Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems

Objectives: To evaluate the influence of implant–abutment connection design, implant diameter, and simulated marginal bone loss on fracture resistance and failure patterns of dental implant systems. Materials and Methods: A total of 180 implant–abutment assemblies were tested, including Morse taper (MT) and internal hex (IH) connections with diameters of 3.5, 4.0, and 5.0 mm. Implants were embedded at two simulated bone levels (0 and 3 mm) and loaded at 30° until failure, in accordance with ISO 14801:2015. Fracture resistance (N) was analyzed using three-way ANOVA. Results: Connection type, implant diameter, bone level, and their interactions significantly affected fracture resistance (p < 0.001). Internal hex implants showed a marked diameter-dependent increase in resistance and greater reduction under simulated bone loss, particularly in reduced diameters. In contrast, Morse taper implants demonstrated similar resistance values regardless of implant diameter or bone level. Failure patterns also differed between systems: internal hex implants exhibited cervical implant fractures in reduced diameters, whereas Morse taper implants showed progressive abutment deformation and/or abutment fracture without implant body fracture. Conclusions: Implant fracture resistance is strongly influenced by implant–abutment connection geometry. Within the limitations of this static in vitro study, MT systems demonstrated diameter-independent mechanical stability and prosthetic-controlled failure patterns, whereas IH systems were highly sensitive to diameter reduction and simulated bone loss.

Authors

Institutions

Publication Details

Journal
Journal of Functional Biomaterials
Published
2026-07-17
DOI
https://doi.org/10.3390/jfb17070346
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems

Márcio de Carvalho Formiga, Sérgio Alexandre Gehrke, Juliana Campos Hasse Fernandes, Fátima de Campos Buzzi et al.
Journal of Functional Biomaterials
Dental Implant Techniques and Outcomes
article

Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems

Márcio de Carvalho Formiga, Sérgio Alexandre Gehrke, Juliana Campos Hasse Fernandes, Fátima de Campos Buzzi, Gustavo Vicentis Oliveira Fernandes, Antônio Scarano, Gustavo Coura, Bruno Freitas Mello
article en

Abstract

Objectives: To evaluate the influence of implant–abutment connection design, implant diameter, and simulated marginal bone loss on fracture resistance and failure patterns of dental implant systems. Materials and Methods: A total of 180 implant–abutment assemblies were tested, including Morse taper (MT) and internal hex (IH) connections with diameters of 3.5, 4.0, and 5.0 mm. Implants were embedded at two simulated bone levels (0 and 3 mm) and loaded at 30° until failure, in accordance with ISO 14801:2015. Fracture resistance (N) was analyzed using three-way ANOVA. Results: Connection type, implant diameter, bone level, and their interactions significantly affected fracture resistance (p < 0.001). Internal hex implants showed a marked diameter-dependent increase in resistance and greater reduction under simulated bone loss, particularly in reduced diameters. In contrast, Morse taper implants demonstrated similar resistance values regardless of implant diameter or bone level. Failure patterns also differed between systems: internal hex implants exhibited cervical implant fractures in reduced diameters, whereas Morse taper implants showed progressive abutment deformation and/or abutment fracture without implant body fracture. Conclusions: Implant fracture resistance is strongly influenced by implant–abutment connection geometry. Within the limitations of this static in vitro study, MT systems demonstrated diameter-independent mechanical stability and prosthetic-controlled failure patterns, whereas IH systems were highly sensitive to diameter reduction and simulated bone loss.

Journal of Functional BiomaterialsVol. 17(7)
Universidade Católica Portuguesa (PT), University of Chieti-Pescara (IT), Universidad Católica del Uruguay (UY), Missouri Foundation for Health (US), Universidade do Vale do Itajaí (BR), Northeast Ohio Medical University (US)
Openalex Percentile: Top 6%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.