Cross-Platform Retrieval of a Chronically Implanted Leadless Pacemaker Using a Dedicated Retrieval System
BACKGROUND: Leadless pacemakers are increasingly used in patients in whom transvenous pacing leads are undesirable. However, management at battery depletion remains challenging, particularly for chronically implanted devices. CASE SUMMARY: A 74-year-old woman with prior bioprosthetic tricuspid valve replacement, atrial fibrillation, and complete atrioventricular block underwent leadless pacemaker implantation in 2019. At follow-up, device interrogation demonstrated battery depletion. A new leadless pacemaker was implanted via right femoral venous access. The chronically implanted device, located in the mid-right ventricular septum, was subsequently retrieved using a trilobed snare system. Engagement required multiple attempts because of difficulty achieving coaxial alignment; however, after capture, the device was recaptured and removed with minimal resistance and without arrhythmia, pacing instability, or complication. CONCLUSION: This case demonstrates the feasibility of cross-platform retrieval of a chronically implanted leadless pacemaker after prolonged dwell time in selected patients.
Authors
- Apichai Pokawattana
- Surachat Jaroonpipatkul (ORCID: https://orcid.org/0000-0003-0200-848X)
Institutions
- Rangsit University (TH)
Publication Details
- Journal
- JACC Case Reports
- Published
- 2026-07-01
- DOI
- https://doi.org/10.1016/j.jaccas.2026.109138
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00