CT-Guided Intervention: Reduction of Metal Artifact Using an Off-Axis Reconstruction Method

OBJECTIVE: Needle artifact redistribution technique (Needle-ART) is a recently developed image artifact-reduction method for CT-guided needle procedures. The principle of this technique is to create an angle between the needle and imaging plane by tilting the CT gantry, then reformatting images along the needle plane, thus redistributing the artifact out of the viewing plane. The purpose of this study is to evaluate the feasibility, artifact reduction, and clinical applicability of Needle-ART for the placement of biopsy introducer needles into the liver, kidneys, and lungs of a deceased animal model. METHODS: A paired before-and-after study was performed in 2 deceased swine to evaluate the effectiveness of Needle-ART in reducing metal artifacts during CT-guided interventions. A 17G co-axial needle was inserted into the liver, kidney, and lung at 6 sites per organ and scanned at each site with both 0-degree gantry tilt (standard axial as the control group) and 6-degree tilted gantry on Needle-ART. A total of 48 scans with 24 control and 24 Needle-ART were collected. Images were reformatted to align with the needle trajectory. Collected images were evaluated for overall artifact severity by 3 blinded radiologists. RESULTS: Needle-ART significantly reduced metal artifacts across all quantitative metrics, which was consistent with blinded reader assessments. Mean artifact length was reduced by 35.5% (P=0.0019) in liver, 65.9% (P=0.003) in kidney, and 59.4% (P=0.0021) in lung. Median artifact width decreased by 30% (P=0.028) in liver, 24.3% (P=0.028) in kidney, and 8.3% (P=0.017) in lung. Median artifact intensity was reduced by 38.8% (P=0.011) in liver, 35.4% (P=0.028) in kidney, and 23.6% (P=0.021) in lung. The radiologists scored conventional CT images (with non-tilted gantry) significantly worse artifacts compared with Needle-ART (with tilted gantry) (P=0.000041). CONCLUSIONS: The off-axis reconstruction method is a novel technique that effectively reduces artifacts by acquiring scans with the CT gantry tilted 6 degrees relative to the plane containing the needle and target.

Authors

Institutions

Publication Details

Journal
Journal of Computer Assisted Tomography
Published
2026-07-16
DOI
https://doi.org/10.1097/rct.0000000000001904
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CT-Guided Intervention: Reduction of Metal Artifact Using an Off-Axis Reconstruction Method

Gretchen Foltz, Martin G. Wagner, A. Kisting, Anna M. Sorensen et al.
Journal of Computer Assisted Tomography
Advanced X-ray and CT Imaging
article

CT-Guided Intervention: Reduction of Metal Artifact Using an Off-Axis Reconstruction Method

Gretchen Foltz, Martin G. Wagner, A. Kisting, Anna M. Sorensen, Timothy P. Szczykutowicz, Ali Mutlu, Annie M. Zlevor, Meridith A. Kisting, Fred T. Lee, Allison B. Couillard, A. M. Salyapongse, J. Louis Hinshaw
article en

Abstract

OBJECTIVE: Needle artifact redistribution technique (Needle-ART) is a recently developed image artifact-reduction method for CT-guided needle procedures. The principle of this technique is to create an angle between the needle and imaging plane by tilting the CT gantry, then reformatting images along the needle plane, thus redistributing the artifact out of the viewing plane. The purpose of this study is to evaluate the feasibility, artifact reduction, and clinical applicability of Needle-ART for the placement of biopsy introducer needles into the liver, kidneys, and lungs of a deceased animal model. METHODS: A paired before-and-after study was performed in 2 deceased swine to evaluate the effectiveness of Needle-ART in reducing metal artifacts during CT-guided interventions. A 17G co-axial needle was inserted into the liver, kidney, and lung at 6 sites per organ and scanned at each site with both 0-degree gantry tilt (standard axial as the control group) and 6-degree tilted gantry on Needle-ART. A total of 48 scans with 24 control and 24 Needle-ART were collected. Images were reformatted to align with the needle trajectory. Collected images were evaluated for overall artifact severity by 3 blinded radiologists. RESULTS: Needle-ART significantly reduced metal artifacts across all quantitative metrics, which was consistent with blinded reader assessments. Mean artifact length was reduced by 35.5% (P=0.0019) in liver, 65.9% (P=0.003) in kidney, and 59.4% (P=0.0021) in lung. Median artifact width decreased by 30% (P=0.028) in liver, 24.3% (P=0.028) in kidney, and 8.3% (P=0.017) in lung. Median artifact intensity was reduced by 38.8% (P=0.011) in liver, 35.4% (P=0.028) in kidney, and 23.6% (P=0.021) in lung. The radiologists scored conventional CT images (with non-tilted gantry) significantly worse artifacts compared with Needle-ART (with tilted gantry) (P=0.000041). CONCLUSIONS: The off-axis reconstruction method is a novel technique that effectively reduces artifacts by acquiring scans with the CT gantry tilted 6 degrees relative to the plane containing the needle and target.

Journal of Computer Assisted Tomography
University of Wisconsin–Madison (US), University of Illinois Urbana-Champaign (US)
Sustainable cities and communities
Openalex Percentile: Top 16%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.