Coronary plaque quantification on energy-integrating and photon-counting detector CT: reproducibility and power modeling for mixed-platform trials

Abstract Objectives To evaluate the reproducibility of quantitative plaque characterization between energy-integrating detector (EID)-CT and photon-counting detector (PCD)-CT, identify reconstruction settings yielding the lowest variability, and model sample-size requirements for trial planning. Materials and methods Patients who underwent coronary CT angiography on dual-source EID-CT and PCD-CT within 30 days were screened retrospectively. EID-CT data were reconstructed using quantitative (Qr40) and vascular (Bv40) kernels, while PCD-CT data were reconstructed with Bv36/40/44 and Qr36/40/44 kernels and quantum iterative reconstruction strengths 2–4. For each plaque component (total, low-attenuation, fibrotic, and calcified), volumes were computed using fixed and adaptive Hounsfield-unit thresholds using an automated deep-learning–based plaque-quantification platform and spatially co-registered. Inter-scanner standard deviation (SD) was calculated, and optimal reconstruction pairs were used for power modeling. Results Thirty-eight patients (age 68.0 [64.0–72.5] years, 30 men) and 77 vessels were included. Inter-scanner correlations were very strong for total ( r = 0.85–0.95), fibrotic ( r = 0.74–0.94), and calcified plaque ( r = 0.92–0.98), and strong for low-attenuation plaque volumes ( r = 0.71–0.85). Mean bias ranged from 8.2 to 164.4 mm³ for total plaque volume. Applying the optimal reconstruction pair (Qr40 EID-CT vs. Bv36 PCD-CT ), inter-scanner SDs were 0.14 (vessel-based) and 0.12 (patient-based). At 80% power and α = 0.05, the estimated sample sizes to detect 5% and 10% changes in total plaque volume were 116 and 29 vessels or 92 and 23 patients per group. Conclusion Quantitative coronary plaque volumes demonstrated high inter-scanner consistency between EID-CT and PCD-CT under optimized conditions. The dual-level power-modeling framework translates inter-scanner variability into actionable sample-size estimates for study design. Key Points Question Quantitative coronary plaque volumes are increasingly used as surrogate biomarkers, yet variability between energy-integrating and photon-counting CT remains a major concern. Findings Harmonized reconstruction settings achieved high inter-scanner consistency with low variability when optimized reconstruction settings and automated plaque quantification are applied. Clinical relevance Standardized coronary plaque quantification across CT technologies enables reliable use of quantitative plaque endpoints, supporting mixed-platform longitudinal trials.

Authors

Publication Details

Journal
European Radiology
Published
2026-07-22
DOI
https://doi.org/10.1007/s00330-026-12773-4
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Coronary plaque quantification on energy-integrating and photon-counting detector CT: reproducibility and power modeling for mixed-platform trials

Mehmet Gulsun, Ákos Varga‐Szemes, Muhammad Taha Hagar, Tilman Emrich et al.
European Radiology
Advanced X-ray and CT Imaging
article

Coronary plaque quantification on energy-integrating and photon-counting detector CT: reproducibility and power modeling for mixed-platform trials

Mehmet Gulsun, Ákos Varga‐Szemes, Muhammad Taha Hagar, Tilman Emrich, Sardi Hyska, José Osoria-Velasquez, Bálint Szilveszter, Pál Maurovich-Horvat, Milan Vecsey-Nagy, Borbála Vattay, Puneet Sharma
article en

Abstract

Abstract Objectives To evaluate the reproducibility of quantitative plaque characterization between energy-integrating detector (EID)-CT and photon-counting detector (PCD)-CT, identify reconstruction settings yielding the lowest variability, and model sample-size requirements for trial planning. Materials and methods Patients who underwent coronary CT angiography on dual-source EID-CT and PCD-CT within 30 days were screened retrospectively. EID-CT data were reconstructed using quantitative (Qr40) and vascular (Bv40) kernels, while PCD-CT data were reconstructed with Bv36/40/44 and Qr36/40/44 kernels and quantum iterative reconstruction strengths 2–4. For each plaque component (total, low-attenuation, fibrotic, and calcified), volumes were computed using fixed and adaptive Hounsfield-unit thresholds using an automated deep-learning–based plaque-quantification platform and spatially co-registered. Inter-scanner standard deviation (SD) was calculated, and optimal reconstruction pairs were used for power modeling. Results Thirty-eight patients (age 68.0 [64.0–72.5] years, 30 men) and 77 vessels were included. Inter-scanner correlations were very strong for total ( r = 0.85–0.95), fibrotic ( r = 0.74–0.94), and calcified plaque ( r = 0.92–0.98), and strong for low-attenuation plaque volumes ( r = 0.71–0.85). Mean bias ranged from 8.2 to 164.4 mm³ for total plaque volume. Applying the optimal reconstruction pair (Qr40 EID-CT vs. Bv36 PCD-CT ), inter-scanner SDs were 0.14 (vessel-based) and 0.12 (patient-based). At 80% power and α = 0.05, the estimated sample sizes to detect 5% and 10% changes in total plaque volume were 116 and 29 vessels or 92 and 23 patients per group. Conclusion Quantitative coronary plaque volumes demonstrated high inter-scanner consistency between EID-CT and PCD-CT under optimized conditions. The dual-level power-modeling framework translates inter-scanner variability into actionable sample-size estimates for study design. Key Points Question Quantitative coronary plaque volumes are increasingly used as surrogate biomarkers, yet variability between energy-integrating and photon-counting CT remains a major concern. Findings Harmonized reconstruction settings achieved high inter-scanner consistency with low variability when optimized reconstruction settings and automated plaque quantification are applied. Clinical relevance Standardized coronary plaque quantification across CT technologies enables reliable use of quantitative plaque endpoints, supporting mixed-platform longitudinal trials.

European Radiology
Affordable and clean energy
Openalex Percentile: Top 16%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.