Performance of Sysmex SP-50 automated slidemakers and stainers in three hematology laboratories

Manual differential count and morphological evaluation of leukocytes remain the gold standard for white blood cell differential (WBC DIFF). However, the accuracy of manual WBC DIFF depends on the quality of the blood slide and the observer's expertise. Nowadays, automated slide preparation and staining systems, such as Sysmex SP-50, have been introduced to improve the quality of this method. In this study, we evaluated the quality of peripheral blood smears prepared and stained using the SP-50 automated system, with a particular focus on variables affecting cell distribution and morphology. This study evaluates the performance of five Sysmex SP-50 systems, from analyzer installation to final verification. The verification protocol included a preliminary performance check of all analyzer setups, followed by a conclusive verification. Conclusive verification was initiated after the performance checks were considered acceptable by laboratory professionals. The results demonstrated cell maldistribution across the film and the presence of smudge cells, both of which were linked to the pressure and height/angle of the spreader glass. In addition, inadequate staining of the blood film was observed, which was influenced by staining conditions and the use of fans. Together, these findings highlight critical quality issues with blood smear preparation that compromise accurate morphological analysis. The results suggest that while automation may offer some advantages, the quality and reliability of Sysmex SP-50-generated smears need laboratory-specific adjustment to meet the quality standards required for diagnostic evaluation.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2026-06-24
DOI
https://doi.org/10.1080/00365513.2026.2686252
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of Sysmex SP-50 automated slidemakers and stainers in three hematology laboratories

Johanna Kultti, Anni-Mari Tuori, Eeva-Liisa Paattiniemi, Päivi Helminen-Pacius et al.
Scandinavian Journal of Clinical and Laboratory Investigation
Clinical Laboratory Practices and Quality Control
article

Performance of Sysmex SP-50 automated slidemakers and stainers in three hematology laboratories

Johanna Kultti, Anni-Mari Tuori, Eeva-Liisa Paattiniemi, Päivi Helminen-Pacius, Kari Pulkki
article en

Abstract

Manual differential count and morphological evaluation of leukocytes remain the gold standard for white blood cell differential (WBC DIFF). However, the accuracy of manual WBC DIFF depends on the quality of the blood slide and the observer's expertise. Nowadays, automated slide preparation and staining systems, such as Sysmex SP-50, have been introduced to improve the quality of this method. In this study, we evaluated the quality of peripheral blood smears prepared and stained using the SP-50 automated system, with a particular focus on variables affecting cell distribution and morphology. This study evaluates the performance of five Sysmex SP-50 systems, from analyzer installation to final verification. The verification protocol included a preliminary performance check of all analyzer setups, followed by a conclusive verification. Conclusive verification was initiated after the performance checks were considered acceptable by laboratory professionals. The results demonstrated cell maldistribution across the film and the presence of smudge cells, both of which were linked to the pressure and height/angle of the spreader glass. In addition, inadequate staining of the blood film was observed, which was influenced by staining conditions and the use of fans. Together, these findings highlight critical quality issues with blood smear preparation that compromise accurate morphological analysis. The results suggest that while automation may offer some advantages, the quality and reliability of Sysmex SP-50-generated smears need laboratory-specific adjustment to meet the quality standards required for diagnostic evaluation.

Scandinavian Journal of Clinical and Laboratory Investigation
University of Helsinki (FI), Helsinki University Hospital (FI), South Karelia Central Hospital (FI), Northern Ostrobothnia Hospital District (FI)
Openalex Percentile: Top 8%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.