Real-Time Comparison of Machine Learning-Enabled Devices for Measuring Compensatory Reserve Status

In this work, the real-time measurement of compensatory reserve metrics is compared during a simulated central hypovolemia clinical research study. We hypothesized that values generated from the compensatory reserve measurement (CRM) device would be statistically similar to those values generated from the compensatory reserve index (CRI) device. A study was conducted to collect data from noninvasive photoplethysmography sensors to validate the CRM against an FDA-cleared CRI device. Participants were placed in a lower body negative pressure (LBNP) chamber and underwent exposure to a stepwise protocol of increasing negative pressure. Data from both algorithmic devices were evaluated using multiple approaches: median error, median absolute error, and pooled Pearson correlations. Data were compared from 20 participants through the duration of the experiment. Pooled correlation analysis between CRM and CRI for all subjects and LBNP steps achieved R2 = 0.859, showing strong linear agreement. Bland–Altman analysis revealed a near-zero mean bias between CRM and CRI (0.13%). Each metric allowed for a mean early measurement time of 20.52 min for CRM and 17.92 min for CRI. Consistent with our hypothesis, the CRM device performed similarly to CRI, demonstrating that both models can track compensatory status changes with progressive central hypovolemia. The early, real-time predictive tracking of compensatory changes can be valuable for continuous monitoring and triaging in prehospital trauma environments.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-07-16
DOI
https://doi.org/10.3390/bioengineering13070817
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real-Time Comparison of Machine Learning-Enabled Devices for Measuring Compensatory Reserve Status

Sofía I. Hernández Torres, Jose M. Gonzalez, Eric J. Snider, Víctor A. Convertino et al.
Bioengineering
Non-Invasive Vital Sign Monitoring
article

Real-Time Comparison of Machine Learning-Enabled Devices for Measuring Compensatory Reserve Status

Sofía I. Hernández Torres, Jose M. Gonzalez, Eric J. Snider, Víctor A. Convertino, Carlos Bedolla, Krysta Amezcua, Ryan Ortiz
article en

Abstract

In this work, the real-time measurement of compensatory reserve metrics is compared during a simulated central hypovolemia clinical research study. We hypothesized that values generated from the compensatory reserve measurement (CRM) device would be statistically similar to those values generated from the compensatory reserve index (CRI) device. A study was conducted to collect data from noninvasive photoplethysmography sensors to validate the CRM against an FDA-cleared CRI device. Participants were placed in a lower body negative pressure (LBNP) chamber and underwent exposure to a stepwise protocol of increasing negative pressure. Data from both algorithmic devices were evaluated using multiple approaches: median error, median absolute error, and pooled Pearson correlations. Data were compared from 20 participants through the duration of the experiment. Pooled correlation analysis between CRM and CRI for all subjects and LBNP steps achieved R2 = 0.859, showing strong linear agreement. Bland–Altman analysis revealed a near-zero mean bias between CRM and CRI (0.13%). Each metric allowed for a mean early measurement time of 20.52 min for CRM and 17.92 min for CRI. Consistent with our hypothesis, the CRM device performed similarly to CRI, demonstrating that both models can track compensatory status changes with progressive central hypovolemia. The early, real-time predictive tracking of compensatory changes can be valuable for continuous monitoring and triaging in prehospital trauma environments.

BioengineeringVol. 13(7)
Uniformed Services University of the Health Sciences (US), The University of Texas at San Antonio Health Science Center (US), United States Army Institute of Surgical Research (US)
Openalex Percentile: Top 15%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.