Dosimetry evaluation of an MRI guided small animal irradiator

OBJECTIVE: Magnetic Resonance guided Radiation Therapy (MRgRT) integrates magnetic resonance imaging (MRI) with radiation delivery, providing superior soft-tissue contrast and real-time treatment guidance. For preclinical MRgRT irradiators, a critical concern is whether the ambient magnetic fields compromise the radiation dose distribution. This study aims to investigate the influence of magnetic fields on X-ray tube electron trajectories and dosimetric outcomes. APPROACH: The system integrates a 0.5 T MRI scanner and a 225 kVp X-ray tube. The localized magnetic field distribution was characterized via finite element analysis. Electron trajectories were simulated across various magnetic field strengths and alignment angles to quantify the X-ray focal-spot displacement. Monte Carlo simulation was then performed to evaluate the dose distribution within homogeneous and heterogeneous tissue-equivalent phantoms. MAIN RESULTS: In the proposed system, the fringe magnetic field across the X-ray tube electrodes measures 40.4 G and aligns nearly parallel to the electric field. This configuration restricts the X-ray focal-spot displacement to 0.15 mm, resulting in negligible dosimetric deviations. There is good agreement between the dose distributions in the phantoms with and without magnetic fields, maintaining ≧99.9% gamma passing rates at 0.5 mm / 1% criteria even under a main magnetic field of up to 10 T. SIGNIFICANCE: This study confirms that, for the small animal MRgRT system, main magnetic field as low as 0.5 T exerts no significant dosimetric impact, provided the magnet and the X-ray tube are well separated, and it may be unnecessary to perform dosimetric corrections.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-07-23
DOI
https://doi.org/10.1088/1361-6560/ae8fb2
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dosimetry evaluation of an MRI guided small animal irradiator

Yi Yang, Yaowen Cao, Cheng Zheng, Xiaogang Yuan et al.
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Dosimetry evaluation of an MRI guided small animal irradiator

Yi Yang, Yaowen Cao, Cheng Zheng, Xiaogang Yuan, Ning Zhao
article en

Abstract

OBJECTIVE: Magnetic Resonance guided Radiation Therapy (MRgRT) integrates magnetic resonance imaging (MRI) with radiation delivery, providing superior soft-tissue contrast and real-time treatment guidance. For preclinical MRgRT irradiators, a critical concern is whether the ambient magnetic fields compromise the radiation dose distribution. This study aims to investigate the influence of magnetic fields on X-ray tube electron trajectories and dosimetric outcomes. APPROACH: The system integrates a 0.5 T MRI scanner and a 225 kVp X-ray tube. The localized magnetic field distribution was characterized via finite element analysis. Electron trajectories were simulated across various magnetic field strengths and alignment angles to quantify the X-ray focal-spot displacement. Monte Carlo simulation was then performed to evaluate the dose distribution within homogeneous and heterogeneous tissue-equivalent phantoms. MAIN RESULTS: In the proposed system, the fringe magnetic field across the X-ray tube electrodes measures 40.4 G and aligns nearly parallel to the electric field. This configuration restricts the X-ray focal-spot displacement to 0.15 mm, resulting in negligible dosimetric deviations. There is good agreement between the dose distributions in the phantoms with and without magnetic fields, maintaining ≧99.9% gamma passing rates at 0.5 mm / 1% criteria even under a main magnetic field of up to 10 T. SIGNIFICANCE: This study confirms that, for the small animal MRgRT system, main magnetic field as low as 0.5 T exerts no significant dosimetric impact, provided the magnet and the X-ray tube are well separated, and it may be unnecessary to perform dosimetric corrections.

Physics in Medicine and Biology
University of Science and Technology of China (CN), MD Precision (Canada) (CA)
Fundamental Research Funds for the Central Universities
Openalex Percentile: Top 9%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.