The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety

Low-cost, scalable tests hold great potential to increase the availability of water safety data globally, especially in resource-limited settings. We evaluated the Lishtot TestDrop Pro, a commercially available, low-cost triboelectric-based sensor marketed for real-time determination of water contamination, including microbiological safety. We prepared a total of 199 water samples by serial dilution of raw wastewater influent into deionized water to produce a range of E. coli counts from <1 to too numerous to count (TNTC, > 300 per dilution plate) CFU/100 mL to evaluate the sensor. We tested these waters using the manufacturer’s instructions. We collected sensor readings using three settings on each of five devices, generating nearly 8,955 individual measurements to compare against E. coli measured via standard membrane filtration assays (EPA Method 1604), a common metric of drinking water safety. We found no statistical correlation between the sensor score and E. coli presence in water: 43% of sensor readings indicated “safe” when wastewater and E. coli was present, and 56% of samples lacking E. coli were deemed “unsafe” by the sensor when wastewater was absent. The Lishtot TestDrop Pro is not an accurate method for measuring microbiological water safety.

Authors

Institutions

Publication Details

Journal
PLOS Water
Published
2026-06-18
DOI
https://doi.org/10.1371/journal.pwat.0000442
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety

J. Wang, Timothy Purvis, Thao V. Nguyen, Joe Brown et al.
PLOS Water
Biosensors and Analytical Detection
article

The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety

J. Wang, Timothy Purvis, Thao V. Nguyen, Joe Brown, Siya Yeolekar, Eddy Ding, Kevin Zhu, Caitlin McHugh
article en

Abstract

Low-cost, scalable tests hold great potential to increase the availability of water safety data globally, especially in resource-limited settings. We evaluated the Lishtot TestDrop Pro, a commercially available, low-cost triboelectric-based sensor marketed for real-time determination of water contamination, including microbiological safety. We prepared a total of 199 water samples by serial dilution of raw wastewater influent into deionized water to produce a range of E. coli counts from <1 to too numerous to count (TNTC, > 300 per dilution plate) CFU/100 mL to evaluate the sensor. We tested these waters using the manufacturer’s instructions. We collected sensor readings using three settings on each of five devices, generating nearly 8,955 individual measurements to compare against E. coli measured via standard membrane filtration assays (EPA Method 1604), a common metric of drinking water safety. We found no statistical correlation between the sensor score and E. coli presence in water: 43% of sensor readings indicated “safe” when wastewater and E. coli was present, and 56% of samples lacking E. coli were deemed “unsafe” by the sensor when wastewater was absent. The Lishtot TestDrop Pro is not an accurate method for measuring microbiological water safety.

PLOS WaterVol. 5(6)
University of North Carolina at Chapel Hill (US), University of North Carolina Health Care (US)
Clean water and sanitation
Openalex Percentile: Top 11%
Biosensors and Analytical Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety — J. Wang, Timothy Purvis, et al. · PLOS Water (2026) | TGRS Research Map | TGRS