Rapid Centrifugation for Coagulation Testing: A Multianalyzer Validation Study Harmonizing Preanalytical Protocols for Total Laboratory Automation

OBJECTIVES: Centrifugation requirements differ between coagulation and clinical chemistry testing, potentially limiting workflow efficiency and total laboratory automation (TLA). Coagulation assays typically require standard centrifugation at 1500 g for 15 min, whereas chemistry samples use a rapid protocol at 2300 g for 5 min. This study evaluated the impact of rapid centrifugation on prothrombin time (PT), activated partial thromboplastin time (APTT), and D-dimer results. METHODS: Blood samples were prospectively collected from volunteers in 3, 2, and pediatric 1 mL citrate tubes and were processed using standard or rapid centrifugation. Testing was performed for PT, APTT, and D-Dimer on Sysmex CN3000 and ACL-TOP 750 analyzers, with residual platelet counts assessed. RESULTS: /L, p < 0.0001). Strong correlations (r > 0.9) were observed for PT and APTT across analyzers, while D-dimer results on ACL-TOP 750 showed slightly lower correlation (r = 0.830). Bland-Altman analysis demonstrated small mean biases (< 1% for PT and APTT). D-dimer biases from CN3000 and ACL-TOP 750 were 1.06% and -2.70%, respectively, which remained within the predefined total allowable error. Residual platelets showed no impact on test results. CONCLUSIONS: Rapid centrifugation at 2300 g for 5 min provides coagulation results comparable to the standard protocol and may facilitate integration of coagulation testing into TLA workflows.

Authors

Institutions

Publication Details

Journal
International Journal of Laboratory Hematology
Published
2026-06-26
DOI
https://doi.org/10.1111/ijlh.70188
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Rapid Centrifugation for Coagulation Testing: A Multianalyzer Validation Study Harmonizing Preanalytical Protocols for Total Laboratory Automation

Eakachai Prompetchara, Chutitorn Ketloy, Kwanlada Chaiwong, Surapat Tanticharoenkarn et al.
International Journal of Laboratory Hematology
Clinical Laboratory Practices and Quality Control
article

Rapid Centrifugation for Coagulation Testing: A Multianalyzer Validation Study Harmonizing Preanalytical Protocols for Total Laboratory Automation

Eakachai Prompetchara, Chutitorn Ketloy, Kwanlada Chaiwong, Surapat Tanticharoenkarn, Orakan Limpornpugdee, Waroonkarn Laiklang, Pitchayaporn Riyagoon, Parinda Limprasert, Phongsapak Pipatnavakij, Jirapa Kaewkhruawan, Wararat Masalae
article en

Abstract

OBJECTIVES: Centrifugation requirements differ between coagulation and clinical chemistry testing, potentially limiting workflow efficiency and total laboratory automation (TLA). Coagulation assays typically require standard centrifugation at 1500 g for 15 min, whereas chemistry samples use a rapid protocol at 2300 g for 5 min. This study evaluated the impact of rapid centrifugation on prothrombin time (PT), activated partial thromboplastin time (APTT), and D-dimer results. METHODS: Blood samples were prospectively collected from volunteers in 3, 2, and pediatric 1 mL citrate tubes and were processed using standard or rapid centrifugation. Testing was performed for PT, APTT, and D-Dimer on Sysmex CN3000 and ACL-TOP 750 analyzers, with residual platelet counts assessed. RESULTS: /L, p < 0.0001). Strong correlations (r > 0.9) were observed for PT and APTT across analyzers, while D-dimer results on ACL-TOP 750 showed slightly lower correlation (r = 0.830). Bland-Altman analysis demonstrated small mean biases (< 1% for PT and APTT). D-dimer biases from CN3000 and ACL-TOP 750 were 1.06% and -2.70%, respectively, which remained within the predefined total allowable error. Residual platelets showed no impact on test results. CONCLUSIONS: Rapid centrifugation at 2300 g for 5 min provides coagulation results comparable to the standard protocol and may facilitate integration of coagulation testing into TLA workflows.

International Journal of Laboratory Hematology
Chulalongkorn University (TH), King Chulalongkorn Memorial Hospital (TH)
Openalex Percentile: Top 7%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.