Precision of an automated surface-based CT-RSA (asCTRSA) for assessing tibial implant migration: a porcine cadaver study and a clinical case

BACKGROUND AND PURPOSE: Radiostereometric analysis (RSA) is the reference standard for migration analysis, but adoption is limited by invasive marker implantation and specialized equipment. We developed an automated, surface-based CT-RSA (asCTRSA) method that reports a registration quality measure (mean rigid-body fitting error, ME). Our aim was to evaluate the precision by comparing asCTRSA with marker-based RSA and CT-based Micromotion Analysis (CTMA, v25.1). METHODS: Precision was assessed in a retrospective, repeated-measures, zero-migration study using a porcine cadaver tibial implant scanned in 7 positions (21 double examinations). Migration was reported as maximum total point motion (MTPM) for RSA and asCTRSA. CTMA provided maximum total translation (mTT) as an estimate of MTPM. Mean rigid-body fitting error (ME) was calculated for RSA and asCTRSA (acceptability threshold 0.35 mm). The minimal important difference for MTPM/mTT was 0.1 mm. asCTRSA was applied to 1 clinical case with known migration 3 and 8 months postoperatively (NCT04017533). RESULTS: In the porcine study, precision was 0.09 mm (95% confidence interval [CI] 0.02-0.18) for asCTRSA, 0.08 mm (CI 0.03-0.12) for CTMA, and 0.45 mm (CI 0.20-0.69) for RSA. In the clinical case, asCTRSA detected increasing migration (MTPM 2.5-3.6 mm) and exceeded CTMA's mTT by 0.4 mm at 8 months. Mean rigid-body fitting error remained below 0.35 mm. CONCLUSION: We showed that asCTRSA enables automated, surface-based CT-RSA and demonstrated higher precision than RSA in this zero-migration cadaver setting.

Authors

Institutions

Publication Details

Journal
Acta Orthopaedica
Published
2026-07-16
DOI
https://doi.org/10.2340/17453674.2026.46156
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Precision of an automated surface-based CT-RSA (asCTRSA) for assessing tibial implant migration: a porcine cadaver study and a clinical case

Benyameen Keelson, Nico Buls, Frank‐David Øhrn, Gert Van Gompel et al.
Acta Orthopaedica
Total Knee Arthroplasty Outcomes
article

Precision of an automated surface-based CT-RSA (asCTRSA) for assessing tibial implant migration: a porcine cadaver study and a clinical case

Benyameen Keelson, Nico Buls, Frank‐David Øhrn, Gert Van Gompel, Johan De Mey, Lars H. W. Engseth, Stephan M. Röhrl, Anselm Schulz, Manou Acke
article en

Abstract

BACKGROUND AND PURPOSE: Radiostereometric analysis (RSA) is the reference standard for migration analysis, but adoption is limited by invasive marker implantation and specialized equipment. We developed an automated, surface-based CT-RSA (asCTRSA) method that reports a registration quality measure (mean rigid-body fitting error, ME). Our aim was to evaluate the precision by comparing asCTRSA with marker-based RSA and CT-based Micromotion Analysis (CTMA, v25.1). METHODS: Precision was assessed in a retrospective, repeated-measures, zero-migration study using a porcine cadaver tibial implant scanned in 7 positions (21 double examinations). Migration was reported as maximum total point motion (MTPM) for RSA and asCTRSA. CTMA provided maximum total translation (mTT) as an estimate of MTPM. Mean rigid-body fitting error (ME) was calculated for RSA and asCTRSA (acceptability threshold 0.35 mm). The minimal important difference for MTPM/mTT was 0.1 mm. asCTRSA was applied to 1 clinical case with known migration 3 and 8 months postoperatively (NCT04017533). RESULTS: In the porcine study, precision was 0.09 mm (95% confidence interval [CI] 0.02-0.18) for asCTRSA, 0.08 mm (CI 0.03-0.12) for CTMA, and 0.45 mm (CI 0.20-0.69) for RSA. In the clinical case, asCTRSA detected increasing migration (MTPM 2.5-3.6 mm) and exceeded CTMA's mTT by 0.4 mm at 8 months. Mean rigid-body fitting error remained below 0.35 mm. CONCLUSION: We showed that asCTRSA enables automated, surface-based CT-RSA and demonstrated higher precision than RSA in this zero-migration cadaver setting.

Acta OrthopaedicaVol. 97
Oslo University Hospital (NO), Vrije Universiteit Brussel (BE), Norwegian University of Science and Technology (NO), Akershus University Hospital (NO), Universitair Ziekenhuis Brussel (BE)
Reduced inequalities
Openalex Percentile: Top 7%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.