Comparison of blood culture contamination rates with standard practice versus two blood diversion devices at a single institution

ABSTRACT Blood culture contamination can lead to false-positive blood culture results, potentially contributing to increased costs and unneeded antibiotic utilization. Initial specimen diversion devices (ISDD) have been proposed to reduce blood culture contamination rates. A single-center, observational study conducted at Mayo Clinic Rochester compared blood culture contamination rates using Steripath or Kurin ISDDs versus standard blood culture collection protocols. Over 255 phlebotomists were trained in the use of the ISDDs, with 31,215 blood cultures, 410 of which were contaminated, collected from adults in the emergency department and inpatient wards. Baseline contamination rates were 1.29%–2.53% with standard collection methods. The Steripath ISDD with direct-to-media inoculation of blood culture bottles (Steripath-DTM) reduced blood culture contamination compared to standard practice (0.34% vs. 1.43%, respectively, P < 0.001). Follow-up studies with the alternative Steripath configuration with an integrated syringe (Steripath-syringe), and then the Kurin ISDD with syringe collection (Kurin-syringe), did not meet the primary endpoint of <0.5% contamination (Steripath-syringe: 1.28%; Kurin-syringe: 1.76%). All three ISDD configurations failed to meet our goal of >75% utilization, with an average measured utilization of 26% for the Steripath-DTM, 26.4% for the Steripath ISDD (combined DTM and syringe) in a follow-up study, and 41.5% in the Kurin study. Additionally, a survey of phlebotomists revealed low staff satisfaction with both Steripath ISDD configurations. Staff reported multiple ergonomic barriers limiting the Steripath ISDD’s utility, which precluded implementation of the Steripath-DTM in routine clinical practice. IMPORTANCE We conducted a very large, real-world study of two blood diversion devices to determine whether either would meet goals for reduced blood culture contamination and a high rate of utilization by phlebotomists.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Microbiology
Published
2026-06-29
DOI
https://doi.org/10.1128/jcm.00085-26
Primary Topic
Bacterial Identification and Susceptibility Testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of blood culture contamination rates with standard practice versus two blood diversion devices at a single institution

Michele Legried, Linda Backus, Casey Vieni, Priya Sampathkumar et al.
Journal of Clinical Microbiology
Bacterial Identification and Susceptibility Testing
article

Comparison of blood culture contamination rates with standard practice versus two blood diversion devices at a single institution

Michele Legried, Linda Backus, Casey Vieni, Priya Sampathkumar, Robin Patel, Brad S. Karon, Audrey N. Schuetz, Kelly Bute, Jill Mainella
article en

Abstract

ABSTRACT Blood culture contamination can lead to false-positive blood culture results, potentially contributing to increased costs and unneeded antibiotic utilization. Initial specimen diversion devices (ISDD) have been proposed to reduce blood culture contamination rates. A single-center, observational study conducted at Mayo Clinic Rochester compared blood culture contamination rates using Steripath or Kurin ISDDs versus standard blood culture collection protocols. Over 255 phlebotomists were trained in the use of the ISDDs, with 31,215 blood cultures, 410 of which were contaminated, collected from adults in the emergency department and inpatient wards. Baseline contamination rates were 1.29%–2.53% with standard collection methods. The Steripath ISDD with direct-to-media inoculation of blood culture bottles (Steripath-DTM) reduced blood culture contamination compared to standard practice (0.34% vs. 1.43%, respectively, P < 0.001). Follow-up studies with the alternative Steripath configuration with an integrated syringe (Steripath-syringe), and then the Kurin ISDD with syringe collection (Kurin-syringe), did not meet the primary endpoint of <0.5% contamination (Steripath-syringe: 1.28%; Kurin-syringe: 1.76%). All three ISDD configurations failed to meet our goal of >75% utilization, with an average measured utilization of 26% for the Steripath-DTM, 26.4% for the Steripath ISDD (combined DTM and syringe) in a follow-up study, and 41.5% in the Kurin study. Additionally, a survey of phlebotomists revealed low staff satisfaction with both Steripath ISDD configurations. Staff reported multiple ergonomic barriers limiting the Steripath ISDD’s utility, which precluded implementation of the Steripath-DTM in routine clinical practice. IMPORTANCE We conducted a very large, real-world study of two blood diversion devices to determine whether either would meet goals for reduced blood culture contamination and a high rate of utilization by phlebotomists.

Journal of Clinical Microbiology
Mayo Clinic in Arizona (US)
Good health and well-being
Openalex Percentile: Top 9%
Bacterial Identification and Susceptibility Testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.