A projection‐domain deep learning approach for respiratory motion correction of myocardial perfusion imaging using a multi‐pinhole solid‐state SPECT

BACKGROUND: Myocardial perfusion imagining (MPI) is a nuclear medicine technique used in the assessment of coronary artery disease. Differences in the perfusion of the myocardium at rest and stress and regions of perfusion defects can be indicators of various cardiac disease states. Respiratory motion (RM) can result in the degradation of image quality and the appearance of artificial regions of perfusion deficit. While RM correction is possible, many methods involve time-consuming calculation or additional equipment for motion tracking. This has limited the clinical uptake of such methods. This work presents a deep learning model for data-driven RM correction on dedicated cardiac SPECT scanners. PURPOSE: An AI-based method for estimating RM from gated projection data for MPI SPECT using a dedicated pinhole scanner was developed. Accuracy of these estimates and impact on image characteristics were assessed. METHODS: RM motion parameters were calculated for rest and stress MPI scans using a minimized root mean squared (RMS) alignment of reconstructed respiratory gates for 90 patients supplemented with 33 NCAT simulated patients. Using these parameters, an AI network was trained to estimate motion based on gated projection data prior to reconstruction. Accuracy of AI predictions relative to RMS data were assessed for clinical scans and down-sampled data to simulate noisy acquisitions. Scans were reconstructed with AI motion correction (MC), RMS MC, and no MC. Anterior myocardial wall thickness and its contrast relative to the interior of the ventricle were calculated in each case. RESULTS: AI motion predictions were accurate to within 1.86 ± 0.05 mm at clinical noise levels, increasing to 2.07 ± 0.07 mm at 1/8 signal. MC with AI was shown to provide a statistically significant increase anterior wall contrast and decrease measured wall thickness relative to no MC, with comparable results to RMS MC. CONCLUSION: Direct AI-based MC parameter estimation using gated projection data was shown to be feasible in clinical settings. Corrections based on AI estimates showed improved contrast and reduced myocardial wall thickness when incorporated into image reconstruction.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-07-22
DOI
https://doi.org/10.1002/mp.70543
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A projection‐domain deep learning approach for respiratory motion correction of myocardial perfusion imaging using a multi‐pinhole solid‐state SPECT

Terrence David Ruddy, R Glenn Wells, Dylan J Malenfant
Medical Physics
Cardiac Imaging and Diagnostics
article

A projection‐domain deep learning approach for respiratory motion correction of myocardial perfusion imaging using a multi‐pinhole solid‐state SPECT

Terrence David Ruddy, R Glenn Wells, Dylan J Malenfant
article en

Abstract

BACKGROUND: Myocardial perfusion imagining (MPI) is a nuclear medicine technique used in the assessment of coronary artery disease. Differences in the perfusion of the myocardium at rest and stress and regions of perfusion defects can be indicators of various cardiac disease states. Respiratory motion (RM) can result in the degradation of image quality and the appearance of artificial regions of perfusion deficit. While RM correction is possible, many methods involve time-consuming calculation or additional equipment for motion tracking. This has limited the clinical uptake of such methods. This work presents a deep learning model for data-driven RM correction on dedicated cardiac SPECT scanners. PURPOSE: An AI-based method for estimating RM from gated projection data for MPI SPECT using a dedicated pinhole scanner was developed. Accuracy of these estimates and impact on image characteristics were assessed. METHODS: RM motion parameters were calculated for rest and stress MPI scans using a minimized root mean squared (RMS) alignment of reconstructed respiratory gates for 90 patients supplemented with 33 NCAT simulated patients. Using these parameters, an AI network was trained to estimate motion based on gated projection data prior to reconstruction. Accuracy of AI predictions relative to RMS data were assessed for clinical scans and down-sampled data to simulate noisy acquisitions. Scans were reconstructed with AI motion correction (MC), RMS MC, and no MC. Anterior myocardial wall thickness and its contrast relative to the interior of the ventricle were calculated in each case. RESULTS: AI motion predictions were accurate to within 1.86 ± 0.05 mm at clinical noise levels, increasing to 2.07 ± 0.07 mm at 1/8 signal. MC with AI was shown to provide a statistically significant increase anterior wall contrast and decrease measured wall thickness relative to no MC, with comparable results to RMS MC. CONCLUSION: Direct AI-based MC parameter estimation using gated projection data was shown to be feasible in clinical settings. Corrections based on AI estimates showed improved contrast and reduced myocardial wall thickness when incorporated into image reconstruction.

Medical PhysicsVol. 53(8)
University of Ottawa (CA), Carleton University (CA)
Good health and well-being
Openalex Percentile: Top 8%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.