Technical note: Clinical real-time multi-target, 3D-motion tracking radiation delivery on a magnetic resonance linear accelerator

BACKGROUND: 
The purpose of this study was to demonstrate the feasibility of multi-target 3D-motion tracking for use in magnetic resonance-imaging (MRI)-guided, stereotactic body radiotherapy for upper abdominal targets. This paper focuses on the clinical testing of a real-time 3D-motion tracking process for both clinical targets and organs-at risk-for automatic beam gating in MRI-guided radiation delivery (MRgRT).
Patient and methods:
A pilot study was performed using a commissioned workflow based on an in-house dual-target phantom. A patient with primary liver cancer in segment 5 with adjacent bowel was selected for dual-plane tracking, two target tracking, and OAR dose-accumulation for auto-beam gating. Data from all five delivery fractions were reviewed and presented, including cine MRI settings, target tracking settings, respiratory motion management, and delivery details.
Results: The initial delivery time for the first fraction was 35 minutes, substantially longer than the expected duration of 11.8 minutes, which was determined based on a 50% delivery time using the calculated monitor units (MU). Subsequent treatments were consistently faster, with the last fraction being 9 minutes. For dual-plane tracking, the clinical team defined tracking parameters based on the motion and location of the secondary tracking structure (critical organ: large bowel) in relation to the tumor, using a 6 mm boundary expansion. This was compared to a tighter 3 mm boundary expansion for the primary target. 
Conclusions: We demonstrated the clinical feasibility of dual-plane tracking in MRgRT which enables multi-target, 3D-motion tracking of both targets and organs at risk. This approach can provide an additional motion‑monitoring capability beyond conventional single‑plane tracking while maintaining effective target gating and treatment time. This initial institutional implementation demonstrates the technical feasibility of dual‑plane tracking and its potential to incorporate motion monitoring of selected critical structures in MRgRT.&#xD.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-06-24
DOI
https://doi.org/10.1088/2057-1976/ae81bc
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Technical note: Clinical real-time multi-target, 3D-motion tracking radiation delivery on a magnetic resonance linear accelerator

Emily McHugh, Minsol Kim, Eric Laugeman, Allen Mo et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Technical note: Clinical real-time multi-target, 3D-motion tracking radiation delivery on a magnetic resonance linear accelerator

Emily McHugh, Minsol Kim, Eric Laugeman, Allen Mo, Taeho Kim, Michael Waters, Justin Ross, Carl DeSelm, Austen Nathaniel Curcuru, Marissa Maxwell
article en

Abstract

BACKGROUND: 
The purpose of this study was to demonstrate the feasibility of multi-target 3D-motion tracking for use in magnetic resonance-imaging (MRI)-guided, stereotactic body radiotherapy for upper abdominal targets. This paper focuses on the clinical testing of a real-time 3D-motion tracking process for both clinical targets and organs-at risk-for automatic beam gating in MRI-guided radiation delivery (MRgRT).
Patient and methods:
A pilot study was performed using a commissioned workflow based on an in-house dual-target phantom. A patient with primary liver cancer in segment 5 with adjacent bowel was selected for dual-plane tracking, two target tracking, and OAR dose-accumulation for auto-beam gating. Data from all five delivery fractions were reviewed and presented, including cine MRI settings, target tracking settings, respiratory motion management, and delivery details.
Results: The initial delivery time for the first fraction was 35 minutes, substantially longer than the expected duration of 11.8 minutes, which was determined based on a 50% delivery time using the calculated monitor units (MU). Subsequent treatments were consistently faster, with the last fraction being 9 minutes. For dual-plane tracking, the clinical team defined tracking parameters based on the motion and location of the secondary tracking structure (critical organ: large bowel) in relation to the tumor, using a 6 mm boundary expansion. This was compared to a tighter 3 mm boundary expansion for the primary target. 
Conclusions: We demonstrated the clinical feasibility of dual-plane tracking in MRgRT which enables multi-target, 3D-motion tracking of both targets and organs at risk. This approach can provide an additional motion‑monitoring capability beyond conventional single‑plane tracking while maintaining effective target gating and treatment time. This initial institutional implementation demonstrates the technical feasibility of dual‑plane tracking and its potential to incorporate motion monitoring of selected critical structures in MRgRT.&#xD.

Biomedical Physics & Engineering Express
BJC HealthCare (US), Washington University in St. Louis (US), Hadassah Medical Center (IL)
Good health and well-being
Openalex Percentile: Top 7%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.