Benchmarking Lipid Quantitation from Skin Swab Sebum-Rich Samples to Establish Mass Spectrometry-Based Diagnostic Methodology

Skin swab sampling provides an accessible, noninvasive method for collecting lipid-rich material from the skin surface, with growing potential for biomarker discovery and diagnostic testing. However, quantitative sebum analysis remains challenging because of variability in the amount collected and the complexity of the lipid matrix. This study evaluates the robustness of skin swab sampling for lipid analysis and the feasibility of quantifying skin-surface components using liquid chromatography–tandem mass spectrometry (LC–MS/MS). Multiple reaction monitoring was used to assess analytical linearity, matrix effects, and recovery following established protocols for swab collection, storage, and extraction. Robust linearity was observed for both spiked isotope-labeled internal standards (IS) and endogenous lipids across concentrations of 0.15–550 ng/mL ( R 2 > 0.97). IS recovery varied between lipid classes, with consistent under-recovery. Although IS corrected for analytical variability during extraction and analysis, it could not fully account for differences in the amount of material collected. Endogenous lipids were, therefore, evaluated as normalization markers. Ratios between selected triacylglycerol (TG) species were highly conserved across samples, with coefficients of variation below 15% for TG 45:1/TG 47:1, TG 46:1/TG 48:1, and TG 47:1/TG 48:1. These ratios were stable across individuals and were independent of disease status. The reproducibility of TG recovery and stability of TG ratios suggest that endogenous lipid normalization can mitigate sampling and extraction variability. This approach provides a framework for improving quantitative reliability in skin swab lipidomics and represents a big step toward the development of reproducible mass spectrometry workflows for skin-swat-based biomarker studies.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2026-06-24
DOI
https://doi.org/10.1021/acs.analchem.5c08056
Primary Topic
Metabolomics and Mass Spectrometry Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Benchmarking Lipid Quantitation from Skin Swab Sebum-Rich Samples to Establish Mass Spectrometry-Based Diagnostic Methodology

Eleanor Sinclair, Michael Morris, Katherine A. Hollywood, Caitlin Walton‐Doyle et al.
Analytical Chemistry
Metabolomics and Mass Spectrometry Studies
article

Benchmarking Lipid Quantitation from Skin Swab Sebum-Rich Samples to Establish Mass Spectrometry-Based Diagnostic Methodology

Eleanor Sinclair, Michael Morris, Katherine A. Hollywood, Caitlin Walton‐Doyle, Perdita E. Barran, Monty Silverdale, Minhui Zhu
article en

Abstract

Skin swab sampling provides an accessible, noninvasive method for collecting lipid-rich material from the skin surface, with growing potential for biomarker discovery and diagnostic testing. However, quantitative sebum analysis remains challenging because of variability in the amount collected and the complexity of the lipid matrix. This study evaluates the robustness of skin swab sampling for lipid analysis and the feasibility of quantifying skin-surface components using liquid chromatography–tandem mass spectrometry (LC–MS/MS). Multiple reaction monitoring was used to assess analytical linearity, matrix effects, and recovery following established protocols for swab collection, storage, and extraction. Robust linearity was observed for both spiked isotope-labeled internal standards (IS) and endogenous lipids across concentrations of 0.15–550 ng/mL ( R 2 > 0.97). IS recovery varied between lipid classes, with consistent under-recovery. Although IS corrected for analytical variability during extraction and analysis, it could not fully account for differences in the amount of material collected. Endogenous lipids were, therefore, evaluated as normalization markers. Ratios between selected triacylglycerol (TG) species were highly conserved across samples, with coefficients of variation below 15% for TG 45:1/TG 47:1, TG 46:1/TG 48:1, and TG 47:1/TG 48:1. These ratios were stable across individuals and were independent of disease status. The reproducibility of TG recovery and stability of TG ratios suggest that endogenous lipid normalization can mitigate sampling and extraction variability. This approach provides a framework for improving quantitative reliability in skin swab lipidomics and represents a big step toward the development of reproducible mass spectrometry workflows for skin-swat-based biomarker studies.

Analytical Chemistry
Waters (United States) (US), Salford Royal NHS Foundation Trust (GB), University of Manchester (GB), Czech Academy of Sciences, Institute of Biotechnology (CZ), IONICS Mass Spectrometry (Canada) (CA)
Openalex Percentile: Top 10%
Metabolomics and Mass Spectrometry Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.