Explainable machine learning for patient‐specific quality assurance in intensity‐modulated radiotherapy based on anatomical structures

BACKGROUND: Patient-specific quality assurance (PSQA) plays a pivotal role in intensity-modulated radiotherapy (IMRT) to ensure accurate dose delivery. However, conventional measurement-based PSQA approaches are labor-intensive and provide limited insight into the underlying factors contributing to variations in gamma passing rates (GPRs). Anatomical characteristics of the planning target volume (PTV) and organs at risk (OARs) may contain predictive information relevant to GPR performance, yet their potential has not been fully explored within interpretable machine learning frameworks. PURPOSE: This study aimed to develop an interpretable machine learning (ML) framework for predicting GPRs in IMRT based on anatomical features extracted from the PTV and OARs. METHODS: A retrospective cohort of 243 clinical chest IMRT plans was analyzed. Radiomic and dosimetric features were extracted for each anatomical structure. Two ML regression models-Random Forest (RF) and eXtreme Gradient Boosting (XGBoost)-were developed to predict GPRs for the PTV and OARs under four gamma criteria (3%/3 mm, 3%/2 mm, 2%/3 mm, and 2%/2 mm). The GPR obtained by comparing the dose distribution reconstructed using the independent Monte Carlo (MC) dose calculation software ArcherQA (Wisdom Technology Company Limited, Hefei, China)-based on linear accelerator delivery log files-with the original planned dose distribution was used as the reference standard, and calculated using global gamma analysis with a 10% dose threshold. Model performance was evaluated using the mean absolute error (MAE), root mean square error (RMSE), and Spearman's rank correlation coefficient. Shapley Additive Explanations (SHAP) were applied to interpret feature contributions in the best-performing model. RESULTS: Both models demonstrated robust predictive performance across different anatomical structures and gamma criteria. As the gamma criteria became less stringent, prediction errors decreased accordingly. Prediction accuracy was relatively high for OARs; for example, under the 3%/3 mm criterion, the test-set MAE was 0.06% ± 0.01% for the heart and 0.26% ± 0.04% for the whole lung. In contrast, the prediction error was relatively larger for the PTV, with a test-set MAE of 1.98% ± 0.31% under the same criterion. SHAP analysis revealed that texture-related radiomic features contributed most substantially to model predictions. Moreover, feature importance patterns varied according to organ type and gamma-criterion stringency. CONCLUSIONS: Multi-omics descriptors derived from anatomical structures can reliably predict GPRs in IMRT. The proposed interpretable ML framework not only achieves accurate prediction but also enhances mechanistic understanding through SHAP-based explanations. These findings provide valuable insights into dose verification variability and offer a practical, transparent tool for IMRT patient-specific quality assurance.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-06-24
DOI
https://doi.org/10.1002/acm2.70667
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Explainable machine learning for patient‐specific quality assurance in intensity‐modulated radiotherapy based on anatomical structures

Jianhao Xie, Xuerou Zhang, H Z Chen, Xi Zhang et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Explainable machine learning for patient‐specific quality assurance in intensity‐modulated radiotherapy based on anatomical structures

Jianhao Xie, Xuerou Zhang, H Z Chen, Xi Zhang, Yi-Xiang Wang, Y H Xu, Ying Huang, Zhiyong Xu, Yehui Luo
article en

Abstract

BACKGROUND: Patient-specific quality assurance (PSQA) plays a pivotal role in intensity-modulated radiotherapy (IMRT) to ensure accurate dose delivery. However, conventional measurement-based PSQA approaches are labor-intensive and provide limited insight into the underlying factors contributing to variations in gamma passing rates (GPRs). Anatomical characteristics of the planning target volume (PTV) and organs at risk (OARs) may contain predictive information relevant to GPR performance, yet their potential has not been fully explored within interpretable machine learning frameworks. PURPOSE: This study aimed to develop an interpretable machine learning (ML) framework for predicting GPRs in IMRT based on anatomical features extracted from the PTV and OARs. METHODS: A retrospective cohort of 243 clinical chest IMRT plans was analyzed. Radiomic and dosimetric features were extracted for each anatomical structure. Two ML regression models-Random Forest (RF) and eXtreme Gradient Boosting (XGBoost)-were developed to predict GPRs for the PTV and OARs under four gamma criteria (3%/3 mm, 3%/2 mm, 2%/3 mm, and 2%/2 mm). The GPR obtained by comparing the dose distribution reconstructed using the independent Monte Carlo (MC) dose calculation software ArcherQA (Wisdom Technology Company Limited, Hefei, China)-based on linear accelerator delivery log files-with the original planned dose distribution was used as the reference standard, and calculated using global gamma analysis with a 10% dose threshold. Model performance was evaluated using the mean absolute error (MAE), root mean square error (RMSE), and Spearman's rank correlation coefficient. Shapley Additive Explanations (SHAP) were applied to interpret feature contributions in the best-performing model. RESULTS: Both models demonstrated robust predictive performance across different anatomical structures and gamma criteria. As the gamma criteria became less stringent, prediction errors decreased accordingly. Prediction accuracy was relatively high for OARs; for example, under the 3%/3 mm criterion, the test-set MAE was 0.06% ± 0.01% for the heart and 0.26% ± 0.04% for the whole lung. In contrast, the prediction error was relatively larger for the PTV, with a test-set MAE of 1.98% ± 0.31% under the same criterion. SHAP analysis revealed that texture-related radiomic features contributed most substantially to model predictions. Moreover, feature importance patterns varied according to organ type and gamma-criterion stringency. CONCLUSIONS: Multi-omics descriptors derived from anatomical structures can reliably predict GPRs in IMRT. The proposed interpretable ML framework not only achieves accurate prediction but also enhances mechanistic understanding through SHAP-based explanations. These findings provide valuable insights into dose verification variability and offer a practical, transparent tool for IMRT patient-specific quality assurance.

Journal of Applied Clinical Medical PhysicsVol. 27(7)
University of Shanghai for Science and Technology (CN), Shanghai Chest Hospital (CN), Children's Hospital of Fudan University (CN)
Natural Science Foundation of Shanghai, National Natural Science Foundation of China
Decent work and economic growth
Openalex Percentile: Top 7%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.