Feasibility study of image reconstruction for a forceps-type positron emission counter: a simulation-based algorithm comparison

OBJECTIVE: The forceps-type positron emission counter (PEC) was developed for intraoperative quantification of 18 F-fluorodeoxyglucose uptake in individual lymph nodes. It employs coincidence detection between two opposing 2 × 2 bismuth germanate crystal arrays, yielding only 16 lines of response (LORs). This study investigates the feasibility of image reconstruction from these 16 LORs to enable repositioning guidance and sensitivity-corrected quantification. APPROACH: Seven reconstruction algorithms-spanning different data-fidelity models (least-squares and Poisson likelihood), regularization priors (none, L 2 Tikhonov, and total variation), and optimization frameworks-were compared using Geant4-based Monte Carlo simulations. Phantom experiments assessed reconstruction quality and count-rate dependence, localization accuracy across the field of view, hot-background robustness, and dual-source separation. MAIN RESULTS: Among the tested design axes, regularization has the largest impact on reconstruction quality, and the optimal algorithm depends on the task: MAP-EM-TV (Poisson-likelihood with TV regularization) achieves the highest structural similarity while PLS-TV (least-squares with TV) achieves the best quantitative accuracy (FoV-sum CV reduced by 21.5 percentage points) and a mean localization error of 1.27 mm. Near-optimal image quality is reached at ∼3,000 counts, with spatial information retained down to ∼100 counts; image quality degrades substantially below source-to-background ratio 12:1. SIGNIFICANCE: The results demonstrate that the 16-LOR PEC can provide spatial information beyond countrate or center-of-gravity estimation, enabling repositioning guidance and sensitivitycorrected quantification for intraoperative use. As the study relies on a shared forward model, validation with a physical prototype is the key next step.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-06-26
DOI
https://doi.org/10.1088/1361-6560/ae8356
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility study of image reconstruction for a forceps-type positron emission counter: a simulation-based algorithm comparison

Sodai Takyu, M. Takahashi, Shigeki Ito, Taiga Yamaya et al.
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
article

Feasibility study of image reconstruction for a forceps-type positron emission counter: a simulation-based algorithm comparison

Sodai Takyu, M. Takahashi, Shigeki Ito, Taiga Yamaya, Ryotaro Ohashi
article en

Abstract

OBJECTIVE: The forceps-type positron emission counter (PEC) was developed for intraoperative quantification of 18 F-fluorodeoxyglucose uptake in individual lymph nodes. It employs coincidence detection between two opposing 2 × 2 bismuth germanate crystal arrays, yielding only 16 lines of response (LORs). This study investigates the feasibility of image reconstruction from these 16 LORs to enable repositioning guidance and sensitivity-corrected quantification. APPROACH: Seven reconstruction algorithms-spanning different data-fidelity models (least-squares and Poisson likelihood), regularization priors (none, L 2 Tikhonov, and total variation), and optimization frameworks-were compared using Geant4-based Monte Carlo simulations. Phantom experiments assessed reconstruction quality and count-rate dependence, localization accuracy across the field of view, hot-background robustness, and dual-source separation. MAIN RESULTS: Among the tested design axes, regularization has the largest impact on reconstruction quality, and the optimal algorithm depends on the task: MAP-EM-TV (Poisson-likelihood with TV regularization) achieves the highest structural similarity while PLS-TV (least-squares with TV) achieves the best quantitative accuracy (FoV-sum CV reduced by 21.5 percentage points) and a mean localization error of 1.27 mm. Near-optimal image quality is reached at ∼3,000 counts, with spatial information retained down to ∼100 counts; image quality degrades substantially below source-to-background ratio 12:1. SIGNIFICANCE: The results demonstrate that the 16-LOR PEC can provide spatial information beyond countrate or center-of-gravity estimation, enabling repositioning guidance and sensitivitycorrected quantification for intraoperative use. As the study relies on a shared forward model, validation with a physical prototype is the key next step.

Physics in Medicine and Biology
National Institutes for Quantum Science and Technology (JP), Mi Corporation (United States) (US)
Support for Pioneering Research Initiated by the Next Generation
Openalex Percentile: Top 6%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.