A Fully MRI-Compatible, Non-Electrical System for Near Real-Time Tremor Monitoring During MR-guided Focused Ultrasound

INTRODUCTION: Tremor quantification during Magnetic Resonance-guided Focused Ultrasound (MRgFUS) for essential tremor (ET) remains a clinical challenge. Current methods, such as accelerometry, can still be limited by electromagnetic artefacts in certain cases, magnetic resonance imaging (MRI) incompatibility, and indirect correlation with sonication timing. We aimed to develop a fully MRI-compatible, non-electrical system capable of continuous, electromagnetically artefact-free tremor monitoring with near real-time frequency analysis, precisely synchronized with ultrasound delivery. The primary purpose of this tool is to assist with targeting during the verification stage by providing the tremor amplitude in situ. METHODS: We designed a pneumatic device using a standard, easily available endotracheal cuff inflated to a fixed pressure, held by the patient during MRgFUS sonications. Tremor-induced pressure fluctuations were transmitted outside the MRI suite and digitized using a custom Arduino-based system. A software tool was implemented to calculate the tremor amplitude in near real-time. RESULTS: The pressure system allowed real-time visualization of the tremor and assessment of the tremor reduction at the verification stage of 10 patients. The mean tremor reduction was 86% immediately after the end of the verification sonication, confirming the correct position of the target. All patients had a lesion at this location except one, in which the target was moved because he experienced sensory side effects. CONCLUSION: We designed and implemented a near real-time pressure system for quantification of the tremor amplitude during sonications. The inexpensive and MR-compatible device is easy to implement on patients requiring MRgFUS treatments and avoid the possible loss of information between the temporary tremor reduction and the clinical evaluation.

Authors

Publication Details

Journal
Stereotactic and Functional Neurosurgery
Published
2026-06-26
DOI
https://doi.org/10.1159/000552899
Primary Topic
Neurological disorders and treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Fully MRI-Compatible, Non-Electrical System for Near Real-Time Tremor Monitoring During MR-guided Focused Ultrasound

Vanessa Fleury, Shahan Momjian, Orane Lorton, Paul Constanthin et al.
Stereotactic and Functional Neurosurgery
Neurological disorders and treatments
article

A Fully MRI-Compatible, Non-Electrical System for Near Real-Time Tremor Monitoring During MR-guided Focused Ultrasound

Vanessa Fleury, Shahan Momjian, Orane Lorton, Paul Constanthin, Adrien Lavalley
article en

Abstract

INTRODUCTION: Tremor quantification during Magnetic Resonance-guided Focused Ultrasound (MRgFUS) for essential tremor (ET) remains a clinical challenge. Current methods, such as accelerometry, can still be limited by electromagnetic artefacts in certain cases, magnetic resonance imaging (MRI) incompatibility, and indirect correlation with sonication timing. We aimed to develop a fully MRI-compatible, non-electrical system capable of continuous, electromagnetically artefact-free tremor monitoring with near real-time frequency analysis, precisely synchronized with ultrasound delivery. The primary purpose of this tool is to assist with targeting during the verification stage by providing the tremor amplitude in situ. METHODS: We designed a pneumatic device using a standard, easily available endotracheal cuff inflated to a fixed pressure, held by the patient during MRgFUS sonications. Tremor-induced pressure fluctuations were transmitted outside the MRI suite and digitized using a custom Arduino-based system. A software tool was implemented to calculate the tremor amplitude in near real-time. RESULTS: The pressure system allowed real-time visualization of the tremor and assessment of the tremor reduction at the verification stage of 10 patients. The mean tremor reduction was 86% immediately after the end of the verification sonication, confirming the correct position of the target. All patients had a lesion at this location except one, in which the target was moved because he experienced sensory side effects. CONCLUSION: We designed and implemented a near real-time pressure system for quantification of the tremor amplitude during sonications. The inexpensive and MR-compatible device is easy to implement on patients requiring MRgFUS treatments and avoid the possible loss of information between the temporary tremor reduction and the clinical evaluation.

Stereotactic and Functional Neurosurgery
Openalex Percentile: Top 7%
Neurological disorders and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Fully MRI-Compatible, Non-Electrical System for Near Real-Time Tremor Monitoring During MR-guided Focused Ultrasound — Vanessa Fleury, Shahan Momjian, et al. · Stereotactic and Functional Neurosurgery (2026) | TGRS Research Map | TGRS