RCPA allowable total error (TEa) specifications. Results: Bland–Altman analysis showed a significant mean bias of −3.30 mm/h (95% CI: −4.25 to −2.35), with increasing dispersion at higher ESR values indicating heteroscedasticity; regression-based limits of agreement therefore widened progressively with ESR magnitude. Passing–Bablok regression showed no proportional bias but a significant constant bias; subgroup regression was not performed due to truncated ranges. Correlation was strong overall (r = 0.961, p < 0.001) and declined across subgroups (r = 0.821, 0.740, 0.687), consistent with range restriction rather than reduced agreement. The analyzer met RCPA TEa criteria across both ranges (≤20 mm/h: bias −2.55 mm/h within ±6.0 mm/h; >20 mm/h: percentage bias −10.49% within ±30%), including the high-ESR subgroup (approximately 6–9%). Conclusions: The Vision C analyzer showed a significant, level-dependent negative bias versus Westergren, widening at higher ESRs; strong correlation and RCPA TEa compliance across the full range, including the high subgroup, support its suitability for routine use. Wide regression-based limits and the limited high subgroup sample size (n = 20) warrant caution and further study but do not support routine manual confirmation, since predefined acceptance criteria are met. Method-specific reporting and consistent instrument use in follow-up are recommended.

DiagnosticsVol. 16(15)
Adnan Menderes University (TR)
Adnan Menderes Üniversitesi
Openalex Percentile: Top 11%
Inflammatory Biomarkers in Disease Prognosis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.