Comparison of two analytical platforms for quantification of neuroglial biomarkers in blood samples: a single molecule array and an electrochemiluminescence assay

Abstract Advances in ultrasensitive immunoassays have enabled reliable quantification of neuroglial biomarkers in blood, providing valuable insights into neurological disorders. However, cross-platform evaluations are necessary to ensure comparability and standardization. This study aimed to compare the analytical performance of a Single Molecule Array (Simoa) and an ultrasensitive electrochemiluminescence (ECL) assay for quantifying neurofilament light chain (NfL) and glial fibrillary acidic protein (GFAP) in human serum. Baseline serum samples from 174 participants in the RIFUND trial were analyzed in parallel on both platforms. Concentrations of NfL and GFAP were compared using Spearman correlation, Bland–Altman analyses, reproducibility assessments, dilution linearity, and cross-platform recovery. All samples were quantifiable on both platforms. NfL concentrations correlated strongly between methods (Spearman correlation r = 0.88, p < 0.0001), whereas GFAP correlated moderately ( r = 0.77, p < 0.0001). Inter- and intra-assay coefficients of variation were comparable between platforms for both analytes. The relationship between the two assays could be described with the following equations: NfL ECL = 7.63 × NfL SIMOA + 1.71 and GFAP ECL = 0.332 × GFAP SIMOA + 4.22. Dilution linearity was excellent on both platforms (R 2 > 0.99), although cross-platform recovery varied systematically across the analytical range. Both Simoa and ECL demonstrated strong analytical performance for neuroglial biomarker quantification in serum. Despite systematic differences in absolute concentrations, relative agreement was high, particularly for NfL. These findings highlight the need for platform harmonization and provide empirically derived conversion factors to support analytical comparability in research and clinical applications.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-06-30
DOI
https://doi.org/10.1038/s41598-026-56794-x
Primary Topic
Advanced biosensing and bioanalysis techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of two analytical platforms for quantification of neuroglial biomarkers in blood samples: a single molecule array and an electrochemiluminescence assay

Anders Svenningsson, Ida Erngren, Karl Sjölin, Kim Kultima et al.
Scientific Reports
Advanced biosensing and bioanalysis techniques
article

Comparison of two analytical platforms for quantification of neuroglial biomarkers in blood samples: a single molecule array and an electrochemiluminescence assay

Anders Svenningsson, Ida Erngren, Karl Sjölin, Kim Kultima, Joachim Burman, Asma Al-Grety, Anders Larsson, Julia Aulin
article en

Abstract

Abstract Advances in ultrasensitive immunoassays have enabled reliable quantification of neuroglial biomarkers in blood, providing valuable insights into neurological disorders. However, cross-platform evaluations are necessary to ensure comparability and standardization. This study aimed to compare the analytical performance of a Single Molecule Array (Simoa) and an ultrasensitive electrochemiluminescence (ECL) assay for quantifying neurofilament light chain (NfL) and glial fibrillary acidic protein (GFAP) in human serum. Baseline serum samples from 174 participants in the RIFUND trial were analyzed in parallel on both platforms. Concentrations of NfL and GFAP were compared using Spearman correlation, Bland–Altman analyses, reproducibility assessments, dilution linearity, and cross-platform recovery. All samples were quantifiable on both platforms. NfL concentrations correlated strongly between methods (Spearman correlation r = 0.88, p < 0.0001), whereas GFAP correlated moderately ( r = 0.77, p < 0.0001). Inter- and intra-assay coefficients of variation were comparable between platforms for both analytes. The relationship between the two assays could be described with the following equations: NfL ECL = 7.63 × NfL SIMOA + 1.71 and GFAP ECL = 0.332 × GFAP SIMOA + 4.22. Dilution linearity was excellent on both platforms (R 2 > 0.99), although cross-platform recovery varied systematically across the analytical range. Both Simoa and ECL demonstrated strong analytical performance for neuroglial biomarker quantification in serum. Despite systematic differences in absolute concentrations, relative agreement was high, particularly for NfL. These findings highlight the need for platform harmonization and provide empirically derived conversion factors to support analytical comparability in research and clinical applications.

Scientific ReportsVol. 16(1)
Uppsala University (SE), Uppsala University Hospital (SE), Karolinska Institutet (SE)
Uppsala Universitet
Openalex Percentile: Top 11%
Advanced biosensing and bioanalysis techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.