Performance Evaluation of the BC-6800Plus Hematology Analyzer for ESR Measurement: Comparison with the Westergren Method and TEST1

Background: Automated erythrocyte sedimentation rate (ESR) measurement streamlines laboratory workflow and shortens turnaround time.This study evaluated the analytical performance of the Mindray BC-6800Plus (BC-6800P, Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China), a fully automated hematology analyzer with integrated ESR testing capability, and compared its performance with those of the Westergren reference method and the TEST1 analyzer (Alifax S.r.l., Polverara, Italy) in a Korean clinical laboratory setting.Methods: Sixty residual EDTA whole-blood specimens were analyzed by all three methods within 4 hours of collection.The analytical performance of the BC-6800P, including precision, carryover, and reference interval verification, was evaluated in accordance with CLSI EP15-A3 and International Council for Standardization in Haematology (ICSH) recommendations.Agreement between methods was assessed using Passing-Bablok regression, Bland-Altman analysis, Spearman's rank correlation, and range-stratified subgroup analyses.Results: The BC-6800P exhibited excellent precision (intra-run CV ≤ 8.2%, inter-run CV ≤ 1.9%) and negligible carryover (-0.18%).All reference interval verification specimens fell within the sex-specific manufacturer-provided intervals.Comparison with the Westergren method demonstrated strong overall agreement (ρ=0.891,P <0.001) with a small negative bias (-3.61 mm/h).However, absolute bias increased markedly at higher ESR values, rising from 3.5 mm/h at ESR 0-20 to 18.2 mm/h at ESR >60.Comparison with TEST1 showed substantial method-dependent disagreement (proportional bias; limits of agreement, -43.1 to 33.6 mm/h), consistent with platform-dependent measurement differences also observed in the TEST1-Westergren comparison.Conclusions: The BC-6800P demonstrated acceptable analytical performance for routine ESR testing.However, the marked increase in absolute bias at higher ESR values necessitates cautious clinical interpretation, and results from different automated platforms should not be considered interchangeable, particularly for longitudinal patient monitoring.

Authors

Institutions

Publication Details

Journal
Laboratory Medicine Online
Published
2026-06-25
DOI
https://doi.org/10.47429/lmo.2026.16.3.185
Primary Topic
Electron Spin Resonance Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Evaluation of the BC-6800Plus Hematology Analyzer for ESR Measurement: Comparison with the Westergren Method and TEST1

H. Son, Jong Ho Lee
Laboratory Medicine Online
Electron Spin Resonance Studies
article

Performance Evaluation of the BC-6800Plus Hematology Analyzer for ESR Measurement: Comparison with the Westergren Method and TEST1

H. Son, Jong Ho Lee
article en

Abstract

Background: Automated erythrocyte sedimentation rate (ESR) measurement streamlines laboratory workflow and shortens turnaround time.This study evaluated the analytical performance of the Mindray BC-6800Plus (BC-6800P, Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China), a fully automated hematology analyzer with integrated ESR testing capability, and compared its performance with those of the Westergren reference method and the TEST1 analyzer (Alifax S.r.l., Polverara, Italy) in a Korean clinical laboratory setting.Methods: Sixty residual EDTA whole-blood specimens were analyzed by all three methods within 4 hours of collection.The analytical performance of the BC-6800P, including precision, carryover, and reference interval verification, was evaluated in accordance with CLSI EP15-A3 and International Council for Standardization in Haematology (ICSH) recommendations.Agreement between methods was assessed using Passing-Bablok regression, Bland-Altman analysis, Spearman's rank correlation, and range-stratified subgroup analyses.Results: The BC-6800P exhibited excellent precision (intra-run CV ≤ 8.2%, inter-run CV ≤ 1.9%) and negligible carryover (-0.18%).All reference interval verification specimens fell within the sex-specific manufacturer-provided intervals.Comparison with the Westergren method demonstrated strong overall agreement (ρ=0.891,P <0.001) with a small negative bias (-3.61 mm/h).However, absolute bias increased markedly at higher ESR values, rising from 3.5 mm/h at ESR 0-20 to 18.2 mm/h at ESR >60.Comparison with TEST1 showed substantial method-dependent disagreement (proportional bias; limits of agreement, -43.1 to 33.6 mm/h), consistent with platform-dependent measurement differences also observed in the TEST1-Westergren comparison.Conclusions: The BC-6800P demonstrated acceptable analytical performance for routine ESR testing.However, the marked increase in absolute bias at higher ESR values necessitates cautious clinical interpretation, and results from different automated platforms should not be considered interchangeable, particularly for longitudinal patient monitoring.

Laboratory Medicine OnlineVol. 16(3)
Yeungnam University Medical Center (KR), Yeungnam University College, Yeungnam University (KR)
Openalex Percentile: Top 7%
Electron Spin Resonance Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.