Influence of Residual Motion Within the Gating Window on Dose Distribution in Respiratory-gated Proton Therapy

<h3>Background/Aim:</h3> Respiratory-gated proton therapy can mitigate interplay effects between the proton beam and target motion. However, the residual motion of the target within the gating window (GW) may still cause clinically relevant dosimetric variations. This study aimed to evaluate target and normal-tissue doses associated with residual motion within the GW. <h3>Patients and Methods:</h3> Eight patients with non-small cell lung cancer were analyzed. For each GW setting, the internal clinical target volume (iCTV) was defined as the union of clinical target volumes (CTVs) within the GW. Maximum residual motion (MRM) was defined as the maximum CTV motion within the GW relative to the 50% phase. Treatment plans were generated for single-field uniform dose (SFUD) and intensity-modulated proton therapy (IMPT). Dynamic 4D dose (D4DD) was calculated, and differences in dose metrics between the treatment plan and D4DD (ΔDM) were evaluated for CTV D<sub>95%</sub>, D<sub>98%</sub>, and D<sub>2%</sub>, and for lungs V<sub>20GyE</sub> and average dose (AD). <h3>Results:</h3> Increasing MRM was associated with decreased CTV D<sub>95%</sub> and D<sub>98%</sub> and increased CTV D<sub>2%</sub> for both SFUD and IMPT. When MRM was ≤3 mm, CTV ΔDM remained within ±3%; however, some metrics exceeded 5% when MRM was &gt;5 mm. Lungs ΔAD showed a slight increasing trend with MRM but remained within 0.25%. Lungs ΔV<sub>20GyE</sub> showed no clear correlation with MRM and varied from −5.7% to 7.1%; however, the values remained within ±3% when MRM was &lt;3 mm. <h3>Conclusion:</h3> When MRM was &lt;3 mm, ΔDM for both the target and lungs were within ±3%. Therefore, selecting a GW that limits MRM to &lt;3 mm may represent an appropriate criterion for both SFUD and IMPT.

Authors

Institutions

Publication Details

Journal
In Vivo
Published
2026-06-30
DOI
https://doi.org/10.21873/invivo.14402
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Residual Motion Within the Gating Window on Dose Distribution in Respiratory-gated Proton Therapy

Sachika Shiraishi, Masashi YAMANAKA, Koichi Tokuuye, Takahiro Shimo et al.
In Vivo
Radiation Therapy and Dosimetry
article

Influence of Residual Motion Within the Gating Window on Dose Distribution in Respiratory-gated Proton Therapy

Sachika Shiraishi, Masashi YAMANAKA, Koichi Tokuuye, Takahiro Shimo, Kazunori Nitta, Taro Murai, Kazuki Matsumoto, AKIHIRO YAMANO
article en

Abstract

<h3>Background/Aim:</h3> Respiratory-gated proton therapy can mitigate interplay effects between the proton beam and target motion. However, the residual motion of the target within the gating window (GW) may still cause clinically relevant dosimetric variations. This study aimed to evaluate target and normal-tissue doses associated with residual motion within the GW. <h3>Patients and Methods:</h3> Eight patients with non-small cell lung cancer were analyzed. For each GW setting, the internal clinical target volume (iCTV) was defined as the union of clinical target volumes (CTVs) within the GW. Maximum residual motion (MRM) was defined as the maximum CTV motion within the GW relative to the 50% phase. Treatment plans were generated for single-field uniform dose (SFUD) and intensity-modulated proton therapy (IMPT). Dynamic 4D dose (D4DD) was calculated, and differences in dose metrics between the treatment plan and D4DD (ΔDM) were evaluated for CTV D<sub>95%</sub>, D<sub>98%</sub>, and D<sub>2%</sub>, and for lungs V<sub>20GyE</sub> and average dose (AD). <h3>Results:</h3> Increasing MRM was associated with decreased CTV D<sub>95%</sub> and D<sub>98%</sub> and increased CTV D<sub>2%</sub> for both SFUD and IMPT. When MRM was ≤3 mm, CTV ΔDM remained within ±3%; however, some metrics exceeded 5% when MRM was &gt;5 mm. Lungs ΔAD showed a slight increasing trend with MRM but remained within 0.25%. Lungs ΔV<sub>20GyE</sub> showed no clear correlation with MRM and varied from −5.7% to 7.1%; however, the values remained within ±3% when MRM was &lt;3 mm. <h3>Conclusion:</h3> When MRM was &lt;3 mm, ΔDM for both the target and lungs were within ±3%. Therefore, selecting a GW that limits MRM to &lt;3 mm may represent an appropriate criterion for both SFUD and IMPT.

In VivoVol. 40(4)
Shonan Kamakura General Hospital (JP)
Zero hunger
Openalex Percentile: Top 8%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.