Ultrasound-based assessment of peri-implant mucosal thickness: an ex vivo comparative study with artificial intelligence-assisted image analysis

BACKGROUND: Mucosal thickness (MT) is a key factor influencing peri-implant soft-tissue response and outcomes in dental implantology. This ex vivo study evaluated the agreement of peri-implant MT measurements obtained using ultrasound (US) standardized with a custom probe holder, compared with transgingival probing (TP) and cone-beam computed tomography (CBCT). METHODS: Porcine hemimandibles (n = 18) underwent guided implant placement. MT was measured at five standardized points using four approaches: (1) US with expert annotation, (2) US with artificial intelligence (AI)-based image segmentation, (3) CBCT, and (4) TP. US images (17 MHz) were independently annotated by two trained specialists; a deep-learning-based method was used to derive automated MT measurements. Method differences were analyzed using a linear mixed-effects model; agreement was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. RESULTS: The overall method effect was not significant (p = 0.105). Pairwise comparisons showed no significant difference between expert-annotated US and TP (p = 0.328), whereas CBCT yielded higher MT values than TP (p = 0.035). Agreement was moderate for expert-annotated US versus TP (ICC = 0.58; 95% confidence interval (CI): 0.42-0.70) and for expert-annotated US versus AI-segmented US (ICC = 0.67; 95% CI: 0.53-0.77), but poor for expert-annotated US versus CBCT (ICC = 0.14; 95% CI: -0.05-0.33). Bland-Altman analysis showed mean differences (95% limits of agreement) of 0.08 mm (- 0.96 mm to + 1.13 mm) for expert-annotated US - TP, - 0.01 mm (- 0.68 mm to + 0.67 mm) for expert-annotated US-AI-segmented US, and - 0.30 mm (- 2.29 mm to + 1.68 mm) for expert-annotated US-CBCT. CONCLUSIONS: Under controlled ex vivo conditions, expert-annotated US standardized with a custom probe holder showed moderate comparative agreement with TP, while AI-segmented measurements showed moderate agreement with expert annotation. CBCT showed limited agreement with US. This integrated approach represents a proof-of-concept requiring further in vivo validation.

Authors

Institutions

Publication Details

Journal
BMC Oral Health
Published
2026-06-27
DOI
https://doi.org/10.1186/s12903-026-08665-0
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ultrasound-based assessment of peri-implant mucosal thickness: an ex vivo comparative study with artificial intelligence-assisted image analysis

Klaus Radermacher, Nikol Petrova, Stefan Wolfart, Peter Broessner et al.
BMC Oral Health
Dental Implant Techniques and Outcomes
article

Ultrasound-based assessment of peri-implant mucosal thickness: an ex vivo comparative study with artificial intelligence-assisted image analysis

Klaus Radermacher, Nikol Petrova, Stefan Wolfart, Peter Broessner, Fabian Christleven, Juliana Marotti
article en

Abstract

BACKGROUND: Mucosal thickness (MT) is a key factor influencing peri-implant soft-tissue response and outcomes in dental implantology. This ex vivo study evaluated the agreement of peri-implant MT measurements obtained using ultrasound (US) standardized with a custom probe holder, compared with transgingival probing (TP) and cone-beam computed tomography (CBCT). METHODS: Porcine hemimandibles (n = 18) underwent guided implant placement. MT was measured at five standardized points using four approaches: (1) US with expert annotation, (2) US with artificial intelligence (AI)-based image segmentation, (3) CBCT, and (4) TP. US images (17 MHz) were independently annotated by two trained specialists; a deep-learning-based method was used to derive automated MT measurements. Method differences were analyzed using a linear mixed-effects model; agreement was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. RESULTS: The overall method effect was not significant (p = 0.105). Pairwise comparisons showed no significant difference between expert-annotated US and TP (p = 0.328), whereas CBCT yielded higher MT values than TP (p = 0.035). Agreement was moderate for expert-annotated US versus TP (ICC = 0.58; 95% confidence interval (CI): 0.42-0.70) and for expert-annotated US versus AI-segmented US (ICC = 0.67; 95% CI: 0.53-0.77), but poor for expert-annotated US versus CBCT (ICC = 0.14; 95% CI: -0.05-0.33). Bland-Altman analysis showed mean differences (95% limits of agreement) of 0.08 mm (- 0.96 mm to + 1.13 mm) for expert-annotated US - TP, - 0.01 mm (- 0.68 mm to + 0.67 mm) for expert-annotated US-AI-segmented US, and - 0.30 mm (- 2.29 mm to + 1.68 mm) for expert-annotated US-CBCT. CONCLUSIONS: Under controlled ex vivo conditions, expert-annotated US standardized with a custom probe holder showed moderate comparative agreement with TP, while AI-segmented measurements showed moderate agreement with expert annotation. CBCT showed limited agreement with US. This integrated approach represents a proof-of-concept requiring further in vivo validation.

BMC Oral Health
Medical University of Graz (AT), RWTH Aachen University (DE), FH Aachen (DE)
RWTH Aachen University
No poverty
Openalex Percentile: Top 6%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.