Flow field characteristics of surgical bioprosthetic valves vs bileaflet mechanical heart valves under different physiological states: A comparative PIV study

The hemodynamic performance of prosthetic heart valves is critical to long-term postoperative outcomes. This study systematically compares the detailed flow field characteristics of aortic surgical bioprosthetic valves (SBV) and bileaflet mechanical heart valves (BMHV) under different physiological conditions, aiming to provide a fluid-dynamics-based rationale for personalized valve selection. An in vitro pulsatile left heart mock circulatory loop was established using a transparent silicone aortic root model reconstructed from the computed tomography data, with a refractive index-matched working fluid to eliminate optical distortion. Time-resolved particle image velocimetry was employed to quantify transient velocity fields, vortex evolution, and viscous shear stress (VSS) distributions downstream of both valve types under healthy, heart failure, and exercise conditions. The results show that the SBV consistently generates a broad, uniform, and stable central jet across all conditions, whereas the BMHV produces a characteristic triple-jet structure accompanied by pronounced wake separation. Under heart failure conditions, the low-velocity wake region downstream of the BMHV expands markedly, increasing the likelihood of flow stagnation. In contrast to the relatively stable vortex dynamics observed in the SBV, the BMHV induces strong shear layers and pronounced vortex shedding, particularly during late systole. Furthermore, the BMHV exhibits significantly higher VSS levels than the SBV in all cases, with a substantial increase under exercise conditions, suggesting potentially elevated risks of hemolysis and platelet activation. Overall, these findings indicate that the SBV provides more favorable hemodynamic performance, while the BMHV may be associated with increased risks of blood damage and thrombogenesis under specific physiological states.

Authors

Institutions

Publication Details

Journal
Physics of Fluids
Published
2026-07-01
DOI
https://doi.org/10.1063/5.0339130
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Flow field characteristics of surgical bioprosthetic valves vs bileaflet mechanical heart valves under different physiological states: A comparative PIV study

Wenyu Geng, Zhixiong Li, Yan Qiang, Liang Qi et al.
Physics of Fluids
Cardiac Valve Diseases and Treatments
article

Flow field characteristics of surgical bioprosthetic valves vs bileaflet mechanical heart valves under different physiological states: A comparative PIV study

Wenyu Geng, Zhixiong Li, Yan Qiang, Liang Qi, Jiabao Qi
article en

Abstract

The hemodynamic performance of prosthetic heart valves is critical to long-term postoperative outcomes. This study systematically compares the detailed flow field characteristics of aortic surgical bioprosthetic valves (SBV) and bileaflet mechanical heart valves (BMHV) under different physiological conditions, aiming to provide a fluid-dynamics-based rationale for personalized valve selection. An in vitro pulsatile left heart mock circulatory loop was established using a transparent silicone aortic root model reconstructed from the computed tomography data, with a refractive index-matched working fluid to eliminate optical distortion. Time-resolved particle image velocimetry was employed to quantify transient velocity fields, vortex evolution, and viscous shear stress (VSS) distributions downstream of both valve types under healthy, heart failure, and exercise conditions. The results show that the SBV consistently generates a broad, uniform, and stable central jet across all conditions, whereas the BMHV produces a characteristic triple-jet structure accompanied by pronounced wake separation. Under heart failure conditions, the low-velocity wake region downstream of the BMHV expands markedly, increasing the likelihood of flow stagnation. In contrast to the relatively stable vortex dynamics observed in the SBV, the BMHV induces strong shear layers and pronounced vortex shedding, particularly during late systole. Furthermore, the BMHV exhibits significantly higher VSS levels than the SBV in all cases, with a substantial increase under exercise conditions, suggesting potentially elevated risks of hemolysis and platelet activation. Overall, these findings indicate that the SBV provides more favorable hemodynamic performance, while the BMHV may be associated with increased risks of blood damage and thrombogenesis under specific physiological states.

Physics of FluidsVol. 38(7)
Nanyang Technological University (SG), Lanzhou University of Technology (CN), First Hospital of Lanzhou University (CN), Lanzhou University (CN)
Good health and well-being
Openalex Percentile: Top 7%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.