The repeatability and comparison between swept-source optical coherence tomography and Scheimpflug imaging in healthy eyes and various grades of keratoconus

To evaluate the repeatability of a swept-source anterior segment optical coherence tomography (SS-OCT) and a high-resolution rotating Scheimpflug system and compare them in patients with various grades of keratoconus. In this prospective, cross-sectional study, 117 eyes (86 keratoconus and 31 healthy controls) with various grades of keratoconus were enrolled during their regular follow-up visit. Three consecutive measurements were taken with SS-OCT (CASIA SS-1000) and Scheimpflug device (Pentacam HR) by the same operator. Anterior and posterior keratometry, elevation, pachymetry, pupil diameter, and Q-value were recorded. Raw data from both devices were used to determine cone location and size. Repeatability was assessed via relative standard deviation (SD), and compared them using Pearson correlation and Bland-Altman analysis. In healthy eyes, repeatability was comparable for most parameters, though SS-OCT showed significantly better repeatability for pachymetry, pupil diameter, and posterior keratometry (p < 0.001). In keratoconus eyes, SS-OCT demonstrated better repeatability in posterior keratometry and pachymetry. Despite strong correlations between devices (r ≥ 0.90 for most parameters), systematic differences were present in almost all measurements, including Kmax (mean difference 4.57 ± 1.68 D), with wide limits of agreement. Cone location also differed significantly between devices (p < 0.001). We found significantly better overall repeatability of keratometry and pachymetry with the SS-OCT in keratoconus eyes, these parameters are essential for progression monitoring and treatment planning. Furthermore, the devices cannot be used interchangeably. Awareness of these differences is essential for interpreting measurements.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-06-24
DOI
https://doi.org/10.1038/s41598-026-59595-4
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The repeatability and comparison between swept-source optical coherence tomography and Scheimpflug imaging in healthy eyes and various grades of keratoconus

Tos T. J. M. Berendschot, Magali M.S. Vandevenne, Rudy M. M. A. Nuijts, Alle Boonstra
Scientific Reports
Corneal surgery and disorders
article

The repeatability and comparison between swept-source optical coherence tomography and Scheimpflug imaging in healthy eyes and various grades of keratoconus

Tos T. J. M. Berendschot, Magali M.S. Vandevenne, Rudy M. M. A. Nuijts, Alle Boonstra
article en

Abstract

To evaluate the repeatability of a swept-source anterior segment optical coherence tomography (SS-OCT) and a high-resolution rotating Scheimpflug system and compare them in patients with various grades of keratoconus. In this prospective, cross-sectional study, 117 eyes (86 keratoconus and 31 healthy controls) with various grades of keratoconus were enrolled during their regular follow-up visit. Three consecutive measurements were taken with SS-OCT (CASIA SS-1000) and Scheimpflug device (Pentacam HR) by the same operator. Anterior and posterior keratometry, elevation, pachymetry, pupil diameter, and Q-value were recorded. Raw data from both devices were used to determine cone location and size. Repeatability was assessed via relative standard deviation (SD), and compared them using Pearson correlation and Bland-Altman analysis. In healthy eyes, repeatability was comparable for most parameters, though SS-OCT showed significantly better repeatability for pachymetry, pupil diameter, and posterior keratometry (p < 0.001). In keratoconus eyes, SS-OCT demonstrated better repeatability in posterior keratometry and pachymetry. Despite strong correlations between devices (r ≥ 0.90 for most parameters), systematic differences were present in almost all measurements, including Kmax (mean difference 4.57 ± 1.68 D), with wide limits of agreement. Cone location also differed significantly between devices (p < 0.001). We found significantly better overall repeatability of keratometry and pachymetry with the SS-OCT in keratoconus eyes, these parameters are essential for progression monitoring and treatment planning. Furthermore, the devices cannot be used interchangeably. Awareness of these differences is essential for interpreting measurements.

Scientific Reports
Maastricht University Medical Centre (NL), Maastricht University (NL)
Openalex Percentile: Top 6%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.