Endoscopic salvage of a N-butyl cyanoacrylate-occluded gastroscope using cholangioscopic electrohydraulic lithotripsy
A 57-year-old man with decompensated cirrhosis was admitted for acute hemorrhage from esophagogastric varices. During endoscopic hemostasis, intra-variceal injection of N -butyl cyanoacrylate (NBCA) was administered to gastric fundal varices. Immediately following injection, NBCA polymerized upon contact with a luminal fluid and refluxed accidentally into the working channel, resulting in the complete obstruction of the therapeutic gastroscope. The occluded scope was withdrawn and exchanged for a new one to complete the hemostatic procedure, as the original was rendered nonfunctional. Standard salvage maneuvers, including high-pressure water irrigation, disposable brushes, and alligator forceps, were used to remove the hardened NBCA of the working channel but failed. The NBCA had solidified and tenaciously adhered to the channel wall, rendering conventional methods ineffective. Therefore, we attempted to use a training cholangioscope to fragment the solidified NBCA via lithotripsy. Direct cholangioscope vision revealed large NBCA casts obstructing the lumen, with an electrohydraulic probe targeting them ([ Fig. 1 ]). Electrohydraulic lithotripsy (EHL) was performed at an energy set of 0.3J per pulse, with continuous irrigation of sterile distilled water at 5 mL/min. The NBCA particles were progressively fragmented into small particles ([ Fig. 2 ]), which were then pushed out ([ Fig. 3 ], [ Video 1 ]). Post-procedural inspection confirmed the complete clearance of the working channel without macroscopic damage to its lining ([ Fig. 4 ]). Fig. 1 Direct cholangioscope vision revealed large NBCA particles obstructing the lumen, with an electrohydraulic probe targeting them. Fig. 2 The NBCA particles were progressively fragmented into small particles by cholangioscopic electrohydraulic lithotripsy. Fig. 3 The NBCA particles were then pushed out via the working channel of the gastroscope. Video 1 Endoscopic salvage of a N -butyl cyanoacrylate-occluded gastroscope using cholangioscopic electrohydraulic lithotripsy. Download Video Fig. 4 Post-procedural inspection confirmed the complete clearance of the working channel without macroscopic damage to its lining. NBCA is highly effective for variceal obliteration[ 1 ] [ 2 ]; however, intraluminal polymerization cause critical damage to the endoscope, leading to substantial repair costs or prolonged downtime.[ 3 ] Conventional mechanical or hydraulic methods are often inadequate for large, solidified NBCA casts. This case demonstrates that cholangioscopic EHL provides a safe, efficient, and cost-effective salvage strategy, potentially obviating the need for costly factory repair and prolonged instrument unavailability. Endoscopy_UCTN_Code_TTT_1AO_2AO Publication History Received: 16 May 2026 Accepted after revision: 15 June 2026 Article published online: 16 July 2026 © 2026. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/). Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Authors
- Yuqing Mao (ORCID: https://orcid.org/0000-0003-4573-6567)
- Bin Shi (ORCID: https://orcid.org/0000-0003-1775-4581)
- Sengwang Fu
Institutions
- Shanghai University of Traditional Chinese Medicine (CN)
- Yueyang Hospital (CN)
- Shanghai First People's Hospital (CN)
Publication Details
- Journal
- Endoscopy
- Published
- 2026-07-16
- DOI
- https://doi.org/10.1055/a-2895-4914
- Primary Topic
- Gallbladder and Bile Duct Disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00