Accurate 20 ms protein sizing with microscope‑based dynamic light scattering
Dynamic Light Scattering (DLS) is a powerful technique to probe nano to microscale objects in solution. Among its main limitations is the requirement of steady-state conditions during measurement, restricting its applicability towards time evolving systems. Here, we present a microscopy-based DLS approach using oblique illumination to significantly enhance the signal-to-background ratio compared to conventional confocal geometries. The coherence factor 𝛽 of the autocorrelation function increases to values above 0.9, approaching the ideal limit. Combined with a custom data analysis pipeline enabling post-acquisition processing over arbitrary time windows, this approach yields narrower and better-defined size distributions when averaging over shorter time intervals. Weak scatterers such as BSA proteins can be accurately measured at temporal resolutions down to 20 ms. These results extend the applicability of DLS beyond steady-state conditions and enable time-resolved size determination in dynamic systems.
Authors
- Alexander Biewald (ORCID: https://orcid.org/0000-0003-0660-2072)
- Natercia Barbosa (ORCID: https://orcid.org/0000-0001-9800-3514)
- Takuji Adachi (ORCID: https://orcid.org/0000-0002-0577-1926)
- QUENTIN STUCKER
Publication Details
- Journal
- Optics Express
- Published
- 2026-11-18
- DOI
- https://doi.org/10.1364/oe.616301
- Primary Topic
- Advanced Fluorescence Microscopy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00