Robotic-arm assistance in revision total hip arthroplasty improves the accuracy of acetabular component positioning: a cadaver-based study

INTRODUCTION: Acetabular component positioning is a key factor for stability, impingement, wear, and biomechanics in revision total hip arthroplasty (THA). However, anatomic distortion and previous implants may complicate positioning. Although evidence on computed-tomography (CT)-based robotic assistance in revision THA is limited, its planning features and precision may improve surgical plan execution. This study aimed to measure the precision and accuracy of acetabular component positioning of revision robotic-arm assisted total hip arthroplasty (RA-THA) compared with a published manual primary total hip arthroplasty (M-THA) benchmark for: (a) acetabular inclination; (b) acetabular version; and (c) center of rotation (COR), with comparison by surgeon case volume. MATERIALS AND METHODS: There were seven cadavers (nine hips) that previously underwent primary THA revised with RA-THA by independent surgeons with varying primary RA-THA case volumes, surgical approaches, and acetabular component types. Due to surgical protocol deviations, two hips were excluded leaving seven hips from six cadavers. Pre- and postoperative CT scans were compared to calculate absolute error for inclination, version, and COR; normality was established with the Anderson-Darling test and two-variance tests assessed differences in standard deviations (α = 0.05). Mean differences between cadaver-based revision RA-THA and published M-THA values were compared with Welch's t-test for superiority (α = 0.05). Effect sizes were reported as mean differences with 95% upper bounds and sensitivity analyses were performed given the fixed sample size. RESULTS: Revision RA-THA demonstrated significantly greater accuracy for acetabular inclination (2.5 ± 1.9° versus 7.2 ± 3.2°, P = 0.008) and version (2.2 ± 1.6° versus 7.8 ± 4.6°, P = 0.014), and significantly greater precision for acetabular version (P = 0.023) compared with M-THA. There was adirectionally lower error for COR compared with M-THA. Mean error reductions were 5.7° for version and 4.7º for inclination with 95% upper bounds of -1.9 for both version and inclination. Low RA-THA surgical volume corresponded with the highest absolute error for shell inclination (5.5º) and version (5.0º), though values remained below published M-THA averages. CONCLUSIONS: Revision RA-THA reduced error in both acetabular inclination and acetabular version, and variability in acetabular version relative to the M-THA benchmark. Although these results relate only to technical performance, accurate acetabular positioning in clinical practice is an important factor in reducing the risk of malposition-related complications in revision THA.

Authors

Institutions

Publication Details

Journal
Archives of Orthopaedic and Trauma Surgery
Published
2026-07-18
DOI
https://doi.org/10.1007/s00402-026-06395-0
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Robotic-arm assistance in revision total hip arthroplasty improves the accuracy of acetabular component positioning: a cadaver-based study

Ilya Borukhov, Sally LiArno, Chase Smitterberg, Varun Chandra et al.
Archives of Orthopaedic and Trauma Surgery
Orthopaedic implants and arthroplasty
article

Robotic-arm assistance in revision total hip arthroplasty improves the accuracy of acetabular component positioning: a cadaver-based study

Ilya Borukhov, Sally LiArno, Chase Smitterberg, Varun Chandra, Kailey Fitzgerald, Michael A. Mont
article en

Abstract

INTRODUCTION: Acetabular component positioning is a key factor for stability, impingement, wear, and biomechanics in revision total hip arthroplasty (THA). However, anatomic distortion and previous implants may complicate positioning. Although evidence on computed-tomography (CT)-based robotic assistance in revision THA is limited, its planning features and precision may improve surgical plan execution. This study aimed to measure the precision and accuracy of acetabular component positioning of revision robotic-arm assisted total hip arthroplasty (RA-THA) compared with a published manual primary total hip arthroplasty (M-THA) benchmark for: (a) acetabular inclination; (b) acetabular version; and (c) center of rotation (COR), with comparison by surgeon case volume. MATERIALS AND METHODS: There were seven cadavers (nine hips) that previously underwent primary THA revised with RA-THA by independent surgeons with varying primary RA-THA case volumes, surgical approaches, and acetabular component types. Due to surgical protocol deviations, two hips were excluded leaving seven hips from six cadavers. Pre- and postoperative CT scans were compared to calculate absolute error for inclination, version, and COR; normality was established with the Anderson-Darling test and two-variance tests assessed differences in standard deviations (α = 0.05). Mean differences between cadaver-based revision RA-THA and published M-THA values were compared with Welch's t-test for superiority (α = 0.05). Effect sizes were reported as mean differences with 95% upper bounds and sensitivity analyses were performed given the fixed sample size. RESULTS: Revision RA-THA demonstrated significantly greater accuracy for acetabular inclination (2.5 ± 1.9° versus 7.2 ± 3.2°, P = 0.008) and version (2.2 ± 1.6° versus 7.8 ± 4.6°, P = 0.014), and significantly greater precision for acetabular version (P = 0.023) compared with M-THA. There was adirectionally lower error for COR compared with M-THA. Mean error reductions were 5.7° for version and 4.7º for inclination with 95% upper bounds of -1.9 for both version and inclination. Low RA-THA surgical volume corresponded with the highest absolute error for shell inclination (5.5º) and version (5.0º), though values remained below published M-THA averages. CONCLUSIONS: Revision RA-THA reduced error in both acetabular inclination and acetabular version, and variability in acetabular version relative to the M-THA benchmark. Although these results relate only to technical performance, accurate acetabular positioning in clinical practice is an important factor in reducing the risk of malposition-related complications in revision THA.

Archives of Orthopaedic and Trauma SurgeryVol. 146(1)
Sinai Hospital (US), Stryker (United States) (US)
Stryker
Openalex Percentile: Top 6%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.