Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance

ABSTRACT Introduction Elevated shock impedance during extravascular implantable cardioverter‐defibrillator (EV‐ICD) implantation may preclude defibrillation threshold (DFT) testing despite appropriate retrosternal lead position. We describe suspected mediastinal air entrapment as a reversible mechanism and a simple preventive strategy. Methods and Results During early EV‐ICD implantation experience, two patients developed shock impedance > 200 Ω immediately after retrosternal lead placement, with associated sensing abnormalities. Lead repositioning did not normalize impedance, no pneumothorax was identified, and DFT testing was deferred. Shock impedance normalized spontaneously within 1–2 days, permitting successful DFT testing without lead revision. Transient mediastinal air entrapment along the defibrillation vector was suspected. A procedural modification was adopted in which 5–10 cm H 2 O positive end‐expiratory pressure was applied immediately before retrosternal introducer removal, followed by prompt EV‐ICD lead advancement. Among five consecutive subsequent implantations using this approach, initial shock impedance remained within the expected range, permitting immediate DFT testing without lead repositioning or delay. Conclusion Transient mediastinal air entrapment may cause elevated shock impedance during EV‐ICD implantation. Positive end‐expiratory pressure during introducer removal may mitigate this complication and facilitate reliable intraoperative DFT testing.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Electrophysiology
Published
2026-06-30
DOI
https://doi.org/10.1111/jce.70426
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance

Sunil Kapur, Robert N. Kerley
Journal of Cardiovascular Electrophysiology
Cardiac pacing and defibrillation studies
article

Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance

Sunil Kapur, Robert N. Kerley
article en

Abstract

ABSTRACT Introduction Elevated shock impedance during extravascular implantable cardioverter‐defibrillator (EV‐ICD) implantation may preclude defibrillation threshold (DFT) testing despite appropriate retrosternal lead position. We describe suspected mediastinal air entrapment as a reversible mechanism and a simple preventive strategy. Methods and Results During early EV‐ICD implantation experience, two patients developed shock impedance > 200 Ω immediately after retrosternal lead placement, with associated sensing abnormalities. Lead repositioning did not normalize impedance, no pneumothorax was identified, and DFT testing was deferred. Shock impedance normalized spontaneously within 1–2 days, permitting successful DFT testing without lead revision. Transient mediastinal air entrapment along the defibrillation vector was suspected. A procedural modification was adopted in which 5–10 cm H 2 O positive end‐expiratory pressure was applied immediately before retrosternal introducer removal, followed by prompt EV‐ICD lead advancement. Among five consecutive subsequent implantations using this approach, initial shock impedance remained within the expected range, permitting immediate DFT testing without lead repositioning or delay. Conclusion Transient mediastinal air entrapment may cause elevated shock impedance during EV‐ICD implantation. Positive end‐expiratory pressure during introducer removal may mitigate this complication and facilitate reliable intraoperative DFT testing.

Journal of Cardiovascular Electrophysiology
Brigham and Women's Hospital (US)
Good health and well-being
Openalex Percentile: Top 8%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.