Optimising $$^{90}Y$$ PET imaging for dosimetry in SIRT: insights from phantom and simulation studies on the Discovery MI scanner

BACKGROUND: The aim of this study was to optimise the acquisition and reconstruction parameters for [Formula: see text] imaging on a digital PET scanner (Discovery MI 4-ring) under high (3 GBq), intermediate (1 GBq), and low (200 MBq) activity conditions. First, quantitative linearity was evaluated. Then, NEMA IEC body phantom acquisitions were reconstructed using various scan durations and Q.Clear reconstruction parameters. Next, optimal protocols were identified by minimising the discrepancies between absorbed dose maps derived from images (using the Local Deposition Model (LDM)) and those obtained from Monte Carlo (MC) simulations. The images reconstructed with these optimal protocols were then used to evaluate effective spatial resolution and to compare the accuracy of LDM with that of the Dose Voxel Kernel (DVK) approach. RESULTS: Quantitative linearity analysis revealed an underestimation of phantom activity at high activity (up to - 15.7% at 3.6 GBq) and an overestimation at lower levels (up to 52.1% at 208 MBq). At high activity, the most accurate results were produced by 15- and 20-minute acquisitions with intermediate beta values (2000-5000). In contrast, intermediate and low activity levels required longer acquisition times and higher beta values (> 6000). Effective spatial resolution ranged from 7 mm at high activity to 24.9 mm at low activity. In terms of absorbed dose accuracy, the mean absorbed dose was underestimated in all spheres of the phantom. However, the error in the mean absorbed dose could be kept within 10% in the background by applying non-linearity corrections. Using MC as reference, LDM achieved greater accuracy in terms of mean absorbed dose and Dose Volume Histogram (DVH) agreement in the spheres, while DVK performed better in the background. CONCLUSIONS: This study enabled the optimisation of scan duration and beta parameter values for Q.Clear reconstructions for dosimetric purposes. Although accurate dosimetry remains challenging in [Formula: see text] PET imaging due to quantitative nonlinearity and limited spatial resolution, an error of less than 10% on the mean absorbed dose can be achieved in large structures if model-specific nonlinearities are corrected for.

Authors

Institutions

Publication Details

Journal
EJNMMI Physics
Published
2026-06-25
DOI
https://doi.org/10.1186/s40658-026-00912-1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimising $$^{90}Y$$ PET imaging for dosimetry in SIRT: insights from phantom and simulation studies on the Discovery MI scanner

Hussein Harb, Romain Popoff, Benoît Presles, Jean‐Marc Vrigneaud et al.
EJNMMI Physics
Medical Imaging Techniques and Applications
article

Optimising $$^{90}Y$$ PET imaging for dosimetry in SIRT: insights from phantom and simulation studies on the Discovery MI scanner

Hussein Harb, Romain Popoff, Benoît Presles, Jean‐Marc Vrigneaud, Farzam Sayah, Tien-Phong Pham, Jean‐Louis Alberini
article en

Abstract

BACKGROUND: The aim of this study was to optimise the acquisition and reconstruction parameters for [Formula: see text] imaging on a digital PET scanner (Discovery MI 4-ring) under high (3 GBq), intermediate (1 GBq), and low (200 MBq) activity conditions. First, quantitative linearity was evaluated. Then, NEMA IEC body phantom acquisitions were reconstructed using various scan durations and Q.Clear reconstruction parameters. Next, optimal protocols were identified by minimising the discrepancies between absorbed dose maps derived from images (using the Local Deposition Model (LDM)) and those obtained from Monte Carlo (MC) simulations. The images reconstructed with these optimal protocols were then used to evaluate effective spatial resolution and to compare the accuracy of LDM with that of the Dose Voxel Kernel (DVK) approach. RESULTS: Quantitative linearity analysis revealed an underestimation of phantom activity at high activity (up to - 15.7% at 3.6 GBq) and an overestimation at lower levels (up to 52.1% at 208 MBq). At high activity, the most accurate results were produced by 15- and 20-minute acquisitions with intermediate beta values (2000-5000). In contrast, intermediate and low activity levels required longer acquisition times and higher beta values (> 6000). Effective spatial resolution ranged from 7 mm at high activity to 24.9 mm at low activity. In terms of absorbed dose accuracy, the mean absorbed dose was underestimated in all spheres of the phantom. However, the error in the mean absorbed dose could be kept within 10% in the background by applying non-linearity corrections. Using MC as reference, LDM achieved greater accuracy in terms of mean absorbed dose and Dose Volume Histogram (DVH) agreement in the spheres, while DVK performed better in the background. CONCLUSIONS: This study enabled the optimisation of scan duration and beta parameter values for Q.Clear reconstructions for dosimetric purposes. Although accurate dosimetry remains challenging in [Formula: see text] PET imaging due to quantitative nonlinearity and limited spatial resolution, an error of less than 10% on the mean absorbed dose can be achieved in large structures if model-specific nonlinearities are corrected for.

EJNMMI Physics
Centre National de la Recherche Scientifique (FR), Université de Bourgogne (FR), Centre Georges François Leclerc (FR)
Openalex Percentile: Top 6%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.