Assessment of adaptive trabecular bone remodeling around a knee prosthesis stem using the Weinans model

Adaptive remodeling of trabecular bone surrounding orthopedic implants plays a critical role in maintaining prosthesis stability and preventing aseptic loosening caused by stress shielding and osteolysis. This study presents a numerical model approach combining Finite Element Analysis and Weinans' bone remodeling model to evaluate the initial density adaptation response of peri-prosthetic femoral trabecular bone under the critical loading condition occurring during the toe-off phase of normal gait. A high-fidelity model incorporating cortical and trabecular bone regions was developed to calculate stress distribution and strain energy density within the peri-prosthetic region. The mechanical stimulus field obtained from the Finite Element Model was subsequently used to evaluate localized remodeling activation and density-related adaptation tendencies. The results showed that localized regions near the stem tip exceeded the ultimate compressive strength of trabecular bone (7.18 MPa), suggesting potential osteolytic activity confined to a limited area. In contrast, stress in the cortical bone remained below reported failure thresholds. Regions subjected to elevated mechanical stimulus exhibited increased trabecular bone density and corresponding improvements in predicted mechanical strength, reaching values close to 10 MPa. The simulations also demonstrated that density adaptation reduced local deformation, thereby promoting mechanical stabilization at the prosthesis-bone interface. These findings support the trabecular bone's capacity to adapt mechanically to implant-induced loading conditions and demonstrate the usefulness of the proposed framework for evaluating peri-prosthetic remodeling behavior and implant load-transfer mechanisms.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2026-06-30
DOI
https://doi.org/10.1177/09544119261462409
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of adaptive trabecular bone remodeling around a knee prosthesis stem using the Weinans model

Agustín Vidal-Lesso, Héctor Plascencia-Mora, Osvaldo Iván Meza-Jr.-Hernández
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Orthopaedic implants and arthroplasty
article

Assessment of adaptive trabecular bone remodeling around a knee prosthesis stem using the Weinans model

Agustín Vidal-Lesso, Héctor Plascencia-Mora, Osvaldo Iván Meza-Jr.-Hernández
article en

Abstract

Adaptive remodeling of trabecular bone surrounding orthopedic implants plays a critical role in maintaining prosthesis stability and preventing aseptic loosening caused by stress shielding and osteolysis. This study presents a numerical model approach combining Finite Element Analysis and Weinans' bone remodeling model to evaluate the initial density adaptation response of peri-prosthetic femoral trabecular bone under the critical loading condition occurring during the toe-off phase of normal gait. A high-fidelity model incorporating cortical and trabecular bone regions was developed to calculate stress distribution and strain energy density within the peri-prosthetic region. The mechanical stimulus field obtained from the Finite Element Model was subsequently used to evaluate localized remodeling activation and density-related adaptation tendencies. The results showed that localized regions near the stem tip exceeded the ultimate compressive strength of trabecular bone (7.18 MPa), suggesting potential osteolytic activity confined to a limited area. In contrast, stress in the cortical bone remained below reported failure thresholds. Regions subjected to elevated mechanical stimulus exhibited increased trabecular bone density and corresponding improvements in predicted mechanical strength, reaching values close to 10 MPa. The simulations also demonstrated that density adaptation reduced local deformation, thereby promoting mechanical stabilization at the prosthesis-bone interface. These findings support the trabecular bone's capacity to adapt mechanically to implant-induced loading conditions and demonstrate the usefulness of the proposed framework for evaluating peri-prosthetic remodeling behavior and implant load-transfer mechanisms.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Universidad de Guanajuato (MX)
Affordable and clean energy
Openalex Percentile: Top 7%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.