Pulse‑level characterization of low monitor unit deliveries on a modern linear accelerator using a plastic scintillation detector

BACKGROUND: Variability in low monitor unit (MU) deliveries has been previously documented, but the underlying pulse‑level behavior has not been fully described due to the integrating nature of most conventional detectors. A plastic scintillation detector (PSD) enables direct visualization of MU microstructure by isolating individual linac pulses. PURPOSE: To use a high temporal resolution PSD to characterize pulse‑level MU substructure on a modern linac and quantify how variation in integral pulse magnitude and pulse count influence the reproducibility of low MU and fractional MU deliveries. METHODS: A Blue Physics Model 11 PSD was used to measure response from individual linac pulses of a Varian TrueBeam operating at energies of 6 MV, 6 MV FFF, 10 MV, and 10 MV FFF. Initial measurements benchmarked reproducibility in pulse count and dose per pulse by delivering 100 MU at a constant dose rate of 400 MU/min. Subsequent measurements focused on low MU cases, delivering 1-3 MU at dose rates of 5-2400 MU/min. RESULTS: The dose per pulse is variable during a single pulse train with a measured coefficient of variation (COV) of up to 13.3%. However, the average dose per pulse across multiple deliveries is stable (< 0.33% COV). This enables highly reproducible readings (< 0.23% COV) for moderate to high MU deliveries. The variation in dose per pulse coupled with a lower number of required pulses leads to challenges when delivering a low number of MUs with FFF beams. Discrepancies in the planned versus delivered MUs were observed on both detectors and the treatment delivery system. CONCLUSIONS: Time-resolved analysis shows that dose per pulse fluctuations and the limited number of pulses comprising low MU FFF deliveries lead to measurable variability in delivered MU. These findings characterize the pulsed MU substructure on a modern linac, in which each MU is composed of a finite number of discrete radiation pulses, and quantify the achievable precision of fractional MU delivery, which is fundamentally limited by the dose per pulse (∼0.1 MU for TrueBeam FFF beams).

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-06-25
DOI
https://doi.org/10.1002/acm2.70670
Primary Topic
Radiation Effects and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pulse‑level characterization of low monitor unit deliveries on a modern linear accelerator using a plastic scintillation detector

Andrew J. White, P Yadunandan Yadav, ahtesham U. Khan, Indra J. Das
Journal of Applied Clinical Medical Physics
Radiation Effects and Dosimetry
article

Pulse‑level characterization of low monitor unit deliveries on a modern linear accelerator using a plastic scintillation detector

Andrew J. White, P Yadunandan Yadav, ahtesham U. Khan, Indra J. Das
article en

Abstract

BACKGROUND: Variability in low monitor unit (MU) deliveries has been previously documented, but the underlying pulse‑level behavior has not been fully described due to the integrating nature of most conventional detectors. A plastic scintillation detector (PSD) enables direct visualization of MU microstructure by isolating individual linac pulses. PURPOSE: To use a high temporal resolution PSD to characterize pulse‑level MU substructure on a modern linac and quantify how variation in integral pulse magnitude and pulse count influence the reproducibility of low MU and fractional MU deliveries. METHODS: A Blue Physics Model 11 PSD was used to measure response from individual linac pulses of a Varian TrueBeam operating at energies of 6 MV, 6 MV FFF, 10 MV, and 10 MV FFF. Initial measurements benchmarked reproducibility in pulse count and dose per pulse by delivering 100 MU at a constant dose rate of 400 MU/min. Subsequent measurements focused on low MU cases, delivering 1-3 MU at dose rates of 5-2400 MU/min. RESULTS: The dose per pulse is variable during a single pulse train with a measured coefficient of variation (COV) of up to 13.3%. However, the average dose per pulse across multiple deliveries is stable (< 0.33% COV). This enables highly reproducible readings (< 0.23% COV) for moderate to high MU deliveries. The variation in dose per pulse coupled with a lower number of required pulses leads to challenges when delivering a low number of MUs with FFF beams. Discrepancies in the planned versus delivered MUs were observed on both detectors and the treatment delivery system. CONCLUSIONS: Time-resolved analysis shows that dose per pulse fluctuations and the limited number of pulses comprising low MU FFF deliveries lead to measurable variability in delivered MU. These findings characterize the pulsed MU substructure on a modern linac, in which each MU is composed of a finite number of discrete radiation pulses, and quantify the achievable precision of fractional MU delivery, which is fundamentally limited by the dose per pulse (∼0.1 MU for TrueBeam FFF beams).

Journal of Applied Clinical Medical PhysicsVol. 27(7)
Northwestern University (US), University of Wisconsin–Madison (US), Northwestern Memorial Hospital (US), Wisconsin Division of Public Health (US)
Openalex Percentile: Top 9%
Radiation Effects and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.