Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung

Background: Contemporary pediatric high-frequency oscillatory ventilators may differ in how set oscillatory parameters are translated into delivered tidal volume under varying respiratory mechanical loads.We evaluated determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung using a device-specific full-factorial bench model.Methods: A Fabian high-frequency oscillatory oscillator was connected to an IngMar pediatric mechanical test lung through a standard dry pediatric circuit.Tidal volume was measured with a Florian respiratory monitor placed between the circuit and lung model.We tested 1,512 combinations of frequency, pressure amplitude, mean airway pressure, inspiratory-to-expiratory ratio, compliance, and resistance.Measurements were recorded after at least 2 minutes of steady state at each condition.Analysis of variance and multiple linear regression were used to evaluate associations with delivered tidal volume.Results: Delivered tidal volume was associated with frequency, pressure amplitude, mean airway pressure, inspiratory-to-expiratory ratio, and resistance (all P<0.001), but not compliance across the tested range (P=0.18).Tidal volume increased with higher pressure amplitude and mean airway pressure and decreased with higher frequency and resistance.In the regression model, tidal volume (mL)=25.9-(1.3×frequency)+(0.14×pressureamplitude)+(0.08×meanairway pressure)-(0.18×resistance).The model explained 80.8% of the variance in delivered tidal volume (R 2 =0.808, adjusted R 2 =0.808).Conclusions: In this pediatric mechanical test lung, delivered tidal volume during high-frequency oscillatory ventilation reflected both oscillator settings and respiratory load.Lower frequency and higher pressure amplitude produced the largest delivered volumes, whereas higher resistance reduced volume delivery.These findings are device-specific and should be interpreted as bench physiology rather than evidence of clinical outcome benefit.

Authors

Institutions

Publication Details

Journal
Archives of Pediatric Critical Care
Published
2026-06-24
DOI
https://doi.org/10.32990/apcc.2026.00038
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung

Weerapong Lilitwat, Yuttiwat Vorakunthada
Archives of Pediatric Critical Care
Respiratory Support and Mechanisms
article

Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung

Weerapong Lilitwat, Yuttiwat Vorakunthada
article en

Abstract

Background: Contemporary pediatric high-frequency oscillatory ventilators may differ in how set oscillatory parameters are translated into delivered tidal volume under varying respiratory mechanical loads.We evaluated determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung using a device-specific full-factorial bench model.Methods: A Fabian high-frequency oscillatory oscillator was connected to an IngMar pediatric mechanical test lung through a standard dry pediatric circuit.Tidal volume was measured with a Florian respiratory monitor placed between the circuit and lung model.We tested 1,512 combinations of frequency, pressure amplitude, mean airway pressure, inspiratory-to-expiratory ratio, compliance, and resistance.Measurements were recorded after at least 2 minutes of steady state at each condition.Analysis of variance and multiple linear regression were used to evaluate associations with delivered tidal volume.Results: Delivered tidal volume was associated with frequency, pressure amplitude, mean airway pressure, inspiratory-to-expiratory ratio, and resistance (all P<0.001), but not compliance across the tested range (P=0.18).Tidal volume increased with higher pressure amplitude and mean airway pressure and decreased with higher frequency and resistance.In the regression model, tidal volume (mL)=25.9-(1.3×frequency)+(0.14×pressureamplitude)+(0.08×meanairway pressure)-(0.18×resistance).The model explained 80.8% of the variance in delivered tidal volume (R 2 =0.808, adjusted R 2 =0.808).Conclusions: In this pediatric mechanical test lung, delivered tidal volume during high-frequency oscillatory ventilation reflected both oscillator settings and respiratory load.Lower frequency and higher pressure amplitude produced the largest delivered volumes, whereas higher resistance reduced volume delivery.These findings are device-specific and should be interpreted as bench physiology rather than evidence of clinical outcome benefit.

Archives of Pediatric Critical Care
Texas Tech University (US), Wentworth Douglass Hospital (US), Texas Tech University Health Sciences Center (US)
Openalex Percentile: Top 7%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.