Comparison of ELISA and MS/MS Methods for Urinary Glyphosate Analysis Reveals Overestimation and Poor Repeatability in Evaluated ELISA Methods

High Resolution Image Download MS PowerPoint Slide Glyphosate is one of the most widely used herbicides, and its human health effects remain debated. Risk assessment requires robust human biomonitoring (HBM) data generated using validated analytical methods. However, recent population studies employing commercial enzyme-linked immunosorbent assays (ELISA) have reported urinary glyphosate concentrations higher than those measured by mass spectrometry (MS), raising concerns about potential overestimation and data comparability. This study compares two MS/MS-based methods, including a novel ion chromatography-MS/MS approach, with two ELISA methods across external quality assessment (EQA) urine samples (n = 13), spiked urine samples (n = 3, 0.1–0.3 ng/mL), and real-world human urine samples (n = 40). The evaluated ELISA methods overestimated urinary glyphosate concentrations below 1 ng/mL in EQA and spiked samples and exhibited poor repeatability (often >20%) relative to MS/MS-based methods. Analyses of real-world samples confirmed this trend, with ELISA yielding higher quantification frequencies (70–100% vs 45–63%), higher median concentrations (1.06–1.10 vs <LOQ–0.11 ng/mL), respectively, and lower repeatability (>30%) than MS/MS methods. These findings indicate that the evaluated ELISA methods lack sufficient reliability for quantifying urinary glyphosate at concentrations relevant to nonoccupational exposure, where matrix-related interferences may compromise analytical performance. These results support the preferential use of MS/MS-based methods for generating robust HBM data in research and regulatory contexts.

Authors

Institutions

Publication Details

Journal
Environmental Science & Technology Letters
Published
2026-06-30
DOI
https://doi.org/10.1021/acs.estlett.6c00374
Primary Topic
Pesticide and Herbicide Environmental Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of ELISA and MS/MS Methods for Urinary Glyphosate Analysis Reveals Overestimation and Poor Repeatability in Evaluated ELISA Methods

Sébastien Denys, Clémentine Dereumeaux, Justine Lanson, Valérie Bouchart et al.
Environmental Science & Technology Letters
Pesticide and Herbicide Environmental Studies
article

Comparison of ELISA and MS/MS Methods for Urinary Glyphosate Analysis Reveals Overestimation and Poor Repeatability in Evaluated ELISA Methods

Sébastien Denys, Clémentine Dereumeaux, Justine Lanson, Valérie Bouchart, Arthur David, Alison Connolly, Clémence Fillol, Holger M. Koch, Fabien Mercier, Mari-Vorgan Louyer, Anne Gérard, Cécile JABALOT, Marie Pécheux
article en

Abstract

High Resolution Image Download MS PowerPoint Slide Glyphosate is one of the most widely used herbicides, and its human health effects remain debated. Risk assessment requires robust human biomonitoring (HBM) data generated using validated analytical methods. However, recent population studies employing commercial enzyme-linked immunosorbent assays (ELISA) have reported urinary glyphosate concentrations higher than those measured by mass spectrometry (MS), raising concerns about potential overestimation and data comparability. This study compares two MS/MS-based methods, including a novel ion chromatography-MS/MS approach, with two ELISA methods across external quality assessment (EQA) urine samples (n = 13), spiked urine samples (n = 3, 0.1–0.3 ng/mL), and real-world human urine samples (n = 40). The evaluated ELISA methods overestimated urinary glyphosate concentrations below 1 ng/mL in EQA and spiked samples and exhibited poor repeatability (often >20%) relative to MS/MS-based methods. Analyses of real-world samples confirmed this trend, with ELISA yielding higher quantification frequencies (70–100% vs 45–63%), higher median concentrations (1.06–1.10 vs <LOQ–0.11 ng/mL), respectively, and lower repeatability (>30%) than MS/MS methods. These findings indicate that the evaluated ELISA methods lack sufficient reliability for quantifying urinary glyphosate at concentrations relevant to nonoccupational exposure, where matrix-related interferences may compromise analytical performance. These results support the preferential use of MS/MS-based methods for generating robust HBM data in research and regulatory contexts.

Environmental Science & Technology Letters
University College Dublin (IE), Exponent (United States) (US), Inserm (FR), École des Hautes Études en Santé Publique (FR), The Centre for Health (New Zealand) (NZ), BC Platforms (Finland) (FI), Santé Publique France (FR), Normandie Université (FR), Institut de Recherche en Santé, Environnement et Travail (FR), Institute for Prevention and Occupational Medicine (DE), Labéo (FR), French Clinical Research Infrastructure Network (FR), Université de Rennes (FR), Université de Caen Normandie (FR)
No poverty
Openalex Percentile: Top 15%
Pesticide and Herbicide Environmental Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.