Optimal dimensionality and fundamental limits of proton stopping power estimation with photon-counting CT material decomposition

Abstract Objective. Proton therapy can achieve high radiation dose to the tumor while sparing normal tissue beyond the dose fall-off. Accurate estimation of proton stopping power ratio (SPR) and range from computed tomography (CT) data is a prerequisite for minimizing range uncertainty and treatment margins. Photon-counting CT (PCCT) could potentially improve the accuracy of SPR estimation with the increased number of energy measurements. This work aims to assess the fundamental limits of SPR and range estimation via material decomposition (MD) with PCCT, while evaluating the feasibility for higher MD dimensionalities for proton treatment planning. Approach. Eigentissue decomposition is used for computing optimal basis materials for elemental composition estimation. We model a water phantom with a centered test tissue insert, and use Cramér-Rao Lower Bound to estimate the covariance of basis material sinogram noise for MD dimensionalities 1, 2, 3, and 4. We model an ideal photon-counting detector with 1 mm 2 detector pixels, and fluence corresponding to 260 mAs at 120 kV. For each dimensionality, the noise is propagated to SPR level and corresponding SPR bias error is calculated. Range uncertainties are estimated through simulations in RayStation, with SPR volumes corresponding to calculated bias and noise for each dimensionality. Main results. The three-material decomposition was optimal for soft tissues, while two-material decomposition was optimal for bone when estimating proton range. The noise increase for the three-material decomposition did not generally translate to greater deviations at range level. The lowest range RMSEs with an ideal photon-counting detector were <0.1-0.7 %, depending on tissue type and depth. Significance. This work indicates the limits for SPR and range estimation accuracy using PCCT with an MD-based approach, and shows that multimaterial decomposition through quantiative imaging with PCCT could improve tissue characterization and range estimation in proton therapy.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-06-26
DOI
https://doi.org/10.1088/1361-6560/ae8359
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimal dimensionality and fundamental limits of proton stopping power estimation with photon-counting CT material decomposition

Torbjörn Näsmark, Mats Persson, Erik Tranéus, Karin Larsson et al.
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Optimal dimensionality and fundamental limits of proton stopping power estimation with photon-counting CT material decomposition

Torbjörn Näsmark, Mats Persson, Erik Tranéus, Karin Larsson, J Andersson
article en

Abstract

Abstract Objective. Proton therapy can achieve high radiation dose to the tumor while sparing normal tissue beyond the dose fall-off. Accurate estimation of proton stopping power ratio (SPR) and range from computed tomography (CT) data is a prerequisite for minimizing range uncertainty and treatment margins. Photon-counting CT (PCCT) could potentially improve the accuracy of SPR estimation with the increased number of energy measurements. This work aims to assess the fundamental limits of SPR and range estimation via material decomposition (MD) with PCCT, while evaluating the feasibility for higher MD dimensionalities for proton treatment planning. Approach. Eigentissue decomposition is used for computing optimal basis materials for elemental composition estimation. We model a water phantom with a centered test tissue insert, and use Cramér-Rao Lower Bound to estimate the covariance of basis material sinogram noise for MD dimensionalities 1, 2, 3, and 4. We model an ideal photon-counting detector with 1 mm 2 detector pixels, and fluence corresponding to 260 mAs at 120 kV. For each dimensionality, the noise is propagated to SPR level and corresponding SPR bias error is calculated. Range uncertainties are estimated through simulations in RayStation, with SPR volumes corresponding to calculated bias and noise for each dimensionality. Main results. The three-material decomposition was optimal for soft tissues, while two-material decomposition was optimal for bone when estimating proton range. The noise increase for the three-material decomposition did not generally translate to greater deviations at range level. The lowest range RMSEs with an ideal photon-counting detector were <0.1-0.7 %, depending on tissue type and depth. Significance. This work indicates the limits for SPR and range estimation accuracy using PCCT with an MD-based approach, and shows that multimaterial decomposition through quantiative imaging with PCCT could improve tissue characterization and range estimation in proton therapy.

Physics in Medicine and Biology
RaySearch Laboratories (Sweden) (SE), KTH Royal Institute of Technology (SE), Umeå University (SE)
Göran Gustafssons Stiftelser, Cancerfonden, Vetenskapsrådet
Openalex Percentile: Top 12%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.