Enhanced 3D-GRID radiotherapy on helical tomotherapy via personalized geometric optimization: A comparative study with IMRT and VMAT

PURPOSE: This study developed and evaluated a personalized cylindrical 3D-GRID radiotherapy strategy adapted to helical tomotherapy (TOMO) for large-volume tumors, and compared it with intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). METHODS: ) were retrospectively included. Anatomy-constrained cylindrical GRID volumes were generated within the GTV with predefined OAR clearance. Paired IMRT, VMAT, and TOMO plans were created using identical GTV and GRID-volume prescription objectives, and dosimetric, integral-dose, OAR-dose, and delivery-efficiency metrics were compared using the same patient-specific GRID geometry for all patients. RESULTS: GTV lower-dose coverage metrics (D90 %, D95 %, and D98 %) did not differ significantly among techniques (all p > 0.05). For IMRT, VMAT, and TOMO, respectively, PVDR1 was 1.953 ± 0.159, 1.983 ± 0.164, and 2.229 ± 0.128, and PVDR2 was 2.466 ± 0.166, 2.474 ± 0.187, and 2.752 ± 0.144 (both omnibus p < 0.001). The higher PVDR values with TOMO were mainly driven by higher D5 % and D10 %, not lower D90 % or D95 %. TOMO also showed higher gradient index and Body-GTV integral dose, indicating less favorable dose fall-off and broader dose spread, whereas VMAT required the fewest MUs and shortest beam-on time. No evaluated OAR endpoint differed significantly among techniques or exceeded its corresponding five-fraction constraint. CONCLUSION: Personalized cylindrical 3D-GRID planning is dosimetrically feasible on TOMO and may serve as a platform-adapted option for TOMO-only or TOMO-dominant centers. Higher DVH-derived PVDR or AVF values should be interpreted with dose fall-off, integral dose, OAR dose, and delivery efficiency rather than as evidence of TOMO superiority.

Authors

Institutions

Publication Details

Journal
Physica Medica
Published
2026-07-17
DOI
https://doi.org/10.1016/j.ejmp.2026.105900
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhanced 3D-GRID radiotherapy on helical tomotherapy via personalized geometric optimization: A comparative study with IMRT and VMAT

尹辉 Yin Hui, Zhitao Dai, Chuangao Wang, J S Li et al.
Physica Medica
Advanced Radiotherapy Techniques
article

Enhanced 3D-GRID radiotherapy on helical tomotherapy via personalized geometric optimization: A comparative study with IMRT and VMAT

尹辉 Yin Hui, Zhitao Dai, Chuangao Wang, J S Li, Wu Weiwei, Liu Zhiwei
article en

Abstract

PURPOSE: This study developed and evaluated a personalized cylindrical 3D-GRID radiotherapy strategy adapted to helical tomotherapy (TOMO) for large-volume tumors, and compared it with intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). METHODS: ) were retrospectively included. Anatomy-constrained cylindrical GRID volumes were generated within the GTV with predefined OAR clearance. Paired IMRT, VMAT, and TOMO plans were created using identical GTV and GRID-volume prescription objectives, and dosimetric, integral-dose, OAR-dose, and delivery-efficiency metrics were compared using the same patient-specific GRID geometry for all patients. RESULTS: GTV lower-dose coverage metrics (D90 %, D95 %, and D98 %) did not differ significantly among techniques (all p > 0.05). For IMRT, VMAT, and TOMO, respectively, PVDR1 was 1.953 ± 0.159, 1.983 ± 0.164, and 2.229 ± 0.128, and PVDR2 was 2.466 ± 0.166, 2.474 ± 0.187, and 2.752 ± 0.144 (both omnibus p < 0.001). The higher PVDR values with TOMO were mainly driven by higher D5 % and D10 %, not lower D90 % or D95 %. TOMO also showed higher gradient index and Body-GTV integral dose, indicating less favorable dose fall-off and broader dose spread, whereas VMAT required the fewest MUs and shortest beam-on time. No evaluated OAR endpoint differed significantly among techniques or exceeded its corresponding five-fraction constraint. CONCLUSION: Personalized cylindrical 3D-GRID planning is dosimetrically feasible on TOMO and may serve as a platform-adapted option for TOMO-only or TOMO-dominant centers. Higher DVH-derived PVDR or AVF values should be interpreted with dose fall-off, integral dose, OAR dose, and delivery efficiency rather than as evidence of TOMO superiority.

Physica MedicaVol. 148
Chinese Academy of Medical Sciences & Peking Union Medical College (CN), National Cancer Center (US), Ganzhou People's Hospital (CN), China Institute of Atomic Energy (CN)
Sanming Project of Medicine in Shenzhen
Openalex Percentile: Top 8%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.