Feasibility of opportunistic peripheral bone mineral density and structural quantification using an ultra-low dose multi-detector CT

Quantitative computed tomography (QCT) enables accurate assessment of volumetric bone mineral density (vBMD) and bone structure. Peripheral quantitative computed tomography (pQCT) and high-resolution peripheral quantitative computed tomography (HR-pQCT) have been proposed as lower-dose alternatives to conventional QCT; however, their limited availability restricts widespread clinical and research use. To investigate the feasibility of using ultra-low-dose multi-detector computed tomography (MDCT) with a hybrid iterative reconstruction (HIR) algorithm for quantifying tibial vBMD, using synchronous liquid and solid calibration phantoms. Fifty healthy volunteers (28 women and 22 men; age range: 30-60 yr; Tscore > -1) were scanned using a Philips-Brilliance MDCT scanner. To minimize radiation exposure, vBMD was quantified at the 38% distal tibial site using either a liquid calibration phantom (50, 100, and 200 mg/cm3 dipotassium hydrogen phosphate [K2HPO4]) or a solid calibration phantom (50, 100, and 250 mg/cm3 calcium hydroxyapatite [CHA]). A cross-calibration equation was applied to derive equivalent vBMD values between phantom types. Cortical bone structure was evaluated using a semi-automated segmentation approach. Following cross-calibration, BMDK2HPO4 = 1.34 BMDCHA + 13.43, mean cortical vBMD values were 1188.07 ± 29.38 mg/cm3 (CHA) and 1600.92 ± 39.51 mg/cm3 (K2HPO4). The CHA-based cortical vBMD values were consistent with solid-based pCT measurements reported in the literature. Mean cortical thickness was approximately 5 mm. The estimated effective radiation dose is approximately 10µSv, corresponding to daily background radiation levels. This preliminary study demonstrates the feasibility of using HIR-based ultra-low-dose MDCT for peripheral tibial vBMD and structural assessment at the 38% distal site. This approach offers a promising opportunistic alternative in clinical and research settings where pQCT or HR-pQCT systems are unavailable.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-06-25
DOI
https://doi.org/10.1088/2057-1976/ae826c
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of opportunistic peripheral bone mineral density and structural quantification using an ultra-low dose multi-detector CT

Elmira Yazdani, Nazanin Mobini, Malakeh Malekzadeh, Joyce H. Keyak et al.
Biomedical Physics & Engineering Express
Dental Radiography and Imaging
article

Feasibility of opportunistic peripheral bone mineral density and structural quantification using an ultra-low dose multi-detector CT

Elmira Yazdani, Nazanin Mobini, Malakeh Malekzadeh, Joyce H. Keyak, Shahrokh Abbasi‐Rad, Mohammad Bagher Shiran, Navid Tofighirad, Hamidreza Saligheh Rad, Mahdi Shamsi, Ali Abbasian Ardakanii
article en

Abstract

Quantitative computed tomography (QCT) enables accurate assessment of volumetric bone mineral density (vBMD) and bone structure. Peripheral quantitative computed tomography (pQCT) and high-resolution peripheral quantitative computed tomography (HR-pQCT) have been proposed as lower-dose alternatives to conventional QCT; however, their limited availability restricts widespread clinical and research use. To investigate the feasibility of using ultra-low-dose multi-detector computed tomography (MDCT) with a hybrid iterative reconstruction (HIR) algorithm for quantifying tibial vBMD, using synchronous liquid and solid calibration phantoms. Fifty healthy volunteers (28 women and 22 men; age range: 30-60 yr; Tscore > -1) were scanned using a Philips-Brilliance MDCT scanner. To minimize radiation exposure, vBMD was quantified at the 38% distal tibial site using either a liquid calibration phantom (50, 100, and 200 mg/cm3 dipotassium hydrogen phosphate [K2HPO4]) or a solid calibration phantom (50, 100, and 250 mg/cm3 calcium hydroxyapatite [CHA]). A cross-calibration equation was applied to derive equivalent vBMD values between phantom types. Cortical bone structure was evaluated using a semi-automated segmentation approach. Following cross-calibration, BMDK2HPO4 = 1.34 BMDCHA + 13.43, mean cortical vBMD values were 1188.07 ± 29.38 mg/cm3 (CHA) and 1600.92 ± 39.51 mg/cm3 (K2HPO4). The CHA-based cortical vBMD values were consistent with solid-based pCT measurements reported in the literature. Mean cortical thickness was approximately 5 mm. The estimated effective radiation dose is approximately 10µSv, corresponding to daily background radiation levels. This preliminary study demonstrates the feasibility of using HIR-based ultra-low-dose MDCT for peripheral tibial vBMD and structural assessment at the 38% distal site. This approach offers a promising opportunistic alternative in clinical and research settings where pQCT or HR-pQCT systems are unavailable.

Biomedical Physics & Engineering Express
Danube Private University (AT), Universität für Weiterbildung Krems (AT), Harvard University (US), Iran University of Medical Sciences (IR), Islamic Azad University Central Tehran Branch (IR), UC Irvine Health (US), La Salle University (US), Tehran University of Medical Sciences (IR)
Iran University of Medical Sciences
Good health and well-being
Openalex Percentile: Top 6%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.