Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign

Objectives: The aim of this study was to compare insertion torque (IT) and primary stability, assessed by resonance frequency analysis (RFA), between two implant systems with a shared macrodesign but different surface morphology and titanium grade in a standardized ex vivo porcine bone model. Methods: Sixty implants (OmniTaper EV and Xive; n = 30 per group) were inserted into standardized porcine rib bone blocks using identical osteotomy protocols. Peak insertion torque and primary stability, assessed by resonance frequency analysis (RFA), were recorded immediately after placement. Between-group differences were evaluated using appropriate statistical methods. Results: Bone density did not differ significantly between groups. OmniTaper EV implants demonstrated significantly higher insertion torque (28.2 ± 6.5 Ncm vs. 24.2 ± 6.2 Ncm; p = 0.020) and higher ISQ values (74.1 ± 6.0 vs. 69.8 ± 6.9; p = 0.013) than Xive implants. Mixed-effects analyses confirmed both findings after adjustment for clustering. No significant correlation was found between insertion torque and ISQ. Conclusions: Despite sharing a common implant design concept and standardized osteotomy preparation, the investigated implant systems demonstrated significant differences in insertion torque and primary stability. The findings indicate that implant-specific characteristics, including surface morphology and titanium grade, may influence implant–bone interface mechanics during insertion. However, their individual contributions cannot be distinguished within the present study design and should therefore be interpreted as the combined influence of implant-specific characteristics.

Authors

Institutions

Publication Details

Journal
Dentistry Journal
Published
2026-07-16
DOI
https://doi.org/10.3390/dj14070446
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign

Peter Gehrke, Günter Dhom, Maria Julia Pietruska, Robert Sader et al.
Dentistry Journal
Dental Implant Techniques and Outcomes
article

Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign

Peter Gehrke, Günter Dhom, Maria Julia Pietruska, Robert Sader, Paul Weigl, Philipp Klose, Jörg Neugebauer, Octavio Weinhold
article en

Abstract

Objectives: The aim of this study was to compare insertion torque (IT) and primary stability, assessed by resonance frequency analysis (RFA), between two implant systems with a shared macrodesign but different surface morphology and titanium grade in a standardized ex vivo porcine bone model. Methods: Sixty implants (OmniTaper EV and Xive; n = 30 per group) were inserted into standardized porcine rib bone blocks using identical osteotomy protocols. Peak insertion torque and primary stability, assessed by resonance frequency analysis (RFA), were recorded immediately after placement. Between-group differences were evaluated using appropriate statistical methods. Results: Bone density did not differ significantly between groups. OmniTaper EV implants demonstrated significantly higher insertion torque (28.2 ± 6.5 Ncm vs. 24.2 ± 6.2 Ncm; p = 0.020) and higher ISQ values (74.1 ± 6.0 vs. 69.8 ± 6.9; p = 0.013) than Xive implants. Mixed-effects analyses confirmed both findings after adjustment for clustering. No significant correlation was found between insertion torque and ISQ. Conclusions: Despite sharing a common implant design concept and standardized osteotomy preparation, the investigated implant systems demonstrated significant differences in insertion torque and primary stability. The findings indicate that implant-specific characteristics, including surface morphology and titanium grade, may influence implant–bone interface mechanics during insertion. However, their individual contributions cannot be distinguished within the present study design and should therefore be interpreted as the combined influence of implant-specific characteristics.

Dentistry JournalVol. 14(7)
Goethe University Frankfurt (DE), University of Cologne (DE), Dornier (Germany) (DE), University Hospital Frankfurt (DE), University Hospital Cologne (DE), Steinbeis University Berlin (DE)
Openalex Percentile: Top 6%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.