Impact of the ICRU/ICRP proposal on operational quantities during interventional procedures: Investigation through spectral measurements

In low-energy photon fields, such as those encountered during fluoroscopy-guided interventions, the current operational quantity for area dosimetry, ambient dose equivalent H*(10), overestimates effective dose by roughly a factor of two. ICRU Report 95 introduces a revised set of operational quantities intended to better align with protection quantities, replacing H*(10) with ambient dose H* for area monitoring. Experimental data validating these changes in realistic clinical environments remain scarce. This study examines the differences between current and newly proposed operational quantities in scattered radiation fields representative of occupational staff exposure during fluoroscopy-guided procedures. Measurements were performed using a Timepix4 hybrid pixel detector operated simultaneously as an X-ray spectrometer and dosimeter. The detector provides energy-resolved photon fluence measurements convertible into operational dose quantities via published fluence-to-dose conversion coefficients. Measurement campaigns were conducted in a urology theatre and an interventional cardiology (IC) theatre under realistic clinical conditions. Photon spectra were recorded at staff positions representative of standard procedures and converted into current and newly proposed quantities relevant to whole-body, skin, and eye lens exposure. The influence of beam angulation and additional shielding was further investigated in the IC theatre.For whole-body exposure, ratios between H* and H*(10) ranged from 0.52 in urology to 0.59 in IC. Quantities relevant to skin and eye lens exposure remained close to unity, with deviations not exceeding 5% and 8%, respectively. Beam angulation and shielding induced only minor variations in these ratios, below 5%.These results provide the first experimental validation of ICRU Report 95 quantities in interventional workplaces. Whole-body dosimetry is most affected by the transition, while skin and eye lens monitoring should remain largely unchanged. Ambient dosimeters calibrated in H*(10) used in low-energy photon fields may be recalibrated using a single procedure-specific calibration factor, whereas calibration procedures for skin and eye lens dosimeters may remain unchanged.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2026-06-25
DOI
https://doi.org/10.1088/1361-6498/ae8272
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of the ICRU/ICRP proposal on operational quantities during interventional procedures: Investigation through spectral measurements

Nicolas Cherbuin, M. Nowak, L. Gallego Manzano, François Bochud et al.
Journal of Radiological Protection
Radiation Dose and Imaging
article

Impact of the ICRU/ICRP proposal on operational quantities during interventional procedures: Investigation through spectral measurements

Nicolas Cherbuin, M. Nowak, L. Gallego Manzano, François Bochud, Pierre Carbonez, Thomas Otto, J. Damet, T. Genetay
article en

Abstract

In low-energy photon fields, such as those encountered during fluoroscopy-guided interventions, the current operational quantity for area dosimetry, ambient dose equivalent H*(10), overestimates effective dose by roughly a factor of two. ICRU Report 95 introduces a revised set of operational quantities intended to better align with protection quantities, replacing H*(10) with ambient dose H* for area monitoring. Experimental data validating these changes in realistic clinical environments remain scarce. This study examines the differences between current and newly proposed operational quantities in scattered radiation fields representative of occupational staff exposure during fluoroscopy-guided procedures. Measurements were performed using a Timepix4 hybrid pixel detector operated simultaneously as an X-ray spectrometer and dosimeter. The detector provides energy-resolved photon fluence measurements convertible into operational dose quantities via published fluence-to-dose conversion coefficients. Measurement campaigns were conducted in a urology theatre and an interventional cardiology (IC) theatre under realistic clinical conditions. Photon spectra were recorded at staff positions representative of standard procedures and converted into current and newly proposed quantities relevant to whole-body, skin, and eye lens exposure. The influence of beam angulation and additional shielding was further investigated in the IC theatre.For whole-body exposure, ratios between H* and H*(10) ranged from 0.52 in urology to 0.59 in IC. Quantities relevant to skin and eye lens exposure remained close to unity, with deviations not exceeding 5% and 8%, respectively. Beam angulation and shielding induced only minor variations in these ratios, below 5%.These results provide the first experimental validation of ICRU Report 95 quantities in interventional workplaces. Whole-body dosimetry is most affected by the transition, while skin and eye lens monitoring should remain largely unchanged. Ambient dosimeters calibrated in H*(10) used in low-energy photon fields may be recalibrated using a single procedure-specific calibration factor, whereas calibration procedures for skin and eye lens dosimeters may remain unchanged.

Journal of Radiological Protection
European Organization for Nuclear Research (CH)
Affordable and clean energy
Openalex Percentile: Top 6%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.