Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment

ABSTRACT Fluorescent capillary gel electrophoresis (CGE) with sodium dodecyl sulfate (CE‐SDS) provides a powerful, high‐sensitivity alternative to ultraviolet (UV)‐based detection for characterizing therapeutic monoclonal antibodies (mAbs). Regulatory and standards organizations, such as the United States Pharmacopeia (USP), include only UV‐based CE‐SDS methods, hindering the adoption of alternative detection methods. There is growing opportunity to expand beyond exclusively UV‐based CE‐SDS methods. In this study, we present an analytical validation of a light‐emitting diode (LED) fluorescence‐based parallel CE‐SDS method for both non‐reduced and reduced analysis of therapeutic antibodies. Using the NISTmAb reference material as a model system, size heterogeneity critical quality attributes (CQAs), including monomeric purity (non‐reduced conditions), percent glycosylation (reduced conditions), and percent thioether (reduced conditions), were assessed. The fluorescence method demonstrated high specificity and precision with relative standard deviation (RSD) values <1% for monomeric purity and glycosylation, and <3% for thioether, as well as robust performance across variations in injection voltage, electrophoresis voltage, labeling temperature, and Labeling Buffer concentration. Ruggedness testing across users and reagent lots confirmed reproducibility, and accuracy assessments showed strong agreement with reported values from the National Institute of Standards and Technology (NIST) and traditional UV detection measurements. Linearity studies yielded the coefficient of determination ( R 2 ) values >0.99 for both non‐reduced and reduced analyses. These results highlight the high sensitivity, stable baseline performance, and suitability of LED fluorescence‐based parallel CE‐SDS as a validated, higher‐throughput alternative to traditional UV‐based methods for mAb quality control (QC).

Authors

Institutions

Publication Details

Journal
Electrophoresis
Published
2026-10-30
DOI
https://doi.org/10.1002/elps.70174
Primary Topic
Biosimilars and Bioanalytical Methods
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment

Sheng‐Yuan Huang, Kyle D. Luttgeharm, Whitney A Pike, Manu Grover
1 links
Electrophoresis
Biosimilars and Bioanalytical Methods
article

Validation of a High‐Throughput Fluorescent Capillary Electrophoresis Sodium Dodecyl Sulfate Method for Monoclonal Antibody Size Heterogeneity Assessment

Sheng‐Yuan Huang, Kyle D. Luttgeharm, Whitney A Pike, Manu Grover
article en

Abstract

ABSTRACT Fluorescent capillary gel electrophoresis (CGE) with sodium dodecyl sulfate (CE‐SDS) provides a powerful, high‐sensitivity alternative to ultraviolet (UV)‐based detection for characterizing therapeutic monoclonal antibodies (mAbs). Regulatory and standards organizations, such as the United States Pharmacopeia (USP), include only UV‐based CE‐SDS methods, hindering the adoption of alternative detection methods. There is growing opportunity to expand beyond exclusively UV‐based CE‐SDS methods. In this study, we present an analytical validation of a light‐emitting diode (LED) fluorescence‐based parallel CE‐SDS method for both non‐reduced and reduced analysis of therapeutic antibodies. Using the NISTmAb reference material as a model system, size heterogeneity critical quality attributes (CQAs), including monomeric purity (non‐reduced conditions), percent glycosylation (reduced conditions), and percent thioether (reduced conditions), were assessed. The fluorescence method demonstrated high specificity and precision with relative standard deviation (RSD) values <1% for monomeric purity and glycosylation, and <3% for thioether, as well as robust performance across variations in injection voltage, electrophoresis voltage, labeling temperature, and Labeling Buffer concentration. Ruggedness testing across users and reagent lots confirmed reproducibility, and accuracy assessments showed strong agreement with reported values from the National Institute of Standards and Technology (NIST) and traditional UV detection measurements. Linearity studies yielded the coefficient of determination ( R 2 ) values >0.99 for both non‐reduced and reduced analyses. These results highlight the high sensitivity, stable baseline performance, and suitability of LED fluorescence‐based parallel CE‐SDS as a validated, higher‐throughput alternative to traditional UV‐based methods for mAb quality control (QC).

Electrophoresis
Agilent Technologies (United States) (US)
Openalex Percentile: Top 56%
Biosimilars and Bioanalytical Methods
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon