Physics‐informed patient‐specific calibration for monocular visual‐based dose reconstruction in breast cancer radiotherapy

BACKGROUND: Interfractional anatomical variations during breast cancer radiotherapy can significantly deviate the delivered dose from the treatment plan. Standard image guidance via Cone-Beam CT (CBCT) is typically sparse due to radiation and workflow constraints, leaving daily geometric changes unmonitored. PURPOSE: To bridge this gap, we propose a physics-informed framework that utilizes monocular surface vision for continuous, non-ionizing 3D dose reconstruction. METHODS: The system utilizes the weekly CBCT scans from the first 2 weeks of treatment (Week 1 and Week 2) for patient-specific calibration. Surface features are extracted via a hybrid HOG-CNN descriptor, while 3D deformation vector fields (DVFs) are compressed into a low-dimensional latent space via Principal Component Analysis (PCA). A linear mapping from surface topography to PCA coefficients is optimized using physics-informed Tikhonov regularization, incorporating an eigenvalue-based penalty matrix to suppress high-order non-physical modes. Post-calibration, the model reconstructs 3D dose distributions for subsequent fractions in real time using only monocular input. RESULTS: Validated on 29 patients (87 fractions), the framework achieved a deformation coefficient correlation of R = 0.796. Reconstructed doses yielded a mean Gamma passing rate (3%/3 mm) of 93.8% ± 3.1% and a dose correlation of 0.940 ± 0.037. Ablation studies demonstrated that removing physics constraints or PCA reduction significantly degraded performance (ΔGPR = -8.5% and -22.1%, respectively; p < 0.001). End-to-end latency was 42 ± 5 ms per fraction, representing a ∼124-fold acceleration over commercial deformable registration solutions. CONCLUSIONS: This framework transforms routine weekly CBCTs into robust "calibration anchors" for continuous visual dose monitoring. It provides clinical-grade accuracy with negligible computational overhead and no additional ionizing dose, offering a practical solution for real-time adaptive radiotherapy.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-06-25
DOI
https://doi.org/10.1002/mp.70533
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Physics‐informed patient‐specific calibration for monocular visual‐based dose reconstruction in breast cancer radiotherapy

Ruiwan Chen, Xuecen Wang, Y F Deng
Medical Physics
Advanced Radiotherapy Techniques
article

Physics‐informed patient‐specific calibration for monocular visual‐based dose reconstruction in breast cancer radiotherapy

Ruiwan Chen, Xuecen Wang, Y F Deng
article en

Abstract

BACKGROUND: Interfractional anatomical variations during breast cancer radiotherapy can significantly deviate the delivered dose from the treatment plan. Standard image guidance via Cone-Beam CT (CBCT) is typically sparse due to radiation and workflow constraints, leaving daily geometric changes unmonitored. PURPOSE: To bridge this gap, we propose a physics-informed framework that utilizes monocular surface vision for continuous, non-ionizing 3D dose reconstruction. METHODS: The system utilizes the weekly CBCT scans from the first 2 weeks of treatment (Week 1 and Week 2) for patient-specific calibration. Surface features are extracted via a hybrid HOG-CNN descriptor, while 3D deformation vector fields (DVFs) are compressed into a low-dimensional latent space via Principal Component Analysis (PCA). A linear mapping from surface topography to PCA coefficients is optimized using physics-informed Tikhonov regularization, incorporating an eigenvalue-based penalty matrix to suppress high-order non-physical modes. Post-calibration, the model reconstructs 3D dose distributions for subsequent fractions in real time using only monocular input. RESULTS: Validated on 29 patients (87 fractions), the framework achieved a deformation coefficient correlation of R = 0.796. Reconstructed doses yielded a mean Gamma passing rate (3%/3 mm) of 93.8% ± 3.1% and a dose correlation of 0.940 ± 0.037. Ablation studies demonstrated that removing physics constraints or PCA reduction significantly degraded performance (ΔGPR = -8.5% and -22.1%, respectively; p < 0.001). End-to-end latency was 42 ± 5 ms per fraction, representing a ∼124-fold acceleration over commercial deformable registration solutions. CONCLUSIONS: This framework transforms routine weekly CBCTs into robust "calibration anchors" for continuous visual dose monitoring. It provides clinical-grade accuracy with negligible computational overhead and no additional ionizing dose, offering a practical solution for real-time adaptive radiotherapy.

Medical PhysicsVol. 53(7)
Sun Yat-sen University (CN), The First Affiliated Hospital, Sun Yat-sen University (CN)
Good health and well-being
Openalex Percentile: Top 7%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.