Invasiveness Study of Supersonic Gas-Curtain-Based Ionization Profile Monitor for Medical Accelerators

In proton beam therapy, ideally, beam monitoring should be non-invasive to provide online real-time feedback, such that the total dose delivered to the patient is not significantly affected. The invasiveness of the Supersonic Gas-Curtain-Based Ionization Profile Monitor (SGC-IPM) system was quantified by perturbation in beam current and transverse beam profile parameters induced by the supersonic gas-curtain for a 4.9–5.3 keV electron beam, representing a worst-case scenario where perturbations can be more easily observable. The experimentally measured gas-curtain effects on transverse beam parameters (≤2%), intensity (≤−1%) and beam current (≤−1%) were small in magnitude and largely below resolution limits. To confirm these effects, order-of-magnitude beam–gas interaction approximations were calculated for the experimental energy range, demonstrating negligible energy loss with minor scattering, broadly consistent with the experimental results. Clinical proton beam gas-curtain predictions (70–250 MeV) indicate a further reduction of ∼104 compared to the experimental observations. Even under the conservative electron beam conditions used in this study, the observed perturbations were minor or unresolvable and measured effects were significantly smaller than spatial and dosimetry scales relevant to proton radiotherapy. Overall, the experimental measurements and supporting order-of-magnitude estimates demonstrate that the SGC-IPM introduces negligible perturbations to beam parameters and is predicted to provide non-invasive beam profile monitoring for clinical proton beam diagnostics.

Authors

Institutions

Publication Details

Journal
Instruments
Published
2026-06-30
DOI
https://doi.org/10.3390/instruments10030035
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Invasiveness Study of Supersonic Gas-Curtain-Based Ionization Profile Monitor for Medical Accelerators

Milaan Patel, W. Deane Butcher, Narender Kumar, Bharat Singh Rawat et al.
Instruments
Radiation Therapy and Dosimetry
article

Invasiveness Study of Supersonic Gas-Curtain-Based Ionization Profile Monitor for Medical Accelerators

Milaan Patel, W. Deane Butcher, Narender Kumar, Bharat Singh Rawat, Carsten Welsch, Oliver Stringer, Hao Zhang, Farhana Thesni Mada Parambil
article en

Abstract

In proton beam therapy, ideally, beam monitoring should be non-invasive to provide online real-time feedback, such that the total dose delivered to the patient is not significantly affected. The invasiveness of the Supersonic Gas-Curtain-Based Ionization Profile Monitor (SGC-IPM) system was quantified by perturbation in beam current and transverse beam profile parameters induced by the supersonic gas-curtain for a 4.9–5.3 keV electron beam, representing a worst-case scenario where perturbations can be more easily observable. The experimentally measured gas-curtain effects on transverse beam parameters (≤2%), intensity (≤−1%) and beam current (≤−1%) were small in magnitude and largely below resolution limits. To confirm these effects, order-of-magnitude beam–gas interaction approximations were calculated for the experimental energy range, demonstrating negligible energy loss with minor scattering, broadly consistent with the experimental results. Clinical proton beam gas-curtain predictions (70–250 MeV) indicate a further reduction of ∼104 compared to the experimental observations. Even under the conservative electron beam conditions used in this study, the observed perturbations were minor or unresolvable and measured effects were significantly smaller than spatial and dosimetry scales relevant to proton radiotherapy. Overall, the experimental measurements and supporting order-of-magnitude estimates demonstrate that the SGC-IPM introduces negligible perturbations to beam parameters and is predicted to provide non-invasive beam profile monitoring for clinical proton beam diagnostics.

InstrumentsVol. 10(3)
University of Liverpool (GB), Cockcroft Institute (GB)
Engineering and Physical Sciences Research Council, Science and Technology Facilities Council
Affordable and clean energy
Openalex Percentile: Top 8%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.