Shielding Performance and Clinical Applicability of Lead-Free Materials in Computed Tomography

Owing to the high radiation exposure associated with computed tomography (CT) examinations and the image quality degradation caused by conventional radiation shielding materials, this study evaluated the dose reduction performance and image quality maintenance potential of a newly developed lead-free composite shielding material. This material was composed of bismuth, tungsten, tungsten carbide, aluminum, and polyurethane. Phantom-based dose measurements demonstrated that the shielding material achieved dose reduction rates ranging from 17.6% to 37.6%, depending on tube voltage. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and changes in tube current-time product (mAs) under a scout-based automatic exposure control (AEC) protocol were analyzed according to the presence or absence of the shielding material across regions. For the clinical evaluation, CT scans were performed on four patients. Furthermore, the images were reviewed to evaluate whether this material affected image quality. The shielding material exhibited radiation reduction levels comparable to those reported in previous studies. SNR and CNR analyses showed minor statistical variations in certain regions; however, most differences were not statistically significant, and even significant differences remained within a range that did not compromise diagnostic image quality. Under the scout-based AEC protocol, the use of the shielding material resulted in less than 1% variation in mAs values. No visually perceptible artifacts or clinically significant image quality degradation were observed. The proposed composite shielding material demonstrated the potential to mitigate some limitations of conventional shielding materials and showed preliminary clinical feasibility as an adjunctive strategy for radiation dose reduction in CT examinations.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2026-07-23
DOI
https://doi.org/10.1088/1361-6498/ae8f51
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Shielding Performance and Clinical Applicability of Lead-Free Materials in Computed Tomography

Young Jin Won, Won Il Jang, Kum Bae Kim, Vannyat Ath et al.
Journal of Radiological Protection
Radiation Dose and Imaging
article

Shielding Performance and Clinical Applicability of Lead-Free Materials in Computed Tomography

Young Jin Won, Won Il Jang, Kum Bae Kim, Vannyat Ath, Jung Ju Jo, Sang Hyoun Choi
article en

Abstract

Owing to the high radiation exposure associated with computed tomography (CT) examinations and the image quality degradation caused by conventional radiation shielding materials, this study evaluated the dose reduction performance and image quality maintenance potential of a newly developed lead-free composite shielding material. This material was composed of bismuth, tungsten, tungsten carbide, aluminum, and polyurethane. Phantom-based dose measurements demonstrated that the shielding material achieved dose reduction rates ranging from 17.6% to 37.6%, depending on tube voltage. Signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and changes in tube current-time product (mAs) under a scout-based automatic exposure control (AEC) protocol were analyzed according to the presence or absence of the shielding material across regions. For the clinical evaluation, CT scans were performed on four patients. Furthermore, the images were reviewed to evaluate whether this material affected image quality. The shielding material exhibited radiation reduction levels comparable to those reported in previous studies. SNR and CNR analyses showed minor statistical variations in certain regions; however, most differences were not statistically significant, and even significant differences remained within a range that did not compromise diagnostic image quality. Under the scout-based AEC protocol, the use of the shielding material resulted in less than 1% variation in mAs values. No visually perceptible artifacts or clinically significant image quality degradation were observed. The proposed composite shielding material demonstrated the potential to mitigate some limitations of conventional shielding materials and showed preliminary clinical feasibility as an adjunctive strategy for radiation dose reduction in CT examinations.

Journal of Radiological Protection
University of Science and Technology (YE), Eulji University (KR), Eulji University Uijeongbu Eulji Medical Center (KR)
National Research Council of Science and Technology
Openalex Percentile: Top 9%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.