Biomechanical effects of varus stem alignment on a cementless metadiaphyseal anchoring hip stem: a biomechanical investigation

BACKGROUND: Minimally invasive surgical approaches have become standard in cementless hip arthroplasty; however, the recent literature still lacks biomechanical evidence regarding the effects of stem malpositioning. This study aimed to biomechanically evaluate the stability of a cementless metadiaphyseal anchoring stem implanted with a varus malposition versus a neutral alignment. METHODS: Twenty paired human cadaveric femora were assigned pairwise for stem implantation featuring either 8° varus malposition (Group 1) or neutral orientation (Group 2). All specimens underwent quasi-static testing and progressively increasing cyclic loading to failure with monitoring via motion tracking. RESULTS: Axial stiffness in Group 1 was significantly lower than in Group 2, p = 0.023. Implant size in Group 1 was significantly smaller than that in Group 2, p = 0.002. Load at 0.15 mm stem subsidence and cycles to 0.15 mm subsidence in Group 1 were not significantly different compared to Group 2, p = 0.214. Load at 1 mm subsidence and cycles to 1 mm subsidence in Group 1 were significantly higher than those in Group 2, p = 0.022. CONCLUSION: An 8° varus-aligned cementless metadiaphyseal anchoring hip stem demonstrates superior load-bearing capacity with higher loads and numbers of cycles until reaching defined subsidence thresholds under dynamic loading, as compared to neutral alignment. These results demonstrate the biomechanical tolerance of unintended intraoperative varus malalignment, but do not support or recommend intentional varus stem positioning.

Authors

Institutions

Publication Details

Journal
Arthroplasty
Published
2026-07-02
DOI
https://doi.org/10.1186/s42836-026-00413-7
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical effects of varus stem alignment on a cementless metadiaphyseal anchoring hip stem: a biomechanical investigation

Matthias Luger, Moritz Kraus, Philipp Kastner, Tobias Gotterbarm et al.
Arthroplasty
Orthopaedic implants and arthroplasty
article

Biomechanical effects of varus stem alignment on a cementless metadiaphyseal anchoring hip stem: a biomechanical investigation

Matthias Luger, Moritz Kraus, Philipp Kastner, Tobias Gotterbarm, Clemens Schopper, Boyko Gueorguiev, Ivan Zderic
article en

Abstract

BACKGROUND: Minimally invasive surgical approaches have become standard in cementless hip arthroplasty; however, the recent literature still lacks biomechanical evidence regarding the effects of stem malpositioning. This study aimed to biomechanically evaluate the stability of a cementless metadiaphyseal anchoring stem implanted with a varus malposition versus a neutral alignment. METHODS: Twenty paired human cadaveric femora were assigned pairwise for stem implantation featuring either 8° varus malposition (Group 1) or neutral orientation (Group 2). All specimens underwent quasi-static testing and progressively increasing cyclic loading to failure with monitoring via motion tracking. RESULTS: Axial stiffness in Group 1 was significantly lower than in Group 2, p = 0.023. Implant size in Group 1 was significantly smaller than that in Group 2, p = 0.002. Load at 0.15 mm stem subsidence and cycles to 0.15 mm subsidence in Group 1 were not significantly different compared to Group 2, p = 0.214. Load at 1 mm subsidence and cycles to 1 mm subsidence in Group 1 were significantly higher than those in Group 2, p = 0.022. CONCLUSION: An 8° varus-aligned cementless metadiaphyseal anchoring hip stem demonstrates superior load-bearing capacity with higher loads and numbers of cycles until reaching defined subsidence thresholds under dynamic loading, as compared to neutral alignment. These results demonstrate the biomechanical tolerance of unintended intraoperative varus malalignment, but do not support or recommend intentional varus stem positioning.

ArthroplastyVol. 8(1)
Johannes Kepler University of Linz (AT), AO Foundation (CH), University Hospital of Zurich (CH), Kepler Universitätsklinikum (AT)
AO Foundation, Universität Zürich
Openalex Percentile: Top 6%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.