Stereotactic Accuracy and Operative Metrics Utilizing the ClearPoint Frameless Intraoperative-CT Image-Based Guidance: A Multi-Institution Analysis

INTRODUCTION: MRI-guided stereotaxy provides high accuracy but is limited by the need for an interventional/intraoperative MRI environment and associated workflow complexity. ClearPoint Software v3.0 extends an established stereotactic workflow to a conventional operating room using intraoperative CT (iCT) imaging. OBJECTIVE: To evaluate stereotactic accuracy and operative metrics of the ClearPoint frameless iCT image-guided system for deep brain stimulation (DBS) lead implantation. METHODS: We performed a multi-center retrospective review of consecutive DBS lead placement procedures supported by ClearPoint Neuro Navigation System Software Version 3.0 across 7 sites during a limited market release (February 1, 2025-May 1, 2025). Primary outcomes were radial targeting error (distance between planned trajectory endpoint and final device tip position in X-Y) and operative duration (skin incision to closure). RESULTS: Twenty-three procedures were analyzed. Median age was 67 years (range 39-89). Most cases were bilateral (87%, n=20). Targets included subthalamic nucleus (39%, n=9), ventral intermediate nucleus (39%, n=9), and globus pallidus internus (22%, n=5). Targeting accuracy was high, with median radial error of 0.20 mm on the left (range 0-1.60) and 0.20 mm on the right (range 0-1.0). Median operative duration was 203 minutes (range 125-330). Median intraoperative scans were 6, and procedures required a median of 2 adjustments per side. CONCLUSION: In this first multi-institution report of ClearPoint Software v3.0 for DBS lead placement, the frameless iCT workflow demonstrated consistent submillimeter accuracy with favorable operative metrics in a standard operating room setting.

Authors

Publication Details

Journal
Stereotactic and Functional Neurosurgery
Published
2026-06-29
DOI
https://doi.org/10.1159/000553285
Primary Topic
Neurological disorders and treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Stereotactic Accuracy and Operative Metrics Utilizing the ClearPoint Frameless Intraoperative-CT Image-Based Guidance: A Multi-Institution Analysis

Nader Pouratian, Matthan Tharakan, Kishore Balasubramanian, Ellen L. Air et al.
Stereotactic and Functional Neurosurgery
Neurological disorders and treatments
article

Stereotactic Accuracy and Operative Metrics Utilizing the ClearPoint Frameless Intraoperative-CT Image-Based Guidance: A Multi-Institution Analysis

Nader Pouratian, Matthan Tharakan, Kishore Balasubramanian, Ellen L. Air, Rohan Aggarwal, Kiana Prather, Paul S. Larson, Brian Dalm, Helen Shi, Ryan Kochanski, Arjun Khanna, Andrew K. Conner
article en

Abstract

INTRODUCTION: MRI-guided stereotaxy provides high accuracy but is limited by the need for an interventional/intraoperative MRI environment and associated workflow complexity. ClearPoint Software v3.0 extends an established stereotactic workflow to a conventional operating room using intraoperative CT (iCT) imaging. OBJECTIVE: To evaluate stereotactic accuracy and operative metrics of the ClearPoint frameless iCT image-guided system for deep brain stimulation (DBS) lead implantation. METHODS: We performed a multi-center retrospective review of consecutive DBS lead placement procedures supported by ClearPoint Neuro Navigation System Software Version 3.0 across 7 sites during a limited market release (February 1, 2025-May 1, 2025). Primary outcomes were radial targeting error (distance between planned trajectory endpoint and final device tip position in X-Y) and operative duration (skin incision to closure). RESULTS: Twenty-three procedures were analyzed. Median age was 67 years (range 39-89). Most cases were bilateral (87%, n=20). Targets included subthalamic nucleus (39%, n=9), ventral intermediate nucleus (39%, n=9), and globus pallidus internus (22%, n=5). Targeting accuracy was high, with median radial error of 0.20 mm on the left (range 0-1.60) and 0.20 mm on the right (range 0-1.0). Median operative duration was 203 minutes (range 125-330). Median intraoperative scans were 6, and procedures required a median of 2 adjustments per side. CONCLUSION: In this first multi-institution report of ClearPoint Software v3.0 for DBS lead placement, the frameless iCT workflow demonstrated consistent submillimeter accuracy with favorable operative metrics in a standard operating room setting.

Stereotactic and Functional Neurosurgery
Partnerships for the goals
Openalex Percentile: Top 7%
Neurological disorders and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Stereotactic Accuracy and Operative Metrics Utilizing the ClearPoint Frameless Intraoperative-CT Image-Based Guidance: A Multi-Institution Analysis — Nader Pouratian, Matthan Tharakan, et al. · Stereotactic and Functional Neurosurgery (2026) | TGRS Research Map | TGRS