Effect of the EzPAP system on tidal volume and airway pressures in healthy adults

Introduction/Background The EzPAP system generates positive inspiratory and expiratory pressures to increase lung volumes, yet limited evidence describes how different flow settings influence its physiological effects. This study examined the impact of EzPAP on tidal volume (VT), minute ventilation (VE), airway pressures, and physiological responses in healthy adults. Methods A prospective repeated‑measures study was conducted in 26 healthy adults. VT, VE, airway pressures, respiratory rate, perceived exertion, discomfort, and hemodynamic variables were recorded at baseline and during EzPAP use across seven flow settings, flowEzPAP (0–15 L/min). Generalized linear mixed models evaluated the effects of flowEzPAP on all outcomes. Results Compared to zero flowEzPAP, VT increased with higher flowEzPAP, with a mean difference of 0.35 L at 15 L/min ( p < .001). VE decreased, with the greatest mean reduction of 0.84 L/min at 13 L/min ( p < .001). Peak inspiratory (PIP), peak expiratory (PEP), and end‑expiratory (PEEP) pressures rose significantly with flowEzPAP, with greatest mean increases of 5.40, 15.06, and 13.33 cmH₂O at 15 L/min (all p < .001). PIP and PEP predicted VT, and all pressures predicted VE ( p < .001). Respiratory rate decreased, while exertion and discomfort increased at higher flows. One transient adverse event occurred at the highest flowEzPAP. Discussion EzPAP produced clinically relevant increases in lung volumes and airway pressures with minimal physiological disturbance in healthy adults. Threshold flowEzPAP settings were identified for achieving therapeutic expiratory pressures. Conclusion EzPAP safely increased VT and airway pressures in healthy adults, supporting further evaluation in clinical populations with reduced lung volumes or risk of atelectasis.

Authors

Institutions

Publication Details

Journal
Canadian Journal of Respiratory Therapy
Published
2026-06-26
DOI
https://doi.org/10.29390/001c.163541
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of the EzPAP system on tidal volume and airway pressures in healthy adults

Bredge McCarren, Clarice Tang, Felicity Blackstock, Kerrie Saliba
Canadian Journal of Respiratory Therapy
Respiratory Support and Mechanisms
article

Effect of the EzPAP system on tidal volume and airway pressures in healthy adults

Bredge McCarren, Clarice Tang, Felicity Blackstock, Kerrie Saliba
article en

Abstract

Introduction/Background The EzPAP system generates positive inspiratory and expiratory pressures to increase lung volumes, yet limited evidence describes how different flow settings influence its physiological effects. This study examined the impact of EzPAP on tidal volume (VT), minute ventilation (VE), airway pressures, and physiological responses in healthy adults. Methods A prospective repeated‑measures study was conducted in 26 healthy adults. VT, VE, airway pressures, respiratory rate, perceived exertion, discomfort, and hemodynamic variables were recorded at baseline and during EzPAP use across seven flow settings, flowEzPAP (0–15 L/min). Generalized linear mixed models evaluated the effects of flowEzPAP on all outcomes. Results Compared to zero flowEzPAP, VT increased with higher flowEzPAP, with a mean difference of 0.35 L at 15 L/min ( p < .001). VE decreased, with the greatest mean reduction of 0.84 L/min at 13 L/min ( p < .001). Peak inspiratory (PIP), peak expiratory (PEP), and end‑expiratory (PEEP) pressures rose significantly with flowEzPAP, with greatest mean increases of 5.40, 15.06, and 13.33 cmH₂O at 15 L/min (all p < .001). PIP and PEP predicted VT, and all pressures predicted VE ( p < .001). Respiratory rate decreased, while exertion and discomfort increased at higher flows. One transient adverse event occurred at the highest flowEzPAP. Discussion EzPAP produced clinically relevant increases in lung volumes and airway pressures with minimal physiological disturbance in healthy adults. Threshold flowEzPAP settings were identified for achieving therapeutic expiratory pressures. Conclusion EzPAP safely increased VT and airway pressures in healthy adults, supporting further evaluation in clinical populations with reduced lung volumes or risk of atelectasis.

Canadian Journal of Respiratory TherapyVol. 62
Victoria School of Management (CH), Western Sydney University (AU)
Good health and well-being
Openalex Percentile: Top 7%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.