Diagnosis of tympanic effusions using hyperspectral imaging to prevent unnecessary paracentesis procedures

Abstract Background Otitis Media with Effusion (OME) is a prevalent childhood condition, affecting up to 90% of children with 10% progressing to a chronic state. It is commonly treated by tympanostomy tube placement. Diagnosis relies on clinical history, otoscopy, audiometry, and tympanometry, but these methods have a high false-positive rate of 15–28%. This highlights the need for an objective, reliable, and non-invasive diagnostic tool to reduce unnecessary surgeries. Methods To evaluate the rate of false-positive OME diagnoses in our tertiary hospital, a retrospective database analysis was conducted. To assess the feasibility of advanced imaging, experimental studies were performed ex vivo on human body donors. Sodium chloride solution was intratympanically injected under endoscopic guidance. Hyperspectral imaging (HSI) datasets were acquired using a customized endoscopic system to quantify tissue water content. Results An analysis of hospital records identified a false positive rate of 29% ( n = 4,564) for OME diagnosis, comparing the number of paracenteses to paracenteses plus tympanostomy tubes. In the ex vivo analysis of HSI, visual evaluation of the Tissue Water Index (TWI) demonstrated sensitivity, specificity, and accuracy of 89%, 78%, and 83%, respectively. The mean TWI of all evaluated tympanic membranes was 50.2 (± 7.1) before injection (control) and increased to 62.8 (± 14.8) after injection. Conclusion HSI presents a highly sensitive and user-friendly potential future diagnostic adjunct for diagnosing OME, which has shown its potential in an ex vivo approach. Now it needs in-vivo confirmation of the results as its implementation in clinical routine could contribute to avoiding unnecessary surgical interventions, minimizing patient burden and healthcare costs.

Authors

Institutions

Publication Details

Journal
BMC Medical Imaging
Published
2026-07-24
DOI
https://doi.org/10.1186/s12880-026-02611-w
Primary Topic
Ear Surgery and Otitis Media
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnosis of tympanic effusions using hyperspectral imaging to prevent unnecessary paracentesis procedures

Andreas Melzer, H. Köhler, Sabine Löffler, Markus Pirlich et al.
BMC Medical Imaging
Ear Surgery and Otitis Media
article

Diagnosis of tympanic effusions using hyperspectral imaging to prevent unnecessary paracentesis procedures

Andreas Melzer, H. Köhler, Sabine Löffler, Markus Pirlich, Denis Gholami Bajestani, M. Unger, Maximilian Gänzle, Andreas Dietz
article en

Abstract

Abstract Background Otitis Media with Effusion (OME) is a prevalent childhood condition, affecting up to 90% of children with 10% progressing to a chronic state. It is commonly treated by tympanostomy tube placement. Diagnosis relies on clinical history, otoscopy, audiometry, and tympanometry, but these methods have a high false-positive rate of 15–28%. This highlights the need for an objective, reliable, and non-invasive diagnostic tool to reduce unnecessary surgeries. Methods To evaluate the rate of false-positive OME diagnoses in our tertiary hospital, a retrospective database analysis was conducted. To assess the feasibility of advanced imaging, experimental studies were performed ex vivo on human body donors. Sodium chloride solution was intratympanically injected under endoscopic guidance. Hyperspectral imaging (HSI) datasets were acquired using a customized endoscopic system to quantify tissue water content. Results An analysis of hospital records identified a false positive rate of 29% ( n = 4,564) for OME diagnosis, comparing the number of paracenteses to paracenteses plus tympanostomy tubes. In the ex vivo analysis of HSI, visual evaluation of the Tissue Water Index (TWI) demonstrated sensitivity, specificity, and accuracy of 89%, 78%, and 83%, respectively. The mean TWI of all evaluated tympanic membranes was 50.2 (± 7.1) before injection (control) and increased to 62.8 (± 14.8) after injection. Conclusion HSI presents a highly sensitive and user-friendly potential future diagnostic adjunct for diagnosing OME, which has shown its potential in an ex vivo approach. Now it needs in-vivo confirmation of the results as its implementation in clinical routine could contribute to avoiding unnecessary surgical interventions, minimizing patient burden and healthcare costs.

BMC Medical Imaging
University Hospital Leipzig (DE), Leipzig University (DE)
Good health and well-being
Openalex Percentile: Top 7%
Ear Surgery and Otitis Media
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.