Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities

[Purpose] We compared the heart rate (HR) measurement accuracy of commercially available PPG-based wearable devices against an ECG-based reference device across different exercise intensities.[Methods] Forty healthy young adults were recruited, and data from 37 participants were analyzed. HR was simultaneously measured using Polar H10 as the ECG-based reference and four PPG-based devices: Polar Verity Sense, Polar Pacer Pro, Apple Watch SE, and Galaxy Watch 6. Participants completed low-, moderate-, and high-intensity treadmill exercise. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), Lin’s concordance correlation coefficient (CCC), and Bland-Altman analysis.[Results] MAE and MAPE generally decreased as intensity increased. Significant differences in MAE were observed among the devices across intensities (<i>p</i> = 0.002). The Polar Verity Sense and Apple Watch SE showed significantly lower MAE than the Polar Pacer Pro and Galaxy Watch 6. During low-intensity exercise, the Apple Watch SE showed lower MAE than the Polar Pacer Pro and Galaxy Watch 6, whereas during high-intensity exercise, the Polar Verity Sense showed lower MAE than the Galaxy Watch 6. CCC values increased with exercise intensity and were consistently higher for the Polar Verity Sense and Apple Watch SE. Bland-Altman analysis showed narrower limits of agreement for the Apple Watch SE and Polar Verity Sense.[Conclusion] HR measurement accuracy differed substantially among the PPG-based wearable devices. The Apple Watch SE and Polar Verity Sense showed the highest agreement with the ECG-based reference, suggesting that wearable HR data should be interpreted based on device-specific validation.

Authors

Institutions

Publication Details

Journal
Physical Activity and Nutrition
Published
2026-06-30
DOI
https://doi.org/10.20463/pan.2026.0032
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities

Min‐Seong Ha, Jaehee Lee, Yoonyoung Jeong, Woo-Min Jeong et al.
Physical Activity and Nutrition
Non-Invasive Vital Sign Monitoring
article

Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities

Min‐Seong Ha, Jaehee Lee, Yoonyoung Jeong, Woo-Min Jeong, Sehwa Yoon
article en

Abstract

[Purpose] We compared the heart rate (HR) measurement accuracy of commercially available PPG-based wearable devices against an ECG-based reference device across different exercise intensities.[Methods] Forty healthy young adults were recruited, and data from 37 participants were analyzed. HR was simultaneously measured using Polar H10 as the ECG-based reference and four PPG-based devices: Polar Verity Sense, Polar Pacer Pro, Apple Watch SE, and Galaxy Watch 6. Participants completed low-, moderate-, and high-intensity treadmill exercise. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), Lin’s concordance correlation coefficient (CCC), and Bland-Altman analysis.[Results] MAE and MAPE generally decreased as intensity increased. Significant differences in MAE were observed among the devices across intensities (<i>p</i> = 0.002). The Polar Verity Sense and Apple Watch SE showed significantly lower MAE than the Polar Pacer Pro and Galaxy Watch 6. During low-intensity exercise, the Apple Watch SE showed lower MAE than the Polar Pacer Pro and Galaxy Watch 6, whereas during high-intensity exercise, the Polar Verity Sense showed lower MAE than the Galaxy Watch 6. CCC values increased with exercise intensity and were consistently higher for the Polar Verity Sense and Apple Watch SE. Bland-Altman analysis showed narrower limits of agreement for the Apple Watch SE and Polar Verity Sense.[Conclusion] HR measurement accuracy differed substantially among the PPG-based wearable devices. The Apple Watch SE and Polar Verity Sense showed the highest agreement with the ECG-based reference, suggesting that wearable HR data should be interpreted based on device-specific validation.

Physical Activity and NutritionVol. 30(2)
University of Seoul (KR), University of Tsukuba (JP), Gimcheon University (KR), Tsukuba International University (JP)
Openalex Percentile: Top 13%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.