Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains

ABSTRACT Purpose To design and develop an 8‐channel transmit 32‐channel receive 11.7T head coil for the strongest human MRI scanner to date, demonstrate safety, and evaluate signal‐to‐noise ratio (SNR) gains compared to 7T, the most used ultra‐high‐field (UHF) platform. Methods The transmit array design incorporated a folded‐end radiofrequency (RF) shield to mitigate wave propagation in the scanner bore. The receive array consisted of 32‐overlapped loops constructed on a helmet size‐matched with the industry standard 7T head coil. The transmit array was characterized using B 1 + mapping and MR thermometry, whereas the receive array performance was evaluated by measuring SNR at 7T and 11.7T using a lightbulb‐shaped phantom filled with a tissue equivalent solution. Results Loss due to wave propagation was 4.5% of the input power for the chosen RF shield. The mean magnitude of the complex correlations between the simulated and measured single channel B 1 + maps was 94%, which highlights minimal interaction between the transmit and receive arrays. An SNR gain of three‐fold was achieved in the phantom centre. In the periphery, representing the cortex, the SNR gain was 2.65‐fold, and the average gain over the whole volume was a factor of 2.69. The g‐factors of the 11.7T array were lower than the reference 7T coil due to the distinct coil sensitivities at higher frequency. Conclusion We report substantial SNR gains over the whole volume at 11.7T using a fully overlapped 32‐channel loop array. The central SNR gain of B 0 2.14 agrees with the theoretical prediction, while we also gain supra‐linearly in the periphery.

Authors

Institutions

Publication Details

Journal
Magnetic Resonance in Medicine
Published
2026-06-25
DOI
https://doi.org/10.1002/mrm.70473
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains

G Shajan, Caroline Le Ster, Alexis Amadon, Rüdiger Stirnberg et al.
Magnetic Resonance in Medicine
Advanced MRI Techniques and Applications
article

Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains

G Shajan, Caroline Le Ster, Alexis Amadon, Rüdiger Stirnberg, Vincent Gras, Nicolas Boulant, Son Chu, Paul‐François Gapais, Jeslin Parackal Issac
article en

Abstract

ABSTRACT Purpose To design and develop an 8‐channel transmit 32‐channel receive 11.7T head coil for the strongest human MRI scanner to date, demonstrate safety, and evaluate signal‐to‐noise ratio (SNR) gains compared to 7T, the most used ultra‐high‐field (UHF) platform. Methods The transmit array design incorporated a folded‐end radiofrequency (RF) shield to mitigate wave propagation in the scanner bore. The receive array consisted of 32‐overlapped loops constructed on a helmet size‐matched with the industry standard 7T head coil. The transmit array was characterized using B 1 + mapping and MR thermometry, whereas the receive array performance was evaluated by measuring SNR at 7T and 11.7T using a lightbulb‐shaped phantom filled with a tissue equivalent solution. Results Loss due to wave propagation was 4.5% of the input power for the chosen RF shield. The mean magnitude of the complex correlations between the simulated and measured single channel B 1 + maps was 94%, which highlights minimal interaction between the transmit and receive arrays. An SNR gain of three‐fold was achieved in the phantom centre. In the periphery, representing the cortex, the SNR gain was 2.65‐fold, and the average gain over the whole volume was a factor of 2.69. The g‐factors of the 11.7T array were lower than the reference 7T coil due to the distinct coil sensitivities at higher frequency. Conclusion We report substantial SNR gains over the whole volume at 11.7T using a fully overlapped 32‐channel loop array. The central SNR gain of B 0 2.14 agrees with the theoretical prediction, while we also gain supra‐linearly in the periphery.

Magnetic Resonance in Medicine
Centre National de la Recherche Scientifique (FR), Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), Université Paris-Saclay (FR), German Center for Neurodegenerative Diseases (DE), CEA Paris-Saclay (FR), University of Glasgow (GB)
Agence Nationale de la Recherche, Horizon 2020 Framework Programme
Industry, innovation and infrastructure
Openalex Percentile: Top 6%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.